Difference between revisions of "Gossen Metrawatt Metrahit 25S"
Line 36: | Line 36: | ||
== Using the device with sigrok == | == Using the device with sigrok == | ||
The device is supported in ''Send Mode'' using a [[Gossen_Metrawatt#Metrahit_RS232 | RS232 interface]] by the driver gmc-mh-1x-2x-rs232. | The device is supported in ''Send Mode'' using a [[Gossen_Metrawatt#Metrahit_RS232 | RS232 interface]] by the driver gmc-mh-1x-2x-rs232 (actually this should work, but with [[User:Matthias Heidbrink | Matthias']] test device it did not, so it's untested). Set the multimeter to use the communication adaptor ''RS232''. | ||
The multimeter is also supported in bidirectional mode using a [[Gossen_Metrawatt#Metrahit_BD232 | BD232]] or [[Gossen_Metrawatt#Metrahit_SI232-II | SI232-II]] interface by the driver gmc-mh-2x-bd232. Set the multimeter to use the communication adaptor ''BD232''. | The multimeter is also supported in bidirectional mode using a [[Gossen_Metrawatt#Metrahit_BD232 | BD232]] or [[Gossen_Metrawatt#Metrahit_SI232-II | SI232-II]] interface by the driver gmc-mh-2x-bd232. Set the multimeter to use the communication adaptor ''BD232''. |
Revision as of 22:27, 14 February 2014
Status | supported |
---|---|
Source code | gmc_mh_1x_2x |
Counts | 31000 |
IEC 61010-1 | CAT III 1000 V, CAT IV 600 V |
Connectivity | Infrared (RS232) |
Measurements | voltage, current, resistance, capacitance, temperature, diode, continuity |
Features | autorange, data hold |
Website | [1] |
The Gossen Metrawatt Metrahit 25S is a 4.5 digit, 31000 counts 1000V CAT III/600 V CAT IV handheld digital multimeter with IR (RS232) connectivity.
Hardware
- Power supply:
- Battery: 2x 1,5 V IEC R 6, IEC LR 6 or corresponding NiCd rechargeable batteries
- Fuse:
- Current 300 mA: FF (UR) 1,6 A/1000 V AC/DC (10 kA) 6,3 mm x 32 mm
- Current 3A/10A: FF (UR) 10 A/1000 V AC/DC (30 kA) 10 mm x 38 mm
Photos
Using the device with sigrok
The device is supported in Send Mode using a RS232 interface by the driver gmc-mh-1x-2x-rs232 (actually this should work, but with Matthias' test device it did not, so it's untested). Set the multimeter to use the communication adaptor RS232.
The multimeter is also supported in bidirectional mode using a BD232 or SI232-II interface by the driver gmc-mh-2x-bd232. Set the multimeter to use the communication adaptor BD232.
The device might be not detected by the driver if it is in a mode where it sends data rarely, either with a huge send interval or with a measurement function that delivers a very low data rate, e.g. temperature measurements with K type sensor. Then it should be set to a mode with higher data rate during the initialisation of the driver.
Measurement functions and ranges
Measurement function | Range | Resolution | Accuracy | sigrok result | Remarks |
---|---|---|---|---|---|
V= | 300 mV | 10 μV | ±(0,05% 3) | mq V, unit V, mqflags DC | |
3 V | 100 μV | " | |||
30 V | 1 mV | " | |||
300 V | 10 mV | " | |||
1000 V | 100 mV | " | |||
V≈ | 300 mV | 100 µV | ±(0,5% + 30) (> 500 Digit)) | mq V, unit V, mqflags AC,RMS | |
3 V | 100 μV | ±(0,2% + 30) | " | ||
30 V | 1 mV | " | |||
300 V | 10 mV | " | |||
1000 V | 100 mV | " | |||
V db≈ | -48 - 63 dB | 0,01 dB | ± 0,1 dB | mq V, unit V, mqflags AC,RMS | Actually the 5 AC ranges, see there for accuracy. |
A= | 300.00 µA | 10 nA | 0,1% + 5 | mq Curr., unit A, mqflags DC | |
3.0000 mA | 100 nA | " | |||
30.000 mA | 1 mA | 0,05% + 5 | " | ||
300.00 mA | 10 µA | 0,5% + 5 | " | ||
3.0000 A | 100 µA | 0,5% + 10 | " | ||
10.000 A | 1 mA | " | |||
A≃ | 300.00 µA | 10 nA | 0,5% + 30 | mq Curr., unit A, mqflags AC,DC,RMS | |
3.0000 mA | 100 nA | " | |||
30.000 mA | 1 µA | " | |||
300.00 mA | 10 µA | " | |||
3.0000 A | 100 µA | 0,75% + 30 | " | ||
10.000 A | 1 mA | " | |||
Ω | 300.00 Ω | 10 mΩ | 0,1% + 5 | mq Res., unit Ω, mqflags - | |
3.0000 kΩ | 100 mΩ | " | |||
30.000 kΩ | 1 Ω | " | |||
300.00 kΩ | 10 Ω | " | |||
3.0000 MΩ | 100 Ω | " | |||
30.000 MΩ | 1k Ω | 2% + 5 | " | ||
Ohm/Cont. | 300 Ω | 100 mΩ | 1% + 3% | mq Res., unit Ω, mqflags - | |
Diode | 3 V | 1 mV | 0.2% + 3 | mq V, unit V, mqflags DC,DIODE | |
Cap. F | 3.000 nF | 1 pF | 1.0% + 5 | mq Cap., unit F, mqflags - | |
30.00 nF | 10 pF | " | |||
300.0 nF | 100 pF | " | |||
3.000 µF | 1 nF | " | |||
30.00 µF | 10 nF | " | |||
300.0 µF | 100 nF | ±(5.0% + 6) | " | ||
3.000 mF | 1 µF | " | |||
30.00 mF | 1 µF | ±(5.0% + 30) | " | ||
Hz≈, ≃ | 300.00 Hz | 0.01 Hz | ±(0.1% + 1) | mq Freq., unit Hz, mqflags AC,DC | To activate in AC, press yellow button twice. |
3.0000 kHz | 0.1 Hz | " | |||
100.000 kHz | 10 Hz | " | |||
°C | Pt100/Pt1000 –200.0 - -100.0 °C |
0.1 °C | 1 K | mq Temp., unit °C, mqflags - | |
Pt100/Pt1000 -100.0 - +100.0 °C |
0,8 K | " | |||
Pt100/Pt1000 +100.0 - +850.0 °C |
±(0,5% + 3) | " | |||
°F | " | " | " | mq Temp., unit °C (!), mqflags - | Details and ranges like °C. The protocol does not allow to distinguish between °C and °F, so the unit is always °C.
|
Event count | mq V, unit V, mqflags AC,RMS,REL | To activate, switch to voltage AC, press yellow button twice. Press up to 4 times to cycle parameters, press long to stop. See manual. Not useful remotely because the voltage measurement value is sent, not the counter. | |||
Stop Watch | 100 ns | (Time data) | Max. 100 Min. Press AUTO for start/stop, DATA for reset. Not useful remotely because the stop watch delivers only zeros. | ||
Data Hold | mqflags += HOLD | Disabled while IR interface is active on Matthias' test device. | |||
MIN | mqflags += MIN | Influences only display, via IR flag is set, but actual current value is delivered. | |||
MAX | mqflags += MAX | Influences only display, via IR flag is set, but actual current value is delivered. |
The Accuracy row was simplified, see manual for exact specs. If the value contains three values, the middle value is % or range.
The column "sigrok result" contains in short form what the driver generates for the resp. data type.