Difference between revisions of "UNI-T UT612"

From sigrok
Jump to navigation Jump to search
m (Uwe Hermann moved page UNI-T ut612 to UNI-T UT612 without leaving a redirect)
(No difference)

Revision as of 20:24, 13 July 2016

The UNI-T ut612 LCR Meter Dual Display comes with an USB interface that encapsulates standard serial over USB/HID. The main chipset inside the device responsible for the meassurements is the Cyrustek ES51919, the serial output from this chip is sent to a Silicon Labs CP2110 HID USB to UART Bridge.

Protocol ES51919

The protocol for the es51919 was reverse engineered as part of the DER-EE 5000 driver that uses the same chipset. Below is the excerpt from the source code.

Public official documentation does not contain the protocol description, so this is all based on reverse engineering.

Packet structure (17 bytes):

0x00: header1 ?? (0x00)

0x01: header2 ?? (0x0d)

0x02: flags

       bit 0 = hold enabled
       bit 1 = reference shown (in delta mode)
       bit 2 = delta mode
       bit 3 = calibration mode
       bit 4 = sorting mode
       bit 5 = LCR mode
       bit 6 = auto mode
       bit 7 = parallel measurement (vs. serial)

0x03: config

       bit 0-4 = ??? (0x10)
       bit 5-7 = test frequency
                   0 = 100 Hz
                   1 = 120 Hz
                   2 = 1 kHz
                   3 = 10 kHz
                   4 = 100 kHz
                   5 = 0 Hz (DC)

0x04: tolerance (sorting mode)

       0 = not set
       3 = +-0.25%
       4 = +-0.5%
       5 = +-1%
       6 = +-2%
       7 = +-5%
       8 = +-10%
       9 = +-20%
      10 = -20+80%

0x05-0x09: primary measurement

 0x05: measured quantity
         1 = inductance
         2 = capacitance
         3 = resistance
         4 = DC resistance
 0x06: measurement MSB  (0x4e20 = 20000 = outside limits)
 0x07: measurement LSB
 0x08: measurement info
         bit 0-2 = decimal point multiplier (10^-val)
         bit 3-7 = unit
                     0 = no unit
                     1 = Ohm
                     2 = kOhm
                     3 = MOhm
                     5 = uH
                     6 = mH
                     7 = H
                     8 = kH
                     9 = pF
                     10 = nF
                     11 = uF
                     12 = mF
                     13 = %
                     14 = degree

0x09: measurement status

         bit 0-3 = status
                     0 = normal (measurement shown)
                     1 = blank (nothing shown)
                     2 = lines ("----")
                     3 = outside limits ("OL")
                     7 = pass ("PASS")
                     8 = fail ("FAIL")
                     9 = open ("OPEn")
                    10 = shorted ("Srt")
         bit 4-6 = ??? (maybe part of same field with 0-3)
         bit 7   = ??? (some independent flag)

0x0a-0x0e: secondary measurement

 0x0a: measured quantity
         0 = none
         1 = dissipation factor
         2 = quality factor
         3 = parallel AC resistance / ESR
         4 = phase angle
 0x0b-0x0e: like primary measurement

0x0f: footer1 (0x0d)

0x10: footer2 (0x0a)

Protocol cp2110

CP2110 Data sheet

Images