Difference between revisions of "Voltcraft 4080"

From sigrok
Jump to navigation Jump to search
Line 7: Line 7:
The description of the commands you find [http://www.produktinfo.conrad.com/datenblaetter/100000-124999/121064-da-01-en-Schnittstellenbeschr_LCR_4080_Handmessg.pdf at Conrad].
The description of the commands you find [http://www.produktinfo.conrad.com/datenblaetter/100000-124999/121064-da-01-en-Schnittstellenbeschr_LCR_4080_Handmessg.pdf at Conrad].


When the device does not respond it might be in setup mode. You can see that in the display. But the software can not. So it might be a good idea to do an initialization with [BXXXXXX] to stop the setup mode.
When the device does not respond it might be in setup mode. You can see that in the display. But the software can not. So it might be a good idea to do an initialization with [BXXXXXX] to stop the setup mode. This only works for a setup mode that was requested by an earlier remote session. You can not leave a manual setup mode this way.





Revision as of 17:15, 20 January 2014

The Voltcraft Plus 4080 LCR Meter Dual Display comes with an optically coupled serial interface.

Protocol

You can access the device e.g. with cutecom. The protocol is basically human readable.

The description of the commands you find at Conrad.

When the device does not respond it might be in setup mode. You can see that in the display. But the software can not. So it might be a good idea to do an initialization with [BXXXXXX] to stop the setup mode. This only works for a setup mode that was requested by an earlier remote session. You can not leave a manual setup mode this way.


Some findings not mentioned in the protocol definition above; without waranty:

disabling keys

sending

A

responds with

REMOTE READY..

and says RS232 in the display

Keys on device are disabled

enabling keys

sending

B

responds with

REMOTE STOP...

Keys on device are enabled

Images

Device complete with rubber bumper and optical/serial cable:

Voltcraft4080 complete.jpg

Screenshot of a short terminal session using CuteCom:

Voltcraft4080 cutecom.png