Difference between revisions of "Gossen Metrawatt Metrahit 27I"
Jump to navigation
Jump to search
Line 2: | Line 2: | ||
| image = <!-- [[File:Siemens B1105 small.png|180px]] --> | | image = <!-- [[File:Siemens B1105 small.png|180px]] --> | ||
| name = Gossen Metrawatt 27I | | name = Gossen Metrawatt 27I | ||
| status = | | status = in progress | ||
| source_code_dir = gmc_mh_1x_2x | | source_code_dir = gmc_mh_1x_2x | ||
| counts = 31000 | | counts = 31000 | ||
Line 14: | Line 14: | ||
'''Please note: This page is a work in progress and still partly incorrect.''' | '''Please note: This page is a work in progress and still partly incorrect.''' | ||
The '''[[Gossen Metrawatt]] Metrahit 27I''' is a 3100/31000 counts 600 V CAT II handheld digital multimeter with IR (RS232/USB) connectivity. It supports milliohm and insulation (megaohm) measurements and has data logging capabilities. | The '''[[Gossen Metrawatt]] Metrahit 27I''' is a 3100/31000 counts 600 V CAT II handheld digital multimeter with bidirectional IR (RS232/USB) connectivity. It supports milliohm and insulation (megaohm) measurements and has data logging capabilities. | ||
There are two versions of this device, an older one with yellow rotary switch and a newer one with red rotary switch. They also have different power supply connectors. | There are two versions of this device, an older one with yellow rotary switch and a newer one with red rotary switch. They also have different power supply connectors. | ||
Line 47: | Line 47: | ||
The device has got a device address that can be set to values from 0 to 15 in the setup. Using the device address 15 is recommended if only a single device is used, it sets a special mode with simplified addressing. The device address 0 sets a special mode where the multimeter does not send replies. It cannot be used with sigrok. | The device has got a device address that can be set to values from 0 to 15 in the setup. Using the device address 15 is recommended if only a single device is used, it sets a special mode with simplified addressing. The device address 0 sets a special mode where the multimeter does not send replies. It cannot be used with sigrok. | ||
If several devices of the Metrahit 27x family are used on the same RS232 interface using cascaded ''RS232'' interfaces, they need to be set to different device addresses from 1 to 15. In opposite to what the manual says, it seems that the device address 15 can be used, too. | If several devices of the Metrahit 27x family are used on the same RS232 interface using cascaded ''RS232'' interfaces, they need to be set to different device addresses from 1 to 15. In opposite to what the manual says, it seems that the device address 15 can be used, too. Using several devices on the same RS232 interface is an untested feature. | ||
Please contact [[User:Matthias Heidbrink | Matthias']] if you can test one of the untested features. | Please contact [[User:Matthias Heidbrink | Matthias']] if you can test one of the untested features. | ||
Line 286: | Line 286: | ||
| | | | ||
| | | | ||
| mqflags += HOLD | |||
| | | | ||
|- | |- | ||
Line 294: | Line 294: | ||
| | | | ||
| | | | ||
| mqflags += MIN | |||
| | | | ||
|- | |- | ||
Line 302: | Line 302: | ||
| | | | ||
| | | | ||
| mqflags += MAX | |||
| | | | ||
|- | |- | ||
Line 314: | Line 314: | ||
[[Category:Device]] | [[Category:Device]] | ||
[[Category:Multimeter]] | [[Category:Multimeter]] | ||
[[Category: | [[Category:In progress]] |
Revision as of 11:17, 31 May 2014
Status | in progress |
---|---|
Source code | gmc_mh_1x_2x |
Counts | 31000 |
IEC 61010-1 | CAT II 600 V |
Connectivity | Infrared (RS232/USB) |
Measurements | voltage, resistance, temperature, diode, continuity |
Features | autorange, data hold, backlight, insulation measurement, data logging |
Website | [1] |
Please note: This page is a work in progress and still partly incorrect.
