MASTECH M9803R
Status | planned |
---|---|
Counts | 4000 |
IEC 61010-1 | CAT II (1000V) |
Connectivity | RS232 |
Measurements | voltage, resistance, diode, capacitance, frequency, current, hFE |
Features | autorange, true-rms, data hold, delay hold, min/max, relative, bargraph, backlight |
Website | p-mastech.com |
The MASTECH M9803R is a 4000 counts, CAT II (1000V) digital bench multimeter with RS232 connectivity.
Hardware
TODO.
Photos
Multimeter:
TODO.
RS232 cable:
TODO.
Protocol
The protocol is documented in: cdmm: M9803R Class Reference
Detailed Description M9803R 11 byte binary protocoll. Reverse engeneered by Matthias Toussaint Port settings seem to be 9600 7E1 or 9600 7O1 (looks like a bug) Protocoll is 11 bytes binary fixed length END OF PACKET is 0x0d 0x0a Multimeter continuously sends data Byte 0: Sign ored together 0x00 -> positive 0x08 -> negative 0x01 -> overflow Byte 1: Digit 4 binary Byte 2: Digit 3 binary Byte 3: Digit 2 binary Byte 4: Digit 1 binary Byte 5: Mode 0x00 DCV 0x01 ACV 0x02 DCA 0x03 ACA 0x04 Ohms 0x05 Ohms + Beep 0x06 Diode V 0x07 ADP 0x08 DCA (10A) 0x09 ACA (10A) 0x0A Freq 0x0C Capacity Byte 6:Decimal point position Display Unit Value Frequency 0.000 kHz 0x00 00.00 kHz 0x01 00.00 Hz 0x05 000.0 Hz 0x06 Voltage 000.0 mV 0x00 0.000 V 0x01 00.00 V 0x02 000.0 V 0x03 0000. V 0x04 Current 0.000 mA 0x00 00.00 mA 0x01 000.0 mA 0x02 Capacity 0.000 nF 0x00 00.00 nF 0x01 000.0 nF 0x02 0.000 uF 0x03 00.00 uF 0x04 Resistance 000.0 Ohm 0x00 0.000 kOhm 0x01 00.00 kOhm 0x02 000.0 kOhm 0x03 0000. kOhm 0x04 00.00 MOhm 0x05 Byte 7: Hold/min/Max/Rel ored together 0x01 Hold 0x02 Rel 0x04 Min 0x08 Max Byte 8: Auto/Manu 0x01 APOF (AutoPowerOff) 0x02 Manu 0x04 Auto 0x08 MEM