Difference between revisions of "UNI-T UT61E"

From sigrok
Jump to navigation Jump to search
(add Usage section)
(mention that UT61+/UT161 is NOT supported)
 
(8 intermediate revisions by 4 users not shown)
Line 4: Line 4:
| status          = supported
| status          = supported
| source_code_dir =  
| source_code_dir =  
| counts          = 22000
| counts          = 22,000
| categories      = CAT II (600V) / CAT III (300V)
| categories      = CAT II (600 V) / CAT III (300 V)
| connectivity    = [[Device_cables#UNI-T_UT-D02|RS232]] / [[Device_cables#UNI-T_UT-D04|USB]]
| connectivity    = [[Device_cables#UNI-T_UT-D02|RS232]] / [[Device_cables#UNI-T_UT-D04|USB]]
| measurements    = voltage, current, resistance, capacitance, frequency, duty cycle, diode, continuity
| measurements    = voltage, current, resistance, capacitance, frequency, duty cycle, diode, continuity
Line 12: Line 12:
}}
}}


The '''UNI-T UT61E''' is a 22000 counts, CAT II (600V) / CAT III (300V) handheld digital multimeter with RS232 or USB connectivity.
The '''UNI-T UT61E''' is a 22,000 counts, CAT II (600 V) / CAT III (300 V) handheld digital multimeter with RS-232 or USB connectivity.


See [[UNI-T UT61E/Info]] for more details (such as '''lsusb -vvv''' output) about the device.
See [[UNI-T UT61E/Info]] for more details (such as '''lsusb -vvv''' output) about the device.
The more recent [https://www.uni-trend.com/meters/html/product/NewProducts/UT61%20161%20Series/ UT61+/UT161 series] is different from the previous UT61 series, and is '''not''' supported, until their protocol is known and a driver for them becomes available.


== Hardware ==
== Hardware ==
Line 54: Line 56:
== Protocol ==
== Protocol ==


See [[Multimeter_ICs#Cyrustek ES51922]] for the DMM IC protocol.
See [[Multimeter_ICs#Cyrustek ES51922|Cyrustek ES51922]] for the DMM IC protocol.


Depending on the cable, additional decoding is needed, though.
Depending on the cable, additional decoding is needed, though.


TODO: Serial cable vs. the two USB HID based cables.
Different cables are available to communicate to the DMM: regular serial cables which provide a COM port, and USB HID based cables where applications are required to handle a proprietary protocol of running serial communication on top of HID requests. See the ''[[Device cables]]'' page for details; the same cables can be used with many different DMM models.
 
Depending on the specific cable in use, either a device driver ending in '''-ser''' or not ending in '''-ser''' must be used. See README.devices for details.
 
The message ''"HID feature report error: LIBUSB_ERROR_PIPE"'' results from using the USB HID driver with a USB-to-serial cable. In this case, try using --driver='''uni-t-ut61e-ser'''


The transmission of the measurement data cannot be disabled, the respective Cyrustek ES51922 pin (111, '''RS232''') is tied to GND (i.e. transmission is always enabled) on this multimeter.<sup>[http://www.steffenvogel.de/2011/01/25/inner-workings-of-uni-trend-ut61e-digital-multimeter/]</sup>
The transmission of the measurement data cannot be disabled. The respective Cyrustek ES51922 pin (111, '''RS232''') is tied to GND (i.e. transmission is always enabled) on this multimeter.<sup>[http://www.steffenvogel.de/2011/01/25/inner-workings-of-uni-trend-ut61e-digital-multimeter/]</sup>


== Usage ==
== Usage ==
The following [[sigrok-cli]] command can be used to receive measured values from a device connected via USB (note that the USB VID/PID after the '''<tt>conn</tt>''' option needs to be changed depending on the exact USB adapter cable used):
The following [[sigrok-cli]] command can be used to receive five measured values from a device connected via USB (note that the USB VID/PID after the '''<tt>conn</tt>''' option needs to be changed depending on the exact USB adapter cable used):
<source lang="bash">
$ sigrok-cli --driver=uni-t-ut61e:conn=1a86.e008 -O analog --samples 5
</source>
 
