The metadata following 0xaa and 0xcc is 7 bytes long:
The metadata following 0xaa and 0xcc denotes the date and time this record was started, and the sampling rate. It is encoded in 7 bytes:


{| border="0" width="18em" class="alternategrey sigroktable"
{| border="0" width="95%" class="alternategrey sigroktable"
|-
|-
| 0xbb
! style="width: 3em; text-align: center;" | Byte
! style="text-align: center;" | Description
|-
|-
| 0xxx
| 0
|- 0xpp
| Last two digits of year (BCD)
|-
| 1
| Month (BCD)
|-
| 2
| Day of month (BCD)
|-
| 3
| Hour (BCD)
|-
| 4
| Minutes (BCD)
|-
| 5
| Seconds (BCD)
|-
| 6
| Samplerate
|}
|}



Revision as of 18:09, 6 June 2013

CEM DT-8852
CEM DT-8852.png
Status in progress
Connectivity USB
Frequency range 31.5Hz - 8kHz
Measurement range (A) 30dB - 130dB
Resolution 0.1dB
Accuracy (94dB@1kHz) 1.4dB
Frequency weighting A, C
Time weighting F, S
Standards IEC 61672-1 Class 2
Website cem-instruments.com

The CEM DT-8852 is an IEC 61672-1 class 2 compliant sound level meter with USB connectivity.

See CEM DT-8852/Info for more details (such as lsusb -vvv output) about the device.

Some random facts:

  • The device only starts logging to USB when the SETUP key is pressed on the keypad. This also disables auto power-off mode.
  • Consumes 9.5mA, regardless of USB or recording status; 15mA when the backlight is on.
  • The battery low warning starts when the battery delivers less than 7V.
  • Sound pressure level measurements are sent to the host at a rate of 20Hz.

Hardware

Photos

Protocol

The device communicates at 9600 baud, no parity, 8 data bits, and 1 stop bit (9600/8n1).

Measurements and settings

Current settings and measurements are continually streamed over the serial bus, without any prompting from the PC. Data is encapsulated in packets of between two and five bytes. The packets are structured as follows:

1 2 3 4 5
0xa5 Token Data

The number of data bytes depend on the token. These are the tokens:

Token Data Description
0x02 0 Time weighting Fast
0x03 0 Time weighting Slow
0x04 0 Max hold mode
0x05 0 Min hold mode
0x06 3 Current time in BCD; first nibble unknown
0x07 0 Current measurement is over measurement range high threshold
0x08 0 Current measurement is under measurement range low threshold
0x0a 0
0x0b 1
0x0c 0
0x0d 2 Current measurement as shown on the display, multiplied by 10, in BCD
0x0e 0 Live measurements mode (not max/min hold)
0x0f 0 Battery is low
0x11 0
0x19 0
0x1a 0
0x1b 1 Frequency weighting dBA
0x1c 1 Frequency weighting dBC
0x1f 0
0x30 0 Measurement range 30-80 dB
0x40 0 Measurement range 30-130 dB (auto)
0x4b 0 Measurement range 50-100 dB
0x4c 0 Measurement range 80-130 dB

Recorded measurements

After the host has requested the recorded measurement log, a packet will be inserted into the regular output stream. Instead of 0xa5, this packet begins with 0xbb:

1 2 3
0xbb Length

The length field is big-endian encoded, with an offset of 100. A length field of 0x0064 thus indicates no stored samples.

This is followed by a series of records:

Token Description
0xaa Measurements in this record use dBA frequency weighting. Metadata follows.
0xcc Measurements in this record use dBC frequency weighting. Metadata follows.
0xac Measurements follow, until the next token. Encoding is BCD time 10, as in the live stream.
0xdd End of recorded measurement dump.

The metadata following 0xaa and 0xcc denotes the date and time this record was started, and the sampling rate. It is encoded in 7 bytes:

Byte Description
0 Last two digits of year (BCD)
1 Month (BCD)
2 Day of month (BCD)
3 Hour (BCD)
4 Minutes (BCD)
5 Seconds (BCD)
6 Samplerate

Resources