{| border="0" width="95%" style="font-size: smaller;" class="alternategrey sigroktable"
|-
! style="width: 10em;" | Bytes
!Command
 
|-
| valign="top" style="white-space=nowrap;" | '''0x00 0xff 0xff'''
| Load device configuration structure.
 
A 3-byte header is returned first in which the first byte is 0x02, and the second and third bytes are the length of the configuration structure to follow, as a little-endian integer representing the number of bytes. Different devices in the EL-USB series have differently-sized configuration structures.
 
|-
| valign="top" style="white-space=nowrap;" | '''0x01''' ''0xnn 0xnn''
| Save device configuration structure.
 
The second and third bytes are the length of the configuration structure to follow, as a little-endian integer representing the number of bytes.
 
The device responds with a single byte with value 0xff when the configuration has been saved.
 
|-
| valign="top" style="white-space=nowrap;" | '''0x03 0xff 0xff'''
| Transfer logged data.
 
A 3-byte header is returned first in which the first byte is 0x02, and the second and third bytes are the length of the logged data to follow, as a little-endian integer representing the number of bytes.
 
The device's entire sample memory is then transferred. The number of stored samples (offset 0x2e in the configuration structure) should be checked before interpreting the data; any remaining data in the transfer should be ignored.
 
|}
 
=== Data buffer ===
 
Every sample consists of a byte representing the temperature and a byte representing relative humidity. There is no timestamp; this must be calculated using the configuration structure's start time and sample period.
 
For temperature measurements in Celcius, the value represents a point on a scale of -40C to 80C, with each point representing a half degree. Thus, to convert to Celcius:
 
temperature = value / 2 - 40
 
For Fahrenheit, the stored value also starts at -40F, but each point represents 1 degree Fahrenheit:
 
temperature = value - 40
 
The relative humidity value is represented as each point representing 0.5:
 
relative_humidity = value / 2
 
=== Device configuration structure ===
 
{| border="0" width="95%" style="font-size: smaller" class="alternategrey sigroktable"
|-
! style="width: 3em;" | Offset
! style="width: 3em;" | Size
! Value
 
|-
| valign="top" | 0x00
| valign="top" | 1
| '''Device type'''
{| border="0" width="100%" style="margin: 0;" class="alternategrey"
|-
| style="width: 3em;" | 1,2
| EL-USB-1
|-
| 3
| EL-USB-2
|-
| 4,6
| EL-USB-3
|-
| 5,7
| EL-USB-4
|-
| 8
| EL-USB-LITE
|-
| 9
| EL-USB-CO
|-
| 10
| EL-USB-TC
|-
| 11
| EL-USB-CO300
|-
| 12
| EL-USB-2-LCD
|-
| 13
| EL-USB-2+
|-
| 14
| EL-USB-1-PRO
|-
| 15
| EL-USB-TC-LCD
|-
| 16
| EL-USB-2-LCD+
|-
| 17
| EL-USB-5
|-
| 18
| EL-USB-1-RCG
|-
| 19
| EL-USB-1-LCD
|-
| 20
| EL-OEM-3
|-
| 21
| EL-USB-1-LCD
|}
 
|-
| 0x01
| 1
| Unused
 
|-
| 0x02
| 16
| NULL-terminated device name, 15 chars max
 
|-
| 0x12
| 1
| Start time, hour (0-23)
 
|-
| 0x13
| 1
| Start time, minute (0-59)
 
|-
| 0x14
| 1
| Start time, second (0-59)
 
|-
| 0x15
| 1
| Start date, day (1-31)
 
|-
| 0x16
| 1
| Start date, month (1-12)
 
|-
| 0x17
| 1
| Start date, year (year - 2000)
 
|-
| 0x18
| 4
| Unsigned little-endian, seconds remaining until logging starts
 
|-
| 0x1c
| 2
| Unsigned little-endian, samplerate (seconds between samples)
 
|-
| 0x1e
| 2
| Unsigned little-endian, number of stored samples
 
|-
| valign="top" | 0x20
| valign="top" | 1
| style="padding: 0;" |
{| border="0" width="100%" style="margin: 0;" class="alternategrey"
|-
! align="left" style="width: 3em;" | Bit
! align="left" | Alarm conditions bitfield (1=enable)
|-
| 0
| temperature high
|-
| 1
| temperature low
|-
| 2
| temperature high hold
|-
| 3
| temperature low hold
|-
| 4
| relative humidity high
|-
| 5
| relative humidity low
|-
| 6
| relative humidity high hold
|-
| 7
| relative humidity low hold
|}
 
|-
| valign="top" | 0x21
| valign="top" | 1
| style="padding: 0;" |
{| border="0" width="100%" style="margin: 0;" class="alternategrey"
|-
! align="left" style="width: 3em;" | Bit
! align="left" | bitfield
|-
| 0
|
|-
| 1
|
|-
| 2
|
|-
| 3
|
|-
| 4
|
|-
| 5
|
|-
| 6
|
|-
| 7
|
|}
 
|-
| 0x22
| 1
| Temperature alarm high (value + 40) * 2
 
|-
| 0x23
| 1
| Temperature alarm low (value + 40) * 2
 
|-
| 0x24
| 4
| unknown (00 00 00 3f, 00 00 80 3f)
 
|-
| 0x28
| 4
| unknown (00 00 20 c2)
 
|-
| 0x2c
| 2
| unknown (00 00)
 
|-
| 0x2e
| 2
| Device-specific configuration
 
|-
| 0x30
| 4
| Firmware version (ASCII, not NULL-terminated
 
|-
| 0x34
| 2
| Unsigned little-endian, serial number
 
|-
| 0x36
| 2
| unknown (00 00)
 
|-
| 0x38
| 1
| Relative humidity alarm high (value * 2)
 
|-
| 0x39
| 1
| Relative humidity alarm low (value * 2)
 
|-
| 0x3a
| 2
| unknown (00 00)
 
|-
| 0x3c
| 4
| unknown (00 00 00 00)
 
|}
 
=== Device-specific information ===
The 16-bit field at offset 0x2e in the configuration structure has different meanings depending on the type of device. It can be either a 16-bit little-endian integer or a bitfield.
 
{| border="0" width="100%" style="margin: 0; font-size: smaller" class="alternategrey sigroktable"
|-
! colspan="2" align="left" | EL-USB-2
|-
{| border="0" width="100%" style="margin: 0;" class="alternategrey"
|-
| style="width: 5em;" | 0x0000
| Celcius
|-
| 0x0001
| Fahrenheit
|}
 
|}


== Resources ==
== Resources ==

Revision as of 17:14, 27 November 2012

EL-USB-2

The Lascar Electronics EL-USB-2 is a USB-based temperature and humidity logger.

See Lascar Electronics EL-USB-2/Info for more details (such as lsusb -vvv output) about the device.

Hardware

Photos

Protocol

See Lascar Electronics EL-USB protocol

Resources

TODO.