Voltcraft 4080

From sigrok
Revision as of 09:05, 17 May 2020 by Gsi (talk | contribs) (reference Extech 380193)
(diff) ← Older revision | Latest revision (diff) | Newer revision → (diff)
Jump to navigation Jump to search
Voltcraft 4080
Voltcraft4080 2.png
Status supported
Source code serial-lcr
Counts 20000
IEC 61010-1
Connectivity USB, RS232
Measurements resistance, capacitance, inductance
Features autorange, relative, auto-poweroff, min-max, tolerance
Website conrad.com

The Voltcraft 4080 is an LCR meter with USB connectivity. It is described as 4.5 digits (20000 count) LCR meter with 0.5% basic accuracy (resistance) that can measure at 120Hz and 1kHz, and comes with USB connectivity (serial protocol).

The Peaktech 2165 appears to be another rebrand. Extech 380193 looks like another compatible.

Hardware

LCR meter:

  • TI TLC7135C ADC, 4 1/2 digits, 1Ksa/s
  • TI MSP430 MCU
  • several discrete TI components

USB to IR cable:

  • FT232R, regular serial port, bidirectional (RX and TX LED)

Photos

Protocol

Serial communication runs at 1200/7e1, uses ASCII characters and is basically human readable. You can access the device with e.g. a terminal program like CuteCom.

Serial packets consist of 39 ASCII characters that end in the CR-LF termination, contain a lot of single character flags, as well as multiple multi-character fields for the primary and secondary displays.

The description of the commands you find at Conrad. The PeakTech 2165 user manual has another description, represented differently.

When the device does not respond it might be in setup mode. You can see that in the display. But the software can not. So it might be a good idea to do an initialization with [BXXXXXX] to stop the setup mode. This only works for a setup mode that was requested by an earlier remote session. You can not leave a manual setup mode this way.

Some findings not mentioned in the protocol definition above; without waranty:

Disabling keys

Sending

A

responds with

REMOTE READY..

and says RS232 in the display.

Keys on device are disabled.

Enabling keys

Sending

B

responds with

REMOTE STOP...

Keys on device are enabled.

Resources