Difference between revisions of "Voltcraft DL-161S"
Jump to navigation
Jump to search
Uwe Hermann (talk | contribs) (Set status to Planned.) |
|||
(2 intermediate revisions by one other user not shown) | |||
Line 2: | Line 2: | ||
| image = [[File:Voltcraft DL-161S.png|180px]] | | image = [[File:Voltcraft DL-161S.png|180px]] | ||
| name = Voltcraft DL-161S | | name = Voltcraft DL-161S | ||
| status = | | status = planned | ||
| source_code_dir = | | source_code_dir = | ||
| connectivity = USB | | connectivity = USB | ||
Line 127: | Line 127: | ||
|- | |- | ||
| style="width: 1em;" | | | style="width: 1em;" | | ||
| style="width: 3em;" | 12 | | style="width: 3em;" | 12-14 | ||
| Number of samples to acquire, as a big-endian integer. | |||
| Number of samples to acquire. | |||
|- | |- | ||
| style="width: 1em;" | | | style="width: 1em;" | | ||
Line 150: | Line 142: | ||
| 0x0d | | 0x0d | ||
| 0xff | | 0xff | ||
| '''Get stored | | '''Get stored measurement count''', 2 bytes payload: | ||
|- | |- | ||
| style="width: 1em;" | | | style="width: 1em;" | | ||
Line 211: | Line 203: | ||
[[Category:Device]] | [[Category:Device]] | ||
[[Category:Sound level meter]] | [[Category:Sound level meter]] | ||
[[Category: | [[Category:Planned]] |
Latest revision as of 19:49, 20 December 2016
Status | planned |
---|---|
Connectivity | USB |
Frequency range | 31.5Hz - 8kHz |
Measurement range (A) | 30dB - 130dB |
Accuracy (94dB@1kHz) | 1.4dB |
Frequency weighting | A, C |
Time weighting | F, S |
Standards | IEC 61672-1 Class 2 |
Website | Voltcraft |
The Voltcraft DL-161S is a sound level meter with USB connectivity.
See Voltcraft DL-161S/Info for more details (such as lsusb -vvv output) about the device.
This is a rebadged CEM DT-173.
Hardware
TODO
Photos
Protocol
The host sends commands and receives responses via a virtual serial port, with USB bulk endpoint 2 as a backend. A command consists of a single byte followed by any arguments to the command.
Command | Response | Description | |||||||||||||||||||||||||||||||||||||||||||||||||||
---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
0x0c | 0xff | Calibrate device, 2 bytes payload: | |||||||||||||||||||||||||||||||||||||||||||||||||||
1 | Calibration value, as a signed integer between -10 and 10, where 1 represents 0.1 dB | ||||||||||||||||||||||||||||||||||||||||||||||||||||
2 | Unused (0x00) | ||||||||||||||||||||||||||||||||||||||||||||||||||||
0x0e | 0xff | Configure device | |||||||||||||||||||||||||||||||||||||||||||||||||||
| |||||||||||||||||||||||||||||||||||||||||||||||||||||
0x0d | 0xff | Get stored measurement count, 2 bytes payload: | |||||||||||||||||||||||||||||||||||||||||||||||||||
1-2 | Always 0x0000? | ||||||||||||||||||||||||||||||||||||||||||||||||||||
Response has 3 byte payload, representing the number of bytes used in sample memory as a little-endian integer. Each sample uses two bytes. | |||||||||||||||||||||||||||||||||||||||||||||||||||||
0x0f | Get stored measurements, 2 bytes payload is offset into buffer (as pages? page size?) | ||||||||||||||||||||||||||||||||||||||||||||||||||||
|