Agilent 54600 Series

From sigrok
Jump to navigation Jump to search
Agilent 54600 series
Agilent 54622A Front.jpg
Status Not Supported‏‎
Source code ?
Channels 2-4 analog, 16 digital (D models only)
Samplerate 200-2000MSa/s (depending on model)
Analog bandwidth 60-500MHz (depending on model)
Vertical resolution 8bits
Triggers edge, pulse width, pattern, video, duration, sequence, CAN, USB, LIN, I2C, SPI
Input impedance 1MΩ‖13pF 300Vrms CAT I, 100Vrms CAT II. 50Ω‖13pF 5Vrms (depending on model)
Memory 4MSa-8MSa (mode/ch-dependent) (depending on model)
Display 7" raster monochrome CRT
Connectivity 3.5" 1.44MiB floppy, RS232, parallel port
Website Keysight Technologies

Models

Model Bandwidth (Hz) Max Sample Rate (Sa/s) Memory (MB) Digital Inputs Analog Inputs Input Impedance (Ω)
54621A 60 M 200 M 4 2 1 M
54621D 60 M 200 M 4 16 2 1 M
54622A 100 M 200 M 4 2 1 M
54622D 100 M 200 M 4 16 2 1 M
54624A 100 M 200 M 4 4 1 M
54641A 350 M 2 G 8 2 1 M, 50
54641D 350 M 2 G 8 16 2 1 M, 50
54642A 500 M 2 G 8 2 1 M, 50
54642D 500 M 2 G 8 16 2 1 M, 50


Protocol

Communication with the 54600 Series is over the RS232 port on the back, or via the floppy drive. The RS232 protocol is described in the Programmer's Guide.

Links

Floppy Drive

The floppy drive is a TEAC FD-05HF.

The Gotek SFR1M44-DU26 USB floppy emulator is compatible once modified & flashed with FlashFloppy firmware. See TDS3000 series oscilloscope floppy drive replacement for an overview of the process

Resources