Siemens B1105
Status | supported |
---|---|
Source code | gmc-mh-1x-2x |
Counts | 310000 |
IEC 61010-1 |
CAT III 1000 V, CAT IV 600 V (older version) CAT III 600 V, CAT IV 300 V (newer version) |
Connectivity | Infrared (RS232/USB) |
Measurements | voltage, current, resistance, capacitance, temperature, diode, continuity |
Features | autorange, data hold |
Website | [1] |
The Siemens B1105 is a 310000 counts 1000V CAT III/600 V CAT IV (older version) or 600V CAT III/300 V CAT IV (newer version) handheld digital multimeter with IR (RS232/USB) connectivity.
It is an OEM version of the Gossen Metrawatt Metrahit 28S. The only differences seem to be the type, logo and color (light grey-blue print on knob, light grey-blue holster) and the voltage rating.
Hardware
- Power supply:
- Battery: 2x 1,5 V IEC R 6, IEC LR 6 or corresponding NiCd rechargeable batteries
- External power supply NA4/500 by Gossen Metrawatt, 4.2 V, 500 mA (for older version)
- External power supply NA5/600 by Gossen Metrawatt, 5 V, 600 mA (for newer version, 3-pin plug)
- Fuse:
- Current 300 mA: FF (UR) 1,6 A/1000 V AC/DC (10 kA) 6,3 mm x 32 mm
- Current 3A/10A: FF (UR) 10 A/1000 V AC/DC (30 kA) 10 mm x 38 mm
Photos
Measurement functions and ranges
Measurement function | Range | Resolution | Accuracy | sigrok result | Remarks |
---|---|---|---|---|---|
V= | 300 mV | 1 μV | ±(0.02% + 0.010% + 5) | mq V, unit V, mqflags DC | |
3 V | 10 μV | ±(0.02% + 0.005% + 5) | " | ||
30 V | 100 μV | " | |||
300 V | 1 mV | " | |||
1000 V | 10 mV | " | |||
V≈ | 300 mV | 10 µV | ±(0.5% + 30) (> 500 Digit)) | mq V, unit V, mqflags AC,RMS | |
3 V | 100 μV | ±(0.2% + 30) | " | ||
30 V | 100 μV | " | |||
300 V | 1 mV | " | |||
1000 V | 10 mV | " | |||
V≃ | 300 mV | 10 μV | ±(0.5% + 30) (> 500 Digit)) | mq V, unit V, mqflags AC,DC,RMS | |
3 V | 100 μV | ±(0.2% + 30) | " | ||
30 V | 100 μV | " | |||
300 V | 1 mV | " | |||
1000 V | 10 mV | " | |||
V db≈ | -48 - 63 dB | 0.01 dB | ± 0.1 dB | mq V, unit V, mqflags AC,RMS | Actually the 5 AC ranges, see there for accuracy. |
V db≈ rel | -48 - 63 dB | 0.01 dB | ± 0.1 dB | mq V, unit V, mqflags AC,RMS | Actually the 5 AC ranges, see there for accuracy. To activate, press yellow button in dB mode, then do second measurement. Cannot be distinguished from non-rel measurement remotely. |
A= | 300.00 µA | 1 nA | 0.05% + 0.02% + 5 | mq Curr., unit A, mqflags DC | |
3.0000 mA | 10 nA | 0.05% + 0.01% + 5 | " | ||
30.000 mA | 100 nA | 0.05% + 0.01% + 5 | " | ||
300.00 mA | 1 µA | 0.1% + 0.01% + 5 | " | ||
3.0000 A | 100 µA | 0.2% + 0.05% + 5 | " | ||
10.000 A | 1 mA | 0.2% + 0.05% + 5 | " | ||
A≃ | 300.00 µA | 10 nA | 0.5% + 30 | mq Curr., unit A, mqflags AC,DC,RMS | |
3.0000 mA | 100 nA | " | |||
30.000 mA | 1 µA | " | |||
300.00 mA | 10 µA | " | |||
3.0000 A | 100 µA | 0.7% + 30 | " | ||
10.000 A | 1 mA | 0.5% + 30 | " | ||
Ω | 300.00 Ω | 1 mΩ | 0.05% + 0.01% + 5 | mq Res., unit Ω, mqflags - | |
3.0000 kΩ | 10 mΩ | " | |||
30.000 kΩ | 100 mΩ | " | |||
300.00 kΩ | 1 Ω | " | |||
3.0000 MΩ | 10 Ω | 0.1% + 0.02% + 5 | " | ||
30.000 MΩ | 100 Ω | 1% + 0.2% + 5 | " | ||
Ohm/Cont. | 300 Ω | 100 mΩ | 1% + 3% | mq Res., unit Ω, mqflags - | |
Diode/Cont. | 300 mV | 100 µV | 0.2% + 3 | mq V, unit V, mqflags DC,DIODE | |
Diode | 3 V | 100 µV | 0.2% + 3 | mq V, unit V, mqflags DC,DIODE | |
Cap. F | 3.000 nF | 1 pF | 1.0% + 0.2 | mq Cap., unit F, mqflags - | |
30.00 nF | 10 pF | " | |||
300.0 nF | 100 pF | " | |||
3.000 µF | 1 nF | " | |||
30.00 µF | 10 nF | " | |||
300.0 µF | 100 nF | ±(5.0% + 6) | " | ||
3.000 mF | 1 µF | " | |||
30.00 mF | 1 µF | " | |||
Hz≈, ≃ | 300.00 Hz | 0.001 Hz | ±(0.05% + 1) | mq Freq., unit Hz, mqflags AC,DC | To activate in AC, press yellow button twice. |
3.0000 kHz | 0.01 Hz | " | |||
300.000 kHz | 1 Hz | " | |||
°C | Pt100/Pt1000 –200.0 - +100.0 °C |
0.1 °C | ±(0.5 K + 3) | mq Temp., unit °C, mqflags - | |
Pt100/Pt1000 +100.0 - 850.0 °C |
±(0.2% + 3) | mq Temp., unit °C, mqflags - | |||
K NiCr-Ni -270-+1372.0 °C | ±(0.7% + 3) | " | |||
J Fe-CuNi -210 - +1200.0 °C | ±(0.8% + 3) | " | Not tested yet due to lack of corresponding sensor. | ||
°F | " | " | " | mq Temp., unit °F, mqflags - | Details and ranges like °C.
|
Event count | mq V, unit V, mqflags AC,RMS,REL | Counts events >= 2500 digits that last >=2s with a break >=2s. To activate, switch to voltage ACDC, press yellow button twice. Press up to 4 times to cycle parameters, press long to stop. See manual. Not useful remotely because the current measurement value is sent, not the counter. | |||
Stop Watch | 100 ns | (Time data) | Max. 100 Min. To activate, select a 3V-1000V≃ range manually, press yellow button, press AUTO for start/stop, DATA for reset. Using the stop watch delivers only voltage data and sometimes seems to bring the device into a mode where it even sends invalid data packets. | ||
Data Hold | Disabled while IR interface is active on Matthias' test device. | ||||
MIN | mqflags += MIN | Influences only display, via IR flag is set, but actual current value is delivered. | |||
MAX | mqflags += MAX | Influences only display, via IR flag is set, but actual current value is delivered. |
The Accuracy row was simplified, see manual for exact specs. If the value contains three values, the middle value is % or range.
The column "sigrok result" contains in short form what the driver generates for the resp. data type.