Voltcraft 4080

From sigrok
Revision as of 17:12, 20 January 2014 by Nobs (talk | contribs)
Jump to navigation Jump to search

The Voltcraft Plus 4080 LCR Meter Dual Display comes with an optically coupled serial interface.

Protocol

You can access the device e.g. with cutecom. The protocol is basically human readable.

The description of the commands you find at Conrad.

When the device does not respond it might be in setup mode. You can see that in the display. But the software can not. So it might be a good idea to do an initialization with [BXXXXXX] to stop the setup mode.


Some findings not mentioned in the protocol definition above; without waranty:

disabling keys

sending

A

responds with

REMOTE READY..

and says RS232 in the display

Keys on device are disabled

enabling keys

sending

B

responds with

REMOTE STOP...

Keys on device are enabled

Images

Device complete with rubber bumper and optical/serial cable:

Voltcraft4080 complete.jpg

Screenshot of a short terminal session using CuteCom:

Voltcraft4080 cutecom.png