Difference between revisions of "Rohde&Schwarz HMO 3000 series"
Line 4: | Line 4: | ||
| status = supported | | status = supported | ||
| source_code_dir = hameg-hmo | | source_code_dir = hameg-hmo | ||
| channels = 2 | | channels = 2 or 4 analog, 16 digital | ||
| samplerate = 4GSa/s (interleaved), 2GSa/s (non interleaved) | | samplerate = 4GSa/s (interleaved), 2GSa/s (non interleaved) | ||
| bandwidth = 300-500MHz (depending on model) | | bandwidth = 300-500MHz (depending on model) |
Revision as of 23:48, 22 October 2018
File:HMO3000.jpg | |
Status | supported |
---|---|
Source code | hameg-hmo |
Channels | 2 or 4 analog, 16 digital |
Samplerate | 4GSa/s (interleaved), 2GSa/s (non interleaved) |
Analog bandwidth | 300-500MHz (depending on model) |
Vertical resolution | 8 bits (HiRes up to 16 bits) |
Triggers | slope, pulse width, logic, video, risetime, runt, serial bus (optional) |
Input impedance | 1MΩ‖13pF 200Vp CAT I, 50Ω 5V RMS |
Memory | 4 or 8 Mpts (depending on number of channels used) |
Display | 6.5" VGA (640x480), 256 colors |
Connectivity | USB host/device, Ethernet, RS232 (optional), GPIB (optional) |
Features | math (+, —, x, /), FFT, pattern generator, statistics, frequency counter, digital voltmeter, vertical sensitivity: 1mV/div - 5V/div |
Website | rohde-schwarz.com |
The Rohde&Schwarz HMO3000 series are 300-500MHz, 4GSa/s, 2-4 analog channels and 16 digital channels mixed-signal oscilloscopes. This series was previously branded Hameg.
All models support different extension modules which provide different connectivity options.
Most models come with a HO730 (USB, Ethernet) module, other modules like the HO720 (RS232, USB) and the HO740 (GPIB) are available optionally.
All models come with analog probes (specific probe might depend on the scope capabilities).
Digital probes (8 or 16 channels) are optional and can be purchased separately, see HO3508.
The HMO3000 series has been discontinued in the first-half of 2018.
Devices
Model | Bandwidth | Analog Channels | Digital Channels |
---|---|---|---|
HMO3032 | 300MHz | 2 | 16 |
HMO3034 | 300MHz | 4 | 16 |
HMO3042 | 400MHz | 2 | 16 |
HMO3044 | 400MHz | 4 | 16 |
HMO3052 | 500MHz | 2 | 16 |
HMO3054 | 500MHz | 4 | 16 |
Protocol
All the devices use the SCPI protocol (for reference, see the "SCPI Programmers Manual").
The serial parameters are configurable on the device, the standard settings are: 8n1, no handshake, 115200 baud.
All models support USB connection to the computer in both TMC and VPC mode, as well as Ethernet connection (raw TCP over a user-configurable port).