Loto OSCxxx series

From sigrok
(Redirected from Loto OSCxxx)
Jump to navigation Jump to search
The printable version is no longer supported and may have rendering errors. Please update your browser bookmarks and please use the default browser print function instead.
Loto OSCxxx series
Status planned
Channels 2 analog (all models) + trigger
Samplerate 50-200 MSa/s (by channel in dual channel)
Samplerate (eq. time) 0.2-1 GSa/s (dual channel combined or single channel)
Analog bandwidth 20-50 MHz (depending on model)
Vertical resolution 8bits
Triggers edge, pulse width, slope, video, alternate
Input impedance 1MΩ‖25pF±5pF
Memory 64Kpts (by channel, 128K total)
Connectivity USB host and device
Features math: + / — / x / FFT / Digital decoder
Website rockemb.com

The Loto OSCxxx series are compact and portable digital storage oscilloscopes, produced by Rocktech.

Devices

Model Bandwidth (MHz) Sample Rate (MS/s by ch) USB VID/PID Notes
Loto OSCH02 100 250 0x8312:0x8a2a Rel. 2020 (250MSa just CH-A)
Loto OSC2002 50 200 0x8312:0x8801 2018-2019 edition
Loto OSCA02 35 100 0x8312:0x8a02
Loto OSC802 25 80 0x8312:0x8312
Loto OSC482 20 50 0x8102:0x8102
Loto BM102 20 50 0x8102:0x8102 out of production
Loto OSCE02 50 200 0x8312:0x8101

Protocol

OSC802

Commands sent via Control Transfer to EP0. Samples buffer available on Bulk EP. List of commands (provisional):

Request Type Request Value Index Data Description
0x80 0x33 0x00 0x00 1 byte Start acquisition.
0x80 0x50 0x00 0x00 1 byte Check if acquisition buffer is ready. Write 0x21 in the data byte if ready.
0x80 0x94 value 0x00 1 byte Set sampling rate with value[0-3] and AC/DC for Channel 1 with value[4].

Rate values: 80MSa/s (B0000), 10MSa/s (B1000), 625KSa/s (B1100)

value[4] = 1 -> DC coupling, value[4] = 0 -> AC coupling

0x80 0x22 0x23 0x24 value 0x00 1 byte Set Channel 1 and Channel 2 voltage. Set AC/DC for Channel 2.
0x80 0xE7 value 0x00 1 byte Enable (value=1) or disable (value=0) hardware trigger on Channel 1.
0x80 0xC5 value 0x00 1 byte Set trigger slope up (value=1) or down (value=0).
0x80 0x16 value 0x00 1 byte Set trigger level (value=0-255).

Resources