MASTECH_MS6514
|
Status |
in progress |
---|
Source code |
mastech_ms6514 |
---|
Connectivity |
USB/serial |
---|
Measurements |
temperature |
---|
Features |
data hold, min/max/avg |
---|
Website |
mastech-group.com |
The MASTECH MS6514 is a logging thermometer with two thermocouple inputs and USB connectivity. It supports K,J,T,E,R,S,N thermocouple types.
The device sends temperature measurements at appr. 2 Hz to the host.
Hardware
- TODO
- SiLabs CP2102 USB interface chip.
Photos
TODO
Protocol
The device communicates with the host via a SiLabs CP2102 USB interface chip. Communication parameters are 9600_8_N_1. Live data is transmitted as soon as the button Setup/PC-Link is pressed for 3 seconds.
Commands
Byte
|
Description
|
0xA1
|
Start sending stored data
|
Received data
Byte
|
Bit
|
Description
|
0 |
|
Always 0x65
|
1 |
|
Always 0x14
|
2 |
0 |
1=READ of stored values, 0=LIVE data
|
4..3 |
|
Index of stored value (0..999) oldest is 0
|
6..5 |
|
Temperature main display (absolute value without sign)
|
8..7 |
|
Temperature aux display (absolute value without sign)
|
9 |
5..4 |
Mode: 10=SETUP, 11=READ
|
2..0 |
Thermocouple type (1=K, 2=J, 3=T, 4=E, 5=R, 6=S, 7=N)
|
10 |
6 |
Hold (1=HOLD, 0=not HOLD)
|
5 |
Record active (1=REC, 0=not REC)
|
1..0 |
Unit (1=°C, 2=°F, 3=K)
|
11 |
7 |
Sign temperatue main display (1=negative, 0=positive)
|
6 |
Overload temperature main display (1="OL" on LCD)
|
3 |
Factor temperature main display (1 means divide by 10)
|
1..0 |
Function main and aux display (00=T1 on main, T2 on aux; 01=T2 on main, T1 on aux; 10=T1-T2 on main, T1 on aux; 11=T1-T2 on main, T2 on aux)
|
12 |
7 |
Sign temperatue aux display (1=negative, 0=positive)
|
6 |
Overload temperature aux display (1="OL" on LCD)
|
3 |
Factor temperature aux display (1 means divide by 10)
|
1..0 |
Function aux display (0=no active, 1=MAX, 2=MIN, 3=AVG). If >0 function defined in byte 11, bit 0-1 is overriden.
|
13 |
|
Hours (hh)
|
14 |
|
Minutes (mm)
|
15 |
|
Seconds (ss)
|
16 |
|
Always 0x0D
|
17 |
|
Always 0x0A
|