Difference between revisions of "Voltcraft DL-161S"
Jump to navigation
Jump to search
Line 2: | Line 2: | ||
| image = [[File:Voltcraft DL-161S.png|180px]] | | image = [[File:Voltcraft DL-161S.png|180px]] | ||
| name = Voltcraft DL-161S | | name = Voltcraft DL-161S | ||
| status = | | status = in progress | ||
| source_code_dir = | | source_code_dir = | ||
| connectivity = USB | | connectivity = USB | ||
Line 211: | Line 211: | ||
[[Category:Device]] | [[Category:Device]] | ||
[[Category:Sound level meter]] | [[Category:Sound level meter]] | ||
[[Category: | [[Category:In progress]] |
Revision as of 08:50, 26 July 2013
Status | in progress |
---|---|
Connectivity | USB |
Frequency range | 31.5Hz - 8kHz |
Measurement range (A) | 30dB - 130dB |
Accuracy (94dB@1kHz) | 1.4dB |
Frequency weighting | A, C |
Time weighting | F, S |
Standards | IEC 61672-1 Class 2 |
Website | Voltcraft |
The Voltcraft DL-161S is a sound level meter with USB connectivity.
See Voltcraft DL-161S/Info for more details (such as lsusb -vvv output) about the device.
This is a rebadged CEM DT-173.
Hardware
TODO
Photos
Protocol
The host sends commands and receives responses via a virtual serial port, with USB bulk endpoint 2 as a backend. A command consists of a single byte followed by any arguments to the command.
Command | Response | Description | |||||||||||||||||||||||||||||||||||||||||||||||||||||||||
---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
0x0c | 0xff | Calibrate device, 2 bytes payload: | |||||||||||||||||||||||||||||||||||||||||||||||||||||||||
1 | Calibration value, as a signed integer between -10 and 10, where 1 represents 0.1 dB | ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
2 | Unused (0x00) | ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
0x0e | 0xff | Configure device | |||||||||||||||||||||||||||||||||||||||||||||||||||||||||
| |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
0x0d | 0xff | Get stored measuremen count, 2 bytes payload: | |||||||||||||||||||||||||||||||||||||||||||||||||||||||||
1-2 | Always 0x0000? | ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Response has 3 byte payload, representing the number of stored samples as a little-endian integer. | |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
0x0f | Get stored measurements, 2 bytes payload is offset into buffer (as pages? page size?) | ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
|