The Gossen Metrawatt Metrahit 27I is a 3100/31000 counts 600 V CAT II handheld digital multimeter with bidirectional IR (RS232/USB) connectivity. It supports milliohm and insulation (megaohm) measurements and has data logging capabilities.
There are two versions of this device, an older one with yellow rotary switch and a newer one with red rotary switch. They also have different power supply connectors.
Hardware
- Power supply:
- Battery: 3x 1,2 V NiMH rechargeable batteries
- External power supply NA HIT 27 (Art. Z218J) by Gossen Metrawatt, 5V, 600 mA (newer version)
- Fuse:
- FF (UR) 1,6 A/1000 V AC/DC (10 kA) 6,3 mm x 32 mm
Photos
Communication Protocol
The complete description is available on request from the manufacturer.
Measurement functions and ranges
Measurement function | Range | Resolution | Accuracy | sigrok result | Remarks |
---|---|---|---|---|---|
V= | 3 V | 100 μV | ±(0.1% + 10) | mq V, unit V, mqflags DC | |
30 V | 1 mV | ±(0.1% + 5) | " | ||
300 V | 10 mV | " | |||
600 V | 100 mV | " | |||
V≈ | 3 V | 100 μV | ±(0.2% + 10) | mq V, unit V, mqflags AC | |
30 V | 1 mV | " | |||
300 V | 10 mV | " | |||
600 V | 100 mV | " | |||
mΩ@1A | 3 mΩ | 0.001 mΩ | ±(1% + 10) | mq Res., unit Ω, mqflags - | 4-wire Kelvin measurement |
30 mΩ | 0.001 mΩ | ±(0.5% + 10) | " | " | |
300 mΩ | 0.01 mΩ | " | " | ||
mΩ@200 mA | 30 mΩ | 0.01 mΩ | ±(0.25% + 10) | " | " |
300 mΩ | 0.01 mΩ | " | " | ||
3 Ω | 0.1 mΩ | " | " | ||
30 Ω | 1 mΩ | " | " | ||
Ω | 300.00 Ω | 1 mΩ | ±(0.1% + 10) | " | ZERO with yellow button |
3.0000 kΩ | 10 mΩ | ±(0.1% + 5) | " | " | |
30.000 kΩ | 100 mΩ | " | |||
300.00 kΩ | 1 Ω | " | |||
3.0000 MΩ | 10 Ω | " | |||
30.000 MΩ | 100 Ω | ±(1.5% + 10) | " | ||
MΩ (Insulation) | 30 MΩ | 10 kΩ | ±(2% + 10) | " | Insulation measurement, test voltage selectable 50/100/250/500 V |
300 MΩ | 100 kΩ | " | " | ||
3 GΩ | 1 MΩ | ±(2% + 10) | " | " | |
Cont. | 310 Ω | 100 mΩ | ±(1% + 5) | mq Res., unit Ω, mqflags - | Signal < about 10 Ω |
Diode | 3 V | 100 µV | ±(1% + 5) | mq V, unit V, mqflags DC,DIODE | |
Hz | 300.00 Hz | 0.01 Hz | ±(0.05% + 5) | mq Freq., unit Hz, mqflags AC | |
3.000 kHz | 0.1 Hz | " | |||
°C | Pt100/Pt1000 –200.0 - +100.0 °C |
0.1 °C | ±(1 K + 5) | mq Temp., unit °C, mqflags - | |
Pt100/Pt1000 +100.0 - 850.0 °C |
±(0.5% + 5) | " | |||
Ni 100/Ni1000 -60-+180 °C |
" | Not tested yet due to lack of corresponding sensor. | |||
°F | " | " | " | mq Temp., unit °F, mqflags - | Details and ranges like °C. |
Data Hold | mqflags += HOLD | ||||
MIN | mqflags += MIN | ||||
MAX | mqflags += MAX |
The Accuracy row was simplified, see manual for exact specs.
The column "sigrok result" contains in short form what the driver generates for the resp. data type.