Difference between revisions of "Gossen Metrawatt Metrahit 14S"
Jump to navigation
Jump to search
m (→Measurement functions and ranges: Fixed Ohms precision values.) |
(→Measurement functions and ranges: Some fixes of technical data) |
||
Line 55: | Line 55: | ||
| 30.00 mV | | 30.00 mV | ||
| 10 μV | | 10 μV | ||
| rowspan="2" | ±( | | rowspan="2" | ±(0,5% + 3) | ||
| mq V, unit V, mqflags DC | | mq V, unit V, mqflags DC | ||
| | | | ||
Line 68: | Line 68: | ||
| 3.000 V | | 3.000 V | ||
| 1 mV | | 1 mV | ||
| rowspan="3" | ±(0, | | rowspan="3" | ±(0,25% + 1) | ||
| " | | " | ||
| | | | ||
Line 87: | Line 87: | ||
| 1000 V | | 1000 V | ||
| 1 V | | 1 V | ||
| ±(0, | | ±(0,35% + 1) | ||
| " | | " | ||
| | | | ||
Line 95: | Line 95: | ||
| 3.000 V | | 3.000 V | ||
| 100 μV | | 100 μV | ||
| rowspan="4" | ±(0, | | rowspan="4" | ±(0,75% + 2) | ||
| " | | " | ||
| | | | ||
Line 121: | Line 121: | ||
| 300.0 µA | | 300.0 µA | ||
| 100 nA | | 100 nA | ||
| | | ±(1,0% + 5) | ||
| mq Curr., unit A, mqflags DC | | mq Curr., unit A, mqflags DC | ||
| | | Disabled on power on, see below. | ||
|- | |- | ||
| 3.000 mA | | 3.000 mA | ||
| 1 µA | | 1 µA | ||
| | | ±(1,0 + 2) | ||
| " | | " | ||
| | | Disabled on power on, see below. | ||
|- | |- | ||
| 30.00 mA | | 30.00 mA | ||
| 10 µA | | 10 µA | ||
| | | ±(1,0% + 5) | ||
| " | | " | ||
| | | | ||
Line 142: | Line 142: | ||
| 300.0 mA | | 300.0 mA | ||
| 100 µA | | 100 µA | ||
| | | ±(1,0% + 2) | ||
| " | | " | ||
| | | | ||
Line 149: | Line 149: | ||
| 3.000 A | | 3.000 A | ||
| 1 mA | | 1 mA | ||
| | | ±(1,0% + 5) | ||
| " | | " | ||
| | | Disabled on power on, see below. | ||
|- | |- | ||
| 10.00 A | | 10.00 A | ||
| 10 mA | | 10 mA | ||
| | | ±(1,0% + 2) | ||
| " | | " | ||
| | | | ||
|- | |- | ||
| rowspan=" | | rowspan="3" | A≈ | ||
| 3.000 mA | | 3.000 mA | ||
| 1 µA | | 1 µA | ||
| | | rowspan="3" | ±(1,5% + 2) | ||
| mq Curr., unit A, mqflags AC,DC,RMS | | mq Curr., unit A, mqflags AC,DC,RMS | ||
| | | | ||
|- | |- | ||
| 300.0 mA | | 300.0 mA | ||
| 100 µA | | 100 µA | ||
| " | | " | ||
| | | | ||
|- | |- | ||
| 10.00 A | | 10.00 A | ||
| 10 mA | | 10 mA | ||
| " | | " | ||
| | | | ||
Line 205: | Line 189: | ||
|- | |- | ||
| 300 Ω | | 300.0 Ω | ||
| 100 mΩ | | 100 mΩ | ||
| | | | ||
Line 305: | Line 289: | ||
|- | |- | ||
|} | |} | ||
To enable the ranges 300.0 µA, 3.000 mA and 3.000 A, switch function selector to circle with dot symbol and press yellow button. Wait until auto calibration of these ranges is finished ("CAL" in display disappears). After that the ranges are available in the mA resp. A function until multimeter powered off automatically or manually. | |||
The column "sigrok result" contains in short form what the driver generates for the resp. data type. | The column "sigrok result" contains in short form what the driver generates for the resp. data type. |
Revision as of 16:58, 12 February 2014
Status | planned |
---|---|
Source code | gmx_mh_1x_2x |
Counts | 3100 |
IEC 61010-1 | CAT III 1000 V, CAT IV 600 V |
Connectivity | Infrared (RS232) |
Measurements | voltage, current, resistance, capacitance, temperature, diode, continuity |
Features | autorange, data hold, bargraph |
The Gossen Metrawatt Metrahit 14S is a 3100 counts handheld digital multimeter with IR (RS232) connectivity.
This page is a work in progress, information incomplete and possibly inaccurate!
Measurement functions and ranges
Measurement function | Range | Resolution | Accuracy | sigrok result | Remarks |
---|---|---|---|---|---|
V= | 30.00 mV | 10 μV | ±(0,5% + 3) | mq V, unit V, mqflags DC | |
300.0 mV | 100 μV | " | |||
3.000 V | 1 mV | ±(0,25% + 1) | " | ||
30.00 V | 10 mV | " | |||
300.0 V | 100 mV | " | |||
1000 V | 1 V | ±(0,35% + 1) | " | ||
V≈ | 3.000 V | 100 μV | ±(0,75% + 2) | " | |
30.00 V | 100 μV | " | |||
300.0 V | 1 mV | " | |||
1000 V | 10 mV | " | |||
A= | 300.0 µA | 100 nA | ±(1,0% + 5) | mq Curr., unit A, mqflags DC | Disabled on power on, see below. |
3.000 mA | 1 µA | ±(1,0 + 2) | " | Disabled on power on, see below. | |
30.00 mA | 10 µA | ±(1,0% + 5) | " | ||
300.0 mA | 100 µA | ±(1,0% + 2) | " | ||
3.000 A | 1 mA | ±(1,0% + 5) | " | Disabled on power on, see below. | |
10.00 A | 10 mA | ±(1,0% + 2) | " | ||
A≈ | 3.000 mA | 1 µA | ±(1,5% + 2) | mq Curr., unit A, mqflags AC,DC,RMS | |
300.0 mA | 100 µA | " | |||
10.00 A | 10 mA | " | |||
Ω | 30.00 Ω | 10 mΩ | mq Res., unit Ω, mqflags - | ||
300.0 Ω | 100 mΩ | " | |||
3.000 kΩ | 1 Ω | " | |||
30.00 kΩ | 10 Ω | " | |||
300.0 kΩ | 100 Ω | " | |||
3.000 MΩ | 1 kΩ | " | |||
30.00 MΩ | 10 kΩ | " | |||
Diode/Cont. | 3.000 V– | 1 mV | mq V, unit V, mqflags DC,DIODE | Diode and cont. are identical on this device, just without beeper for diode. | |
°C | Pt100 –200.0 - +200.0 °C | 0.1 °C | 2 Kelvin + 5 | mq Temp., unit °C, mqflags - | |
Pt100 +200.0 - +850.0 °C | ±(1.0% + 5) | " | |||
Pt1000 –100.0 - +200.0 °C | 2 Kelvin + 2 | " | |||
Pt1000 +200.0 - +850.0 °C | ±(1.0% + 2) | " | |||
Data Hold | mqflags += HOLD | Disabled while IR interface is active. | |||
MIN | mqflags += MIN | Influences only display, via IR flag is set, but actual current value is delivered. | |||
MAX | mqflags += MAX | Influences only display, via IR flag is set, but actual current value is delivered. |
To enable the ranges 300.0 µA, 3.000 mA and 3.000 A, switch function selector to circle with dot symbol and press yellow button. Wait until auto calibration of these ranges is finished ("CAL" in display disappears). After that the ranges are available in the mA resp. A function until multimeter powered off automatically or manually.
The column "sigrok result" contains in short form what the driver generates for the resp. data type.