UNI-T UT612
Revision as of 20:57, 13 July 2016 by Uwe Hermann (talk | contribs)
The UNI-T UT612 is a 20000 counts, dual-display, LCR meter with USB/HID connectivity.
See UNI-T UT612/Info for more details (such as lsusb -v output) about the device.
Hardware
- LCR meter chipset: Cyrustek ES51919 (datasheet)
- HID USB to UART bridge: Silicon Labs CP2110 (datasheet)
Photos
Protocol
The protocol for the ES51919 was reverse engineered as part of the DER-EE 5000 driver that uses the same chipset. Below is the excerpt from the source code.
Public official documentation does not contain the protocol description, so this is all based on reverse engineering. Packet structure (17 bytes): 0x00: header1 ?? (0x00) 0x01: header2 ?? (0x0d) 0x02: flags bit 0 = hold enabled bit 1 = reference shown (in delta mode) bit 2 = delta mode bit 3 = calibration mode bit 4 = sorting mode bit 5 = LCR mode bit 6 = auto mode bit 7 = parallel measurement (vs. serial) 0x03: config bit 0-4 = ??? (0x10) bit 5-7 = test frequency 0 = 100 Hz 1 = 120 Hz 2 = 1 kHz 3 = 10 kHz 4 = 100 kHz 5 = 0 Hz (DC) 0x04: tolerance (sorting mode) 0 = not set 3 = +-0.25% 4 = +-0.5% 5 = +-1% 6 = +-2% 7 = +-5% 8 = +-10% 9 = +-20% 10 = -20+80% 0x05-0x09: primary measurement 0x05: measured quantity 1 = inductance 2 = capacitance 3 = resistance 4 = DC resistance 0x06: measurement MSB (0x4e20 = 20000 = outside limits) 0x07: measurement LSB 0x08: measurement info bit 0-2 = decimal point multiplier (10^-val) bit 3-7 = unit 0 = no unit 1 = Ohm 2 = kOhm 3 = MOhm 5 = uH 6 = mH 7 = H 8 = kH 9 = pF 10 = nF 11 = uF 12 = mF 13 = % 14 = degree 0x09: measurement status bit 0-3 = status 0 = normal (measurement shown) 1 = blank (nothing shown) 2 = lines ("----") 3 = outside limits ("OL") 7 = pass ("PASS") 8 = fail ("FAIL") 9 = open ("OPEn") 10 = shorted ("Srt") bit 4-6 = ??? (maybe part of same field with 0-3) bit 7 = ??? (some independent flag) 0x0a-0x0e: secondary measurement 0x0a: measured quantity 0 = none 1 = dissipation factor 2 = quality factor 3 = parallel AC resistance / ESR 4 = phase angle 0x0b-0x0e: like primary measurement 0x0f: footer1 (0x0d) 0x10: footer2 (0x0a)
Resources
- Official UNI-T UT612 related documents (manual, software, ...)
- eevblog.com: Review and tear-down of UNI-T UT612 LCR meter