Gossen Metrawatt Metrahit 16I

From sigrok
Revision as of 16:39, 7 May 2015 by Matthias Heidbrink (talk | contribs) (Fixed source code path)
Jump to navigation Jump to search
Gossen Metrawatt Metrahit 16I
Gossen Metrawatt Metrahit 16I small.png
Status supported
Source code gmc-mh-1x-2x
Counts 3100 (31000)
IEC 61010-1 CAT III 1000 V, CAT IV 600 V
Connectivity Infrared (RS232)
Measurements voltage, current, resistance, capacitance, temperature, diode, continuity
Features autorange, data hold, bargraph
Website [1]

The Gossen Metrawatt Metrahit 16I is a 3100 counts handheld digital multimeter with IR (RS232) connectivity and insulation measurement (500/1000 V).

The resolution is one digit higher on the data interface.

Hardware

  • Battery: 9 V IEC 6 LR 61
  • Fuse: None! (not required, no internal amps ranges)
  • TODO.

Photos

Measurement functions and ranges

Measurement function Range Resolution Deviation sigrok result Remarks
V= 30 mV 10 μV ±(0,5% + 3) mq V, unit V, mqflags DC Input impendance >10 GΩ; not covered by autorange, use manual range
300 mV 100 μV ±(0,5% + 3) " "
3 V 1 mV ±(0,25% + 1) "
30 V 10 mV "
300 V 100 mV "
600 V 1 V ±(0,35% + 1) "
V≈ 3 V 100 μV ±(1.0% + 3) (> 10 Digit) mq V, unit V, mqflags AC,RMS
30 V 100 μV "
300 V 1 mV "
1000 V 10 mV "
V≃ 3 V 100 μV ±(1.0% + 3) (> 10 Digit) mq V, unit V, mqflags AC,DC,RMS
30 V 100 μV "
300 V 1 mV "
1000 V 10 mV "
A≈ (Clamp only)  30 A 10 mA ±(2,5% + 3) (>10 Digit) mq Curr., unit A, mqflags AC,DC,RMS Current clamp with 100 mV/A required, e.g. GMC WZ12B
100 A 100 mA "
Ω 30.00 Ω 10 mΩ ±(0,5% + 3) mq Res., unit Ω, mqflags -
300.0 Ω 100 mΩ "
3.000 kΩ 1 Ω ±(0,4% + 1) "
30.00 kΩ 10 Ω "
300.0 kΩ 100 Ω "
3.000 MΩ 1 kΩ ±(0,6% + 1) "
30.00 MΩ 10 kΩ ±(2,0% + 1) "
Ω Iso 500V 1.600 GΩ 1 kΩ - 1 MΩ ±(3% + 2) mq Res., unit Ω, mqflags - Altogether 4 ranges with different resolution
Ω Iso 1000V 3.100 GΩ 1 kΩ - 1 MΩ mq Res., unit Ω, mqflags - Altogether 4 ranges with different resolution
Diode/Cont. 3.000 V– 1 mV ±(0,25% + 1) mq V, unit V, mqflags DC,DIODE Diode and cont. are identical on this device, just without beeper for diode.
Cap. F 30.00 nF 10 pF ±(1,0% + 3) mq Cap., unit F, mqflags -
300.0 nF 100 pF "
3.000 µF 1 nF "
30.00 µF 10 nF ±(3,0% + 3) "
Hz= 300.0 Hz 0.01 Hz ±(0.1% + 3) mq Freq., unit Hz The protocol does not allow to differ AC and DC.
3.000 kHz 0.1 Hz "
30.00 kHz 1 Hz "
100.00 kHz 10 Hz "
Hz≈ 300.0 Hz 0.01 Hz ±(0.1% + 3) mq Freq., unit Hz The protocol does not allow to differ AC and DC.
3.000 kHz 0.1 Hz "
30.00 kHz 1 Hz "
100.00 kHz 10 Hz "
°C Pt100 –200.0 - +200.0 °C 0.1 °C 2 Kelvin + 5 mq Temp., unit °C, mqflags -
Pt100 +200.0 - +800.0 °C ±(1.0% + 5) "
Pt1000 –100.0 - +200.0 °C 2 Kelvin + 5 "
Pt1000 +200.0 - +800.0 °C ±(1.0% + 5) "
Event count Seems not to be present on data output.
Data Hold mqflags += HOLD Disabled while IR interface is active on  Matthias' device.
MIN mqflags += MIN Influences only display, via IR flag is set, but actual current value is delivered.
MAX mqflags += MAX Influences only display, via IR flag is set, but actual current value is delivered.

The column "sigrok result" contains in short form what the driver generates for the resp. data type.