The '''[[Gossen Metrawatt]] Metrahit 27I''' is a 3100/31000 counts 600 V CAT II handheld digital multimeter with IR (RS232) connectivity. It supports milliohm and insulation (megaohm) measurements and has data logging capabilities. | The '''[[Gossen Metrawatt]] Metrahit 27I''' is a 3100/31000 counts 600 V CAT II handheld digital multimeter with IR (RS232) connectivity. It supports milliohm and insulation (megaohm) measurements and has data logging capabilities. | ||
There are two versions of this device, an older one with yellow rotary switch and a newer one with red rotary switch. They also have different power supply connectors. | |||
== Hardware == | == Hardware == | ||
Line 20: | Line 22: | ||
* '''Power supply''': | * '''Power supply''': | ||
** Battery: 3x ''1,2 V NiMH'' rechargeable batteries | ** Battery: 3x ''1,2 V NiMH'' rechargeable batteries | ||
** External power supply ''NA HIT 27'' (Art. Z218J) by Gossen Metrawatt, 5V, 600 mA | ** External power supply ''NA HIT 27'' (Art. Z218J) by Gossen Metrawatt, 5V, 600 mA (newer version) | ||
* '''Fuse''': | * '''Fuse''': | ||
** FF (UR) 1,6 A/1000 V AC/DC (10 kA) 6,3 mm x 32 mm | ** FF (UR) 1,6 A/1000 V AC/DC (10 kA) 6,3 mm x 32 mm |
Revision as of 11:00, 7 April 2014
Status | planned |
---|---|
Source code | gmc_mh_1x_2x |
Counts | 31000 |
IEC 61010-1 | CAT II 600 V |
Connectivity | Infrared (RS232/USB) |
Measurements | voltage, resistance, temperature, diode, continuity |
Features | autorange, data hold, backlight, insulation measurement, data logging |
Website | [1] |
Please note: This page is a work in progress and still partly incorrect.
The Gossen Metrawatt Metrahit 27I is a 3100/31000 counts 600 V CAT II handheld digital multimeter with IR (RS232) connectivity. It supports milliohm and insulation (megaohm) measurements and has data logging capabilities.
There are two versions of this device, an older one with yellow rotary switch and a newer one with red rotary switch. They also have different power supply connectors.
Hardware
- Power supply:
- Battery: 3x 1,2 V NiMH rechargeable batteries
- External power supply NA HIT 27 (Art. Z218J) by Gossen Metrawatt, 5V, