Difference between revisions of "HT Instruments HT410"

From sigrok
Jump to navigation Jump to search
m
 
(One intermediate revision by the same user not shown)
Line 3: Line 3:
| name            = HT Instruments HT410
| name            = HT Instruments HT410
| status          = supported
| status          = supported
| source_code_dir = gmx_mh_1x_2x
| source_code_dir = gmc-mh-1x-2x
| counts          = 3100
| counts          = 3100
| categories      = CAT III 1000 V, CAT IV 600 V
| categories      = CAT III 1000 V, CAT IV 600 V
Line 325: Line 325:
|   
|   
|  
|  
| style="background: #ff0000; "| mqflags += HOLD
| style="background: #ff0000; "| <s>mqflags += HOLD </s>
| Disabled while IR interface is active.
| Disabled while IR interface is active.
   
   
Line 333: Line 333:
|   
|   
|  
|  
| style="background: #ff0000; "| mqflags += MIN
| style="background: #ff0000; "| <s>mqflags += MIN </s>
| Disabled while IR interface is active.
| Disabled while IR interface is active.
   
   
Line 341: Line 341:
|   
|   
|  
|  
| style="background: #ff0000; "| mqflags += MAX  
| style="background: #ff0000; "| <s>mqflags += MAX</s>
| Disabled while IR interface is active.
| Disabled while IR interface is active.



Latest revision as of 22:34, 23 August 2015

HT Instruments HT410
HT410 logo.png
Status supported
Source code gmc-mh-1x-2x
Counts 3100
IEC 61010-1 CAT III 1000 V, CAT IV 600 V
Connectivity Infrared (RS232)
Measurements voltage, current, resistance, temperature, diode, continuity
Features autorange, data hold, bargraph
Website ht-instruments.com

The HT Instruments HT410 is a 3100 counts handheld digital multimeter with IR (RS232) connectivity.

It is probably a low-cost OEM version of the Rishabh 15S (see Gossen Metrawatt, Metrahit 15S) or a similar model with much worse specs, although the actual accuracy of Matthias' device seems to be better than the specs, according to the calibration certificate (!) that came with it. At least the test cords that come with it are Rishabh brand and cheaper quality than orginal Gossen Metrawatt ones, although they look the same. The same seems to be true for case and sleeve.

The IR connectivity is undocumented, possibly because no IR interface or software is available from the manufacturer. To enable sending data, press the DATA button while powering it on. See here for fitting interfaces.

Hardware

  • Battery: 9 V NEDA1604, JIS006P, IEC6F22
  • Fuse:
    • Current 300 mA: FF 1,6 A/1000V, 6,3x32 mm, 10 kA
    • Current 3A, 10A: FF 16 A/1000V, 10x38mm, 30 kA
  • TODO.

Photos

Measurement functions and ranges

Measurement function Range Resolution Accuracy sigrok result Remarks
V= 30.00 mV 10 μV ±(0,5% + 7) mq V, unit V, mqflags DC Manual range; accuracy with ZERO, otherwise +35 digits.
300.0 mV 100 μV " Manual range
3.000 V 1 mV ±(0,25% + 5) "
30.00 V 10 mV "
300.0 V 100 mV "
1000 V 1 V ±(0,35% + 5) "
V≈ 3.000 V 1 mV ±(0,75% + 6) mq V, unit V, mqflags AC
30.00 V 10 mV "
300.0 V 100 mV "
1000 V 1 V "
A= 300 µA 0.1 µA ±(1.0% + 9) mq Curr., unit A, mqflags DC Manual range
3.000 mA 1 µA ±(1.0% + 6) "
30.00 mA 10 µA ±(1.0% + 9) " Manual range
300.0 mA 100 µA ±(1,0% + 6) "
3.000 A 1 mA ±(1,0% + 9) " Manual range
10.00 A 10 mA ±(1,0% + 6) "
A≈ 3.00 mA 1 µA ±(1,5% + 6) mq Curr., unit A, mqflags AC
300.0 mA 100 µA "
10.00 A 10 mA "
Ω 30.00 Ω 10 mΩ ±(0.5% + 7) mq Res., unit Ω, mqflags - Accuracy with ZERO, otherwise +30 digits.
300.0 Ω 100 mΩ "
3.000 kΩ 1 Ω ±(0.4% + 5) "
30.00 kΩ 10 Ω "
300.0 kΩ 100 Ω "
3.000 MΩ 1 kΩ ±(0.6% + 5) "
30.00 MΩ 10 kΩ ±(2.0% + 5) "
Diode/Cont. 3.000 V– 1 mV mq V, unit V, mqflags DC,DIODE Diode and cont. are identical on this device, just without beeper for diode. Cont. beeps < 120 Ω.
Cap. F 30.00 nF 10 pF ±(1.0% + 7) mq Cap., unit F, mqflags - Accuracy with ZERO, otherwise +50 digits.
300.0 nF 100 pF "
3.000 µF 1 nF "
30.00 µF 10 nF ±(3.0% + 7) "
Hz≃ 300.0 Hz 0.1 Hz ±(0.5% + 5) mq Freq., unit Hz, mqflags AC,DC
3.000 kHz 1 Hz "
30.00 kHz 10 Hz "
100.00 kHz 100 Hz "
Duty cycle 2% - 98% 0.1 % ±5 mq Duty cycle, unit %, mqflags -
°C Pt100 –200.0 - +200.0 °C 0.1 °C ±(2 K + 9) mq Temp., unit °C, mqflags -
Pt100 +200.0 - +850.0 °C ±(1.0% + 9) "
Pt1000 –100.0 - +200.0 °C ±(2 K + 6) "
Pt1000 +200.0 - +850.0 °C ±(1.0% + 6) "
Data Hold mqflags += HOLD Disabled while IR interface is active.
MIN mqflags += MIN Disabled while IR interface is active.
MAX mqflags += MAX Disabled while IR interface is active.

The column "sigrok result" contains in short form what the driver generates for the resp. data type.

The accuracy specification is incomplete, see manual for exact specs.