The ''UNI-T ut612 LCR Meter Dual Display'' comes with an USB interface that encapsulates standard serial over USB/HID. The main chipset inside the device responsible for the meassurements is the Cyrustek ES51919, the serial output from this chip is sent to a Silicon Labs CP2110 HID USB to UART Bridge.
The '''UNI-T UT612''' is a 20000 counts, dual-display, LCR meter with USB/HID connectivity.


== Protocol ES51919 ==
See [[UNI-T UT612/Info]] for more details (such as '''lsusb -v''' output) about the device.


The protocol for the es51919 was reverse engineered as part of the DER-EE 5000 driver that uses the same chipset. Below is the excerpt from the source code.
== Hardware ==


Public official documentation does not contain the protocol
* '''LCR meter chipset''': [http://www.cyrustek.com.tw/product-1-LCR%20Meter%20Series.htm Cyrustek ES51919] ([http://www.cyrustek.com.tw/spec/ES51920.pdf datasheet])
description, so this is all based on reverse engineering.
* '''HID USB to UART bridge''': [http://www.silabs.com/products/interface/usb-bridges/classic-usb-bridges/pages/usb-to-uart-bridge.aspx Silicon Labs CP2110] ([https://www.silabs.com/Support%20Documents/TechnicalDocs/CP2110.pdf datasheet])
 
== Photos ==
 
== Protocol ==
 
The protocol for the ES51919 was [http://sigrok.org/gitweb/?p=libsigrok.git;a=blob;f=src/lcr/es51919.c;h=0244851a2b6c9594ab093dc4d7429523fcc9e791;hb=HEAD#l310 reverse engineered as part of the DER-EE 5000 driver] that uses the same chipset. Below is the excerpt from the source code.
 
<small>
<pre>
Public official documentation does not contain the protocol description, so this is all based on reverse engineering.


Packet structure (17 bytes):
Packet structure (17 bytes):
Line 97: Line 107:
0x10: footer2 (0x0a)
0x10: footer2 (0x0a)


=== Protocol cp2110 ===
</pre>
</small>


[https://www.silabs.com/Support%20Documents/TechnicalDocs/CP2110.pdf CP2110 Data sheet]
== Resources ==


== Images ==
* [https://drive.google.com/folderview?id=0B4Jyby-tjH5oX2NEZk1xZXVlNlE&usp=drive_web Official UNI-T UT612 related documents] (manual, software, ...)
* [http://www.eevblog.com/forum/testgear/review-and-tear-down-of-uni-t-ut612-lcr-meter/ eevblog.com: Review and tear-down of UNI-T UT612 LCR meter]


[[Category:Device]]
[[Category:Device]]
[[Category:LCR meter]]
[[Category:LCR meter]]
[[Category:Planned]]
[[Category:Planned]]

Revision as of 20:57, 13 July 2016

The UNI-T UT612 is a 20000 counts, dual-display, LCR meter with USB/HID connectivity.

See UNI-T UT612/Info for more details (such as lsusb -v output) about the device.

Hardware

Photos

Protocol

The protocol for the ES51919 was reverse engineered as part of the DER-EE 5000 driver that uses the same chipset. Below is the excerpt from the source code.

Public official documentation does not contain the protocol description, so this is all based on reverse engineering.

Packet structure (17 bytes):

0x00: header1 ?? (0x00)

0x01: header2 ?? (0x0d)

0x02: flags
        bit 0 = hold enabled
        bit 1 = reference shown (in delta mode)
        bit 2 = delta mode
        bit 3 = calibration mode
        bit 4 = sorting mode
        bit 5 = LCR mode
        bit 6 = auto mode
        bit 7 = parallel measurement (vs. serial)

0x03: config
        bit 0-4 = ??? (0x10)
        bit 5-7 = test frequency
                    0 = 100 Hz
                    1 = 120 Hz
                    2 = 1 kHz
                    3 = 10 kHz
                    4 = 100 kHz
                    5 = 0 Hz (DC)

0x04: tolerance (sorting mode)
        0 = not set
        3 = +-0.25%
        4 = +-0.5%
        5 = +-1%
        6 = +-2%
        7 = +-5%
        8 = +-10%
        9 = +-20%
       10 = -20+80%

0x05-0x09: primary measurement
  0x05: measured quantity
          1 = inductance
          2 = capacitance
          3 = resistance
          4 = DC resistance
  0x06: measurement MSB  (0x4e20 = 20000 = outside limits)
  0x07: measurement LSB
  0x08: measurement info
          bit 0-2 = decimal point multiplier (10^-val)
          bit 3-7 = unit
                      0 = no unit
                      1 = Ohm
                      2 = kOhm
                      3 = MOhm
                      5 = uH
                      6 = mH
                      7 = H
                      8 = kH
                      9 = pF
                      10 = nF
                      11 = uF
                      12 = mF
                      13 = %
                      14 = degree

0x09: measurement status
          bit 0-3 = status
                      0 = normal (measurement shown)
                      1 = blank (nothing shown)
                      2 = lines ("----")
                      3 = outside limits ("OL")
                      7 = pass ("PASS")
                      8 = fail ("FAIL")
                      9 = open ("OPEn")
                     10 = shorted ("Srt")
          bit 4-6 = ??? (maybe part of same field with 0-3)
          bit 7   = ??? (some independent flag)

0x0a-0x0e: secondary measurement
  0x0a: measured quantity
          0 = none
          1 = dissipation factor
          2 = quality factor
          3 = parallel AC resistance / ESR
          4 = phase angle
  0x0b-0x0e: like primary measurement

0x0f: footer1 (0x0d)

0x10: footer2 (0x0a)

Resources