Difference between revisions of "Voltcraft DL-161S"

From sigrok
Jump to navigation Jump to search
Line 57: Line 57:
| style="width: 1em;" |
| style="width: 1em;" |
| style="width: 3em;" | 3
| style="width: 3em;" | 3
| ?
| "REC" LED flash interval, in seconds: 0=none, 30=30s, 20=20s, 24=10s (!)
|-
|-
| style="width: 1em;" |
| style="width: 1em;" |
| style="width: 3em; vertical-align: top;" | 4
| Bitfield representing the following flags:
<div style="padding: 0;">
{| border="0" width="100%" style="margin: 0; border-collapse: collapse;" class="alternategrey"
|-
| style="width: 3em;" | 7
| Activation: 0=immediate, 1=manual
|-
| style="width: 3em;" | 6-5
| <i>Unused</i>
|-
| style="width: 3em;" | 4
| style="width: 3em;" | 4
| ?
| 0=store measurements, 1=real-time
|-
| style="width: 3em;" | 3
| Frequency weighting: 0=dBC, 1=dBA
|-
| style="width: 3em;" | 2
| Time weighting: 0=Slow, 1=Fast
|-
| style="width: 3em;" | 1-0
| <i>Unused</i>
|}
</div>
 
|-
|-
| style="width: 1em;" |
| style="width: 1em;" |

Revision as of 11:06, 25 July 2013

Voltcraft DL-161S
Voltcraft DL-161S.png
Status planned
Connectivity USB
Frequency range 31.5Hz - 8kHz
Measurement range (A) 30dB - 130dB
Accuracy (94dB@1kHz) 1.4dB
Frequency weighting A, C
Time weighting F, S
Standards IEC 61672-1 Class 2
Website Voltcraft

The Voltcraft DL-161S is a sound level meter with USB connectivity.

See Voltcraft DL-161S/Info for more details (such as lsusb -vvv output) about the device.

This is a rebadged CEM DT-173.

Hardware

TODO

Photos

Protocol

The host sends commands and receives responses via a virtual serial port, with USB bulk endpoint 2 as a backend. A command consists of a single byte followed by any arguments to the command.

Command Response Description
0x0e 0xff Configure device
1-2 Length of data to follow, as a little-endian integer.
3 "REC" LED flash interval, in seconds: 0=none, 30=30s, 20=20s, 24=10s (!)
4 Bitfield representing the following flags:
7 Activation: 0=immediate, 1=manual
6-5 Unused
4 0=store measurements, 1=real-time
3 Frequency weighting: 0=dBC, 1=dBA
2 Time weighting: 0=Slow, 1=Fast
1-0 Unused
5 Sample interval 1-7, representing 50ms, 500ms, 1s, 2s, 5s, 10s, 60s respectively.
6 Last two digits of year, e.g. 0x0d for 2013
7 Month, 1-12
8 Day of month, 1-31
9 Hours
10 Minutes
11 Seconds
12 ?
13 ?
14 Number of samples to acquire.
15 Alarm low threshold
16 Alarm high threshold

Resources