If your meter has a serial (RS-232) cable connected to a USB-to-serial adapter, a different driver is used. Example for ttyUSB0:
<source lang="bash">
<source lang="bash">
$ sigrok-cli --driver=uni-t-ut61e:conn=1a86.e008 -O analog
$ sigrok-cli --driver=uni-t-ut61e-ser:conn=/dev/ttyUSB0 -O analog --samples 5
</source>
Same example for COM1 (Windows), please note the special syntax for specifying the COM port:
<source lang="text">
C:\> sigrok-cli --driver=uni-t-ut61e-ser:conn=\\.\COM1 -O analog --samples 5
</source>
</source>
With the '''<tt>--samples <n></tt>''' option it is posible to stop the acquisition after the specified number of measurements. More information can be found in the [http://sigrok.org/gitweb/?p=libsigrok.git;a=blob;f=README.devices README.devices] file of the [[libsigrok]] source tree.
 
'''<tt>--samples <n></tt>''' stops acquisition after the specified number of measurements, while '''<tt>--continuous</tt>''' does not stop.  Type just '''sigrok-cli''' by itself for a summary of options.  
More information on drivers can be found in the [http://sigrok.org/gitweb/?p=libsigrok.git;a=blob;f=README.devices README.devices] file of the [[libsigrok]] source tree.


== Resources ==
== Resources ==

Latest revision as of 18:38, 19 March 2021

UNI-T UT61E
Old ver front.png
Status supported
Counts 22,000
IEC 61010-1 CAT II (600 V) / CAT III (300 V)
Connectivity RS232 / USB
Measurements voltage, current, resistance, capacitance, frequency, duty cycle, diode, continuity
Features autorange, true-rms, data hold, min/max, relative, bargraph, backlight
Website uni-trend.com

The UNI-T UT61E is a 22,000 counts, CAT II (600 V) / CAT III (300 V) handheld digital multimeter with RS-232 or USB connectivity.

See UNI-T UT61E/Info for more details (such as lsusb -vvv output) about the device.

The more recent UT61+/UT161 series is different from the previous UT61 series, and is not supported, until their protocol is known and a driver for them becomes available.

Hardware

Photos

Older version:

Newer version:

Protocol

See Cyrustek ES51922 for the DMM IC protocol.

Depending on the cable, additional decoding is needed, though.

Different cables are available to communicate to the DMM: regular serial cables which provide a COM port, and USB HID based cables where applications are required to handle a proprietary protocol of running serial communication on top of HID requests. See the Device cables page for details; the same cables can be used with many different DMM models.

Depending on the specific cable in use, either a device driver ending in -ser or not ending in -ser must be used. See README.devices for details.

The message "HID feature report error: LIBUSB_ERROR_PIPE" results from using the USB HID driver with a USB-to-serial cable. In this case, try using --driver=uni-t-ut61e-ser

The transmission of the measurement data cannot be disabled. The respective Cyrustek ES51922 pin (111, RS232) is tied to GND (i.e. transmission is always enabled) on this multimeter.[1]

Usage

The following sigrok-cli command can be used to receive five measured values from a device connected via USB (note that the USB VID/PID after the conn option needs to be changed depending on the exact USB adapter cable used):

$ sigrok-cli --driver=uni-t-ut61e:conn=1a86.e008 -O analog --samples 5

If your meter has a serial (RS-232) cable connected to a USB-to-serial adapter, a different driver is used. Example for ttyUSB0:

$ sigrok-cli --driver=uni-t-ut61e-ser:conn=/dev/ttyUSB0 -O analog --samples 5

Same example for COM1 (Windows), please note the special syntax for specifying the COM port:

C:\> sigrok-cli --driver=uni-t-ut61e-ser:conn=\\.\COM1 -O analog --samples 5

--samples <n> stops acquisition after the specified number of measurements, while --continuous does not stop. Type just sigrok-cli by itself for a summary of options. More information on drivers can be found in the README.devices file of the libsigrok source tree.

Resources