Loto OSCxxx series
Jump to navigation
Jump to search
The printable version is no longer supported and may have rendering errors. Please update your browser bookmarks and please use the default browser print function instead.
Status | planned |
---|---|
Channels | 2 analog (all models) + trigger |
Samplerate | 50-200 MSa/s (by channel in dual channel) |
Samplerate (eq. time) | 0.2-1 GSa/s (dual channel combined or single channel) |
Analog bandwidth | 20-50 MHz (depending on model) |
Vertical resolution | 8bits |
Triggers | edge, pulse width, slope, video, alternate |
Input impedance | 1MΩ‖25pF±5pF |
Memory | 64Kpts (by channel, 128K total) |
Connectivity | USB host and device |
Features | math: + / — / x / FFT / Digital decoder |
Website | rockemb.com |
The Loto OSCxxx series are compact and portable digital storage oscilloscopes, produced by Rocktech.
Devices
Model | Bandwidth (MHz) | Sample Rate (MS/s by ch) | USB VID/PID | Notes |
---|---|---|---|---|
Loto OSCH02 | 100 | 250 | 0x8312:0x8a2a | Rel. 2020 (250MSa just CH-A) |
Loto OSC2002 | 50 | 200 | 0x8312:0x8801 | 2018-2019 edition |
Loto OSCA02 | 35 | 100 | 0x8312:0x8a02 | |
Loto OSC802 | 25 | 80 | 0x8312:0x8312 | |
Loto OSC482 | 20 | 50 | 0x8102:0x8102 | |
Loto BM102 | 20 | 50 | 0x8102:0x8102 | out of production |
Loto OSCE02 | 50 | 200 | 0x8312:0x8101 |
Protocol
OSC802
Commands sent via Control Transfer to EP0. Samples buffer available on Bulk EP. List of commands (provisional):
Request Type | Request | Value | Index | Data | Description |
---|---|---|---|---|---|
0x80 | 0x33 | 0x00 | 0x00 | 1 byte | Start acquisition. |
0x80 | 0x50 | 0x00 | 0x00 | 1 byte | Check if acquisition buffer is ready. Write 0x21 in the data byte if ready. |
0x80 | 0x94 | value | 0x00 | 1 byte | Set sampling rate with value[0-3] and AC/DC for Channel 1 with value[4].
Rate values: 80MSa/s (B0000), 10MSa/s (B1000), 625KSa/s (B1100) value[4] = 1 -> DC coupling, value[4] = 0 -> AC coupling |
0x80 | 0x22 0x23 0x24 | value | 0x00 | 1 byte | Set Channel 1 and Channel 2 voltage. Set AC/DC for Channel 2. |
0x80 | 0xE7 | value | 0x00 | 1 byte | Enable (value=1) or disable (value=0) hardware trigger on Channel 1. |
0x80 | 0xC5 | value | 0x00 | 1 byte | Set trigger slope up (value=1) or down (value=0). |
0x80 | 0x16 | value | 0x00 | 1 byte | Set trigger level (value=0-255). |