Line 145: Line 145:
| 8
| 8
|
|
|  
| '''Temp/aux mode'''. When on and rotary switch is in position 5 (capacitance), CONF? returns <code>MV,1,DC</code>, but the value returned from FETC? instead denotes the temperature, in the unit (F or C) according to the meter's configuration.
 


|-
|-

Revision as of 10:21, 1 September 2012

Agilent U1232A

The Agilent U123xA are a series of low-cost 6000-count handheld digital multimeters. PC connectivity is via an IR port.

Hardware

Photos

Multimeter:

IR-USB cable:

See Multimeter cables#Agilent_U1173A.

IR-bluetooth interface:

See Multimeter cables#Agilent_U1177A.

Protocol

STAT

The STAT? command is used to query the current state of the device. The response is a quoted 21-byte string, like this:

"000000000110L00000000"
Byte Value Meaning
1 Max/min mode
0 Off
1 On
2
0
3 Trig hold-log
0 Off
1 On
4 Auto hold-log
0 Off
1 On
5 LED flashlight
0 Off
1 On
6 LCD backlight
0 Off
1 On
7
0 Off
1 On
8 Temp/aux mode. When on and rotary switch is in position 5 (capacitance), CONF? returns MV,1,DC, but the value returned from FETC? instead denotes the temperature, in the unit (F or C) according to the meter's configuration.


0 Off
1 On
9
0 Off
1 On
10
0 Off
1 On
11 Auto power-off (APO)
0 Off
1 On
12
0 Off
1 On
13
L
0 Off
1 On
14
0 Off
1 On
15
0 Off
1 On
16 Rotary switch position
0 V/Zlow
1 Voltage AC
2 Voltage DC
3 Resistance
4 Diode
5 Capacitance
6 Current
7 Microcurrent
17
0 Off
1 On
18
0 Off
1 On
19
0 Off
1 On
20
0 Off
1 On
21
0 Off
1 On