Difference between revisions of "Gossen Metrawatt Metrahit 16I"

From sigrok
Jump to navigation Jump to search
m
Line 39: Line 39:
-->
-->
== Measurement functions and ranges ==
== Measurement functions and ranges ==
The following table is a work in progress and currently inaccurate!


{|border="0" style="font-size: smaller" class="alternategrey sortable sigroktable"
{|border="0" style="font-size: smaller" class="alternategrey sortable sigroktable"
Line 89: Line 87:
| 3 V
| 3 V
| 100 μV
| 100 μV
| rowspan="4" | ±(0,5% + 30) (> 300 Digit)
| rowspan="4" | ±(1.0% + 3) (> 10 Digit)
|  
|  


Line 111: Line 109:
| 3 V
| 3 V
| 100 μV
| 100 μV
| rowspan="4" | ±(0,5% + 30) (> 300 Digit)
| rowspan="4" | ±(1.0% + 3) (> 10 Digit)
|  
|  


Line 133: Line 131:
| 30 A
| 30 A
| 10 mA
| 10 mA
|  
| rowspan="2" | ±(2,5% + 3) (>10 Digit)
| rowspan="2" | Current clamp with 100 mV/A required, e.g. GMC WZ12B
| rowspan="2" | Current clamp with 100 mV/A required, e.g. GMC WZ12B


Line 139: Line 137:
| 100 A
| 100 A
| 100 mA
| 100 mA
|


|-
|-
Line 145: Line 142:
| 30.00 Ω
| 30.00 Ω
| 10 mΩ
| 10 mΩ
|  
| rowspan="2" | ±(0,5% + 3)
|
|


Line 151: Line 148:
| 300.0 Ω
| 300.0 Ω
| 100 mΩ
| 100 mΩ
|
|
|


Line 157: Line 153:
| 3.000 kΩ
| 3.000 kΩ
| 1 Ω
| 1 Ω
|  
| rowspan="3" | ±(0,4% + 1)
|  
|  


Line 163: Line 159:
| 30.00 kΩ
| 30.00 kΩ
| 10 Ω
| 10 Ω
|
|  
|  


Line 169: Line 164:
| 300.0 kΩ
| 300.0 kΩ
| 100 Ω
| 100 Ω
|
|  
|  


Line 175: Line 169:
| 3.000 MΩ
| 3.000 MΩ
| 1 kΩ
| 1 kΩ
|  
| ±(0,6% + 1)
|  
|  


Line 181: Line 175:
| 30.00 MΩ
| 30.00 MΩ
| 10 kΩ
| 10 kΩ
|  
| ±(2,0% + 1)
|  
|  


Line 188: Line 182:
| 3.0000 V–
| 3.0000 V–
| 1 mV
| 1 mV
|  
| ±(0,25% + 1)
|
|


Line 195: Line 189:
| 30.00 nF
| 30.00 nF
| 10 pF
| 10 pF
|  
| rowspan="3" | ±(1,0% + 3)
|  
|  


|-
|-
| 300.0 nF
| 300.0 nF
| 100 pF
| 100 pF  
|
|  
|  


Line 207: Line 200:
| 3.000 µF
| 3.000 µF
| 1 nF
| 1 nF
|
|  
|  


Line 213: Line 205:
| 30.00 µF
| 30.00 µF
| 10 nF
| 10 nF
|  
| ±(3,0% + 3)
|  
|  


Line 241: Line 233:
| rowspan="4" | °C
| rowspan="4" | °C
| rowspan="2" | Pt100
| rowspan="2" | Pt100
| –200.0 - +100.0 °C
| –200.0 - +200.0 °C
| rowspan="4" | 0.1 °C
| 2 Kelvin + 5
| rowspan="4" | 0,5 °C + 3
|  


|-
|-
| +100.0 - +850.0 °C
| +200.0 - +800.0 °C
| ±(1.0% + 5)
|


|-
|-
| rowspan="2" | Pt1000
| rowspan="2" | Pt1000
| –100.0 - +100.0 °C
| –100.0 - +200.0 °C
| 2 Kelvin + 5
|


|-
|-
| +100.0 - +850.0 °C
| +200.0 - +800.0 °C
| ±(1.0% + 5)
|


|-
|-

Revision as of 01:18, 2 November 2013

Gossen Metrawatt Metrahit 16I
Status in progress
Counts 3100 (31000)
IEC 61010-1 CAT III 1000 V, CAT IV 600 V
Connectivity Infrared (RS232)
Measurements voltage, current, resistance, capacitance, temperature, diode, continuity
Features autorange, data hold, bargraph
Website [1]

The Gossen Metrawatt Metrahit 16I is a 3100 counts handheld digital multimeter with IR (RS232) connectivity and insulation measurement (500/1000 V).

The resolution is one digit higher on the data interface.

Hardware

  • Fuse: None!
  • TODO.

Photos

Measurement functions and ranges

Measurement function Range Resolution Error Remarks
V= 30 mV 10 μV ±(0,5% + 3) Not within autorange, use manual range
300 mV 100 μV ±(0,5% + 3) Not within autorange, use manual range
3 V 1 mV ±(0,25% + 1)
30 V 10 mV
300 V 100 mV
600 V 1 V ±(0,35% + 1)
V≈ 3 V 100 μV ±(1.0% + 3) (> 10 Digit)
30 V 100 μV
300 V 1 mV
1000 V 10 mV
V≃ 3 V 100 μV ±(1.0% + 3) (> 10 Digit)
30 V 100 μV
300 V 1 mV
1000 V 10 mV
A≈ (Clamp only)  30 A 10 mA ±(2,5% + 3) (>10 Digit) Current clamp with 100 mV/A required, e.g. GMC WZ12B
100 A 100 mA
Ω 30.00 Ω 10 mΩ ±(0,5% + 3)
300.0 Ω 100 mΩ
3.000 kΩ 1 Ω ±(0,4% + 1)
30.00 kΩ 10 Ω
300.0 kΩ 100 Ω
3.000 MΩ 1 kΩ ±(0,6% + 1)
30.00 MΩ 10 kΩ ±(2,0% + 1)
Diode/Cont. 3.0000 V– 1 mV ±(0,25% + 1)
Cap. F 30.00 nF 10 pF ±(1,0% + 3)
300.0 nF 100 pF
3.000 µF 1 nF
30.00 µF 10 nF ±(3,0% + 3)
Hz 300.00 Hz 0.01 Hz ±(0.1% + 3)
3.0000 kHz 0.1 Hz
30.000 kHz 1 Hz
100.000 kHz 10 Hz
°C Pt100 –200.0 - +200.0 °C 2 Kelvin + 5
+200.0 - +800.0 °C ±(1.0% + 5)
Pt1000 –100.0 - +200.0 °C 2 Kelvin + 5
+200.0 - +800.0 °C ±(1.0% + 5)
Event count
Stop Watch