Difference between revisions of "Agilent U12xxx series"
(Created page with "The Agilent U12xxx series are handheld digital multimeters, ranging from inexpensive all-purpose meters to industrial clamp meters. {| border="0" style="width: 40em; font-siz...") |
(→CONF) |
||
Line 81: | Line 81: | ||
=== CONF === | === CONF === | ||
The <code>CONF?</code> command queries the device's current measurement configuration. | The <code>CONF?</code> command queries the device's current measurement configuration. | ||
==== U123xx models ==== | ==== U123xx models ==== | ||
Line 307: | Line 266: | ||
| '''Diode''' | | '''Diode''' | ||
|} | |||
==== U124xx and U125xx models ==== | |||
The output consists of the meter's mode, followed by a space, and then the meter's upper and lower ranges in that mode, separated by a comma. The mode may optionally include a colon followed by a code denoting a submode relevant to that major mode. | |||
For example, this is the AC voltage in the 1000mV range: | |||
VOLT:AC +1.000000E+00,+1.000000E-04 | |||
{| border="0" width="95%" style="font-size: smaller" class="alternategrey sortable sigroktable" | |||
|- | |||
! style="width: 5em;" | Mode | |||
! Meaning | |||
|- | |||
| VOLT | |||
| Voltage, submode can be AC | |||
|- | |||
| CURR | |||
| Current, submode can be AC | |||
|- | |||
| FREQ | |||
| Frequency counter | |||
|- | |||
| DIOD | |||
| Diode check | |||
|- | |||
| CONT | |||
| Continuity | |||
|- | |||
| RES | |||
| Resistance | |||
|- | |||
| CAP | |||
| Capacitance | |||
|- | |||
| SCOU | |||
| Switch counter | |||
|- | |||
| CPER | |||
| Current loop, submode can be e.g. <code>0-20mA</code> or <code>4-20mA</code>. | |||
|- | |||
| T1 or T2 | |||
| Temperature, submode can be K or J, denoting the thermocouple type. Instead of range, CEL or FAR is given, denoting the measurement is in Celcius or Fahrenheit. | |||
|} | |} | ||
Revision as of 12:51, 30 August 2014
The Agilent U12xxx series are handheld digital multimeters, ranging from inexpensive all-purpose meters to industrial clamp meters.
Model | Counts | Supported |
---|---|---|
U1231A | 6000 | Yes |
U1232A | 6000 | Yes |
U1233A | 6000 | Yes |
U1241A / U1241B | 10000 | Yes |
U1242A / U1242B | 10000 | Yes |
U1251A / U1251B | 50000 | Yes |
U1252A / U1252B | 50000 | Yes |
U1253A / U1253B | 50000 | Yes |
U1271A | 30000 | No |
U1272A | 30000 | No |
U1273A | 30000 | No |
U1273AX | 30000 | No |
Is your model not supported yet? Get in touch! We need people to help us test other models in this series, and getting a new model supported is generally quick & easy.
Connection
All models communicate via an optically isolated serial connection, at 9600 bps, 8 bits, no parity, and 1 stop bit (9600/8n1). Connecting to the devices can be done either via USB (U1173A cable) or Bluetooth (U1177A interface).
Protocol
All models in this series use generally the same protocol, with only minor differences between models -- and none within the same range.
IDN
Sending the command *IDN?
causes the device to return a comma-separated four-field string, compatible with the identical GPIB command. The four fields are vendor name, model name, serial number and version:
Agilent Technologies,U1232A,MY52020136,V1.00
FETC
The FETC?
command returns the current measured value, as a floating point ASCII string:
+9.25000000E-03
The value always takes exactly this form. If no valid measurement is available, i.e. the display shows "OL", the transmitted value is +9.90000000E+37
.
CONF
The CONF?
command queries the device's current measurement configuration.
U123xx models
This can be either one, two or three comma-separated fields. The number of fields depends on the first field's value, but the second field always denotes the measurement mode's range and resolution, and the third field is the current type (AC or DC).
Mode | Resolution | Current | Meaning |
---|---|---|---|
V | y | y | Volts |
0 | Range 600mV, resolution 0.1mV | ||
1 | Range 6V, resolution 1mV | ||
2 | Range 60V, resolution 0.01V | ||
3 | Range 600V, resolution 0.1V | ||
MV | y | y | Millivolts |
1 | Range 600mV, resolution 0.1mV | ||
A | y | y | Ampere |
0 | Range 6A, resolution 1mA | ||
1 | Range 10A, resolution 0.01A | ||
UA | y | y | Microampere |
0 | Range 60μA, resolution 0.01μA | ||
1 | Range 600μA, resolution 0.1μA | ||
FREQ | y | y | AC frequency |
0 | Always shows 0, even when the range is changed manually on the device. This is possibly a firmware bug. | ||
RES | y | n | Resistance |
0 | Range 600Ω, resolution 0.1Ω | ||
1 | Range 6kΩ, resolution 0.001kΩ | ||
2 | Range 60kΩ, resolution 0.01kΩ | ||
3 | Range 600kΩ, resolution 0.1kΩ | ||
4 | Range 6MΩ, resolution 0.001MΩ | ||
5 | Range 60MΩ, resolution 0.01MΩ | ||
CAP | y | n | Capacitance |
0 | Range 1000nF, resolution 1nF | ||
1 | Range 10μF, resolution 0.01μF | ||
2 | Range 100μF, resolution 0.1μF | ||
3 | Range 1000μF, resolution 1μF | ||
4 | Range 10mF, resolution 0.01mF | ||
DIOD | n | n | Diode |
U124xx and U125xx models
The output consists of the meter's mode, followed by a space, and then the meter's upper and lower ranges in that mode, separated by a comma. The mode may optionally include a colon followed by a code denoting a submode relevant to that major mode. For example, this is the AC voltage in the 1000mV range:
VOLT:AC +1.000000E+00,+1.000000E-04
Mode | Meaning |
---|---|
VOLT | Voltage, submode can be AC |
CURR | Current, submode can be AC |
FREQ | Frequency counter |
DIOD | Diode check |
CONT | Continuity |
RES | Resistance |
CAP | Capacitance |
SCOU | Switch counter |
CPER | Current loop, submode can be e.g. 0-20mA or 4-20mA .
|
T1 or T2 | Temperature, submode can be K or J, denoting the thermocouple type. Instead of range, CEL or FAR is given, denoting the measurement is in Celcius or Fahrenheit. |
STAT
The STAT?
command is used to query the current state of the device. The response is a quoted 21-byte string, like this:
"000000000110L00000000"
U123xx models
Byte | Value | Meaning |
---|---|---|
1 | Max/min/avg mode | |
0 | Off | |
1 | On | |
2 | Relative measurement. The last value returned from FETC? before this mode was enabled denotes the stored reference value. | |
0 | Off | |
1 | On | |
3 | Trig hold-log | |
0 | Off | |
1 | On | |
4 | Auto hold-log | |
0 | Off | |
1 | On | |
5 | LED flashlight | |
0 | Off | |
1 | On | |
6 | LCD backlight | |
0 | Off | |
1 | On | |
7 | Smoothing | |
0 | Off | |
1 | On | |
8 | Temp/aux mode. When on and rotary switch is in position 5 (capacitance), CONF? returns MV,1,DC , but the value returned from FETC? instead denotes the temperature, in the unit (F or C) according to the meter's configuration.
| |
0 | Off | |
1 | On | |
9 | Unknown (always 0) | |
10 | Beep frequency | |
0 | 4.2kHz | |
1 | 3.8kHz (default) | |
2 | 3.4kHz | |
3 | 3.2kHz | |
4 | Off | |
11 | Auto power-off (APO) | |
0 | Off | |
1 | On | |
12 | Unknown (always 0) | |
13 | Unknown (always L) | |
14 | Unknown (always 0) | |
15 | Unknown (always 0) | |
16 | Rotary switch position | |
0 | V/Zlow | |
1 | Voltage AC | |
2 | Voltage DC | |
3 | Resistance | |
4 | Diode | |
5 | Capacitance | |
6 | Current | |
7 | Microcurrent | |
17 | Continuity mode. When enabled, the value returned from FETC? is either NAN (no continuity) or any real value to denote continuity. | |
0 | Off | |
1 | On | |
18 | Unknown (always 0) | |
19 | Unknown (always 0, but briefly 1 at startup) | |
20 | Unknown (always 0) | |
21 | Unknown (always 0) |