Kecheng KC-330B
|
Status |
in progress |
---|
Connectivity |
USB |
---|
Measurement range (A) |
30dB - 130dB |
---|
Resolution |
0.1dB |
---|
Accuracy (94dB@1kHz) |
1.5dB |
---|
Frequency weighting |
A, C |
---|
Time weighting |
F, S |
The Kecheng KC-330B is a sound level meter and data logger with USB connectivity.
See Kecheng KC-330B/Info for more details (such as lsusb -vvv output) about the device.
Hardware
Photos
Protocol
The host sends commands to the device via endpoint 2, and receives responses via endpoint 1. A command consists of a single byte, with the MSB (bit 7) cleared, followed by any arguments to the command. The response from the device consists of at least one byte: the command, with the MSB set.
Description
|
Command
|
Response
|
Configure
|
0x01 + 6 bytes
|
0x81
|
|
1
|
Sample rate 0-6, representing 125ms, 500ms, 1s, 2s, 5s, 10s, 60s respectively.
|
|
2
|
Alarm low threshold
|
|
3
|
Alarm high threshold
|
|
4
|
Time weighting: 0=Fast, 1=Slow
|
|
5
|
Frequency weighting: 0=dBA, 1=dBC
|
|
6
|
Data source: 0=memory, 1=real time
|
|
Identify
|
0x02
|
0x82 + length byte + device model (ASCII)
|
Set date and time
|
0x03 + 6 bytes
|
0x83
|
|
1
|
Last two digits of year, e.g. 0x0d for 2013
|
|
2
|
Month, 1-12
|
|
3
|
Day of month, 1-31
|
|
4
|
Hours
|
|
5
|
Minutes
|
|
6
|
Seconds
|
|
Check status
|
0x04
|
0x84
|
Start acquisition (depending on data source)
|
0x05
|
0x85 + 8 bytes
|
|
1-6
|
Configuration settings of acquired data, same format as 0x01 command (except 0x11...)
|
|
7
|
?
|
|
8
|
?
|
|
Get next measurement
|
0x08
|