Difference between revisions of "Gossen Metrawatt Metrahit 14A"
Jump to navigation
Jump to search
m (→Measurement functions and ranges: More fixes) |
(→Measurement functions and ranges: Added explanation for mysterious 3 mA and 3 A ranges.) |
||
Line 123: | Line 123: | ||
| | | | ||
| mq Curr., unit A, mqflags DC | | mq Curr., unit A, mqflags DC | ||
| | | Disabled on power on, see below. | ||
|- | |- | ||
Line 144: | Line 144: | ||
| | | | ||
| " | | " | ||
| | | Disabled on power on, see below. | ||
|- | |- | ||
Line 284: | Line 284: | ||
|- | |- | ||
|} | |} | ||
To enable the ranges 3.000 mA and 3.000 A, switch function selector to circle with dot symbol and press yellow button. Wait until auto calibration of these ranges is finished ("CAL" in display disappears). After that the ranges are available in the mA resp. A function until multimeter powered off automatically or manually. | |||
The column "sigrok result" contains in short form what the driver generates for the resp. data type. | The column "sigrok result" contains in short form what the driver generates for the resp. data type. |
Revision as of 23:26, 7 January 2014
Status | supported |
---|---|
Source code | gmx_mh_1x_2x |
Counts | 3100 |
IEC 61010-1 | CAT III 1000 V, CAT IV 600 V |
Connectivity | Infrared (RS232) |
Measurements | voltage, current, resistance, temperature, diode, continuity |
Features | autorange, data hold, bargraph |
The Gossen Metrawatt Metrahit 14A is a 3100 counts handheld digital multimeter with IR (RS232) connectivity.
It is very similar to the Metrahit 14S, but has no ranges of 3 µA= and 3 mA≈.
Hardware
- TODO.
Photos
Measurement functions and ranges
Measurement function | Range | Resolution | Accuracy | sigrok result | Remarks |
---|---|---|---|---|---|
V= | 30.00 mV | 10 μV | ±(0,5% + 3) | mq V, unit V, mqflags DC | Manual range |
300.0 mV | 100 μV | " | " | ||
3.000 V | 1 mV | ±(0,25% + 1) | " | ||
30.00 V | 10 mV | " | |||
300.0 V | 100 mV | " | |||
1000 V | 1 V | ±(0,35% + 1) | " | ||
V≈ | 3.000 V | 1 mV | ±(0,75% + 2) | " | |
30.00 V | 10 mV | " | |||
300.0 V | 100 mV | " | |||
1000 V | 1 V | " | |||
A= | 3.000 mA | 1 µA | mq Curr., unit A, mqflags DC | Disabled on power on, see below. | |
30.00 mA | 10 µA | " | |||
300.0 mA | 100 µA | " | |||
3.000 A | 1 mA | " | Disabled on power on, see below. | ||
10.00 A | 10 mA | " | |||
A≈ | 30.00 mA | 10 µA | mq Curr., unit A, mqflags AC | ||
300.0 mA | 100 µA | " | |||
10.00 A | 10 mA | " | |||
Ω | 30.00 Ω | 10 mΩ | mq Res., unit Ω, mqflags - | ||
300 Ω | 100 mΩ | " | |||
3.000 kΩ | 1 Ω | " | |||
30.00 kΩ | 10 Ω | " | |||
300.0 kΩ | 100 Ω | " | |||
3.000 MΩ | 1 kΩ | " | |||
30.00 MΩ | 10 kΩ | " | |||
Diode/Cont. | 3.000 V– | 1 mV | mq V, unit V, mqflags DC,DIODE | Diode and cont. are identical on this device, just without beeper for diode. | |
°C | Pt100 –200.0 - +200.0 °C | 0.1 °C | 2 Kelvin + 5 | mq Temp., unit °C, mqflags - | |
Pt100 +200.0 - +850.0 °C | ±(1.0% + 5) | " | |||
Pt1000 –100.0 - +200.0 °C | 2 Kelvin + 2 | " | |||
Pt1000 +200.0 - +850.0 °C | ±(1.0% + 2) | " | |||
Data Hold | mqflags += HOLD | Disabled while IR interface is active. | |||
MIN | mqflags += MIN | Influences only display, via IR flag is set, but actual current value is delivered. | |||
MAX | mqflags += MAX | Influences only display, via IR flag is set, but actual current value is delivered. |
To enable the ranges 3.000 mA and 3.000 A, switch function selector to circle with dot symbol and press yellow button. Wait until auto calibration of these ranges is finished ("CAL" in display disappears). After that the ranges are available in the mA resp. A function until multimeter powered off automatically or manually.
The column "sigrok result" contains in short form what the driver generates for the resp. data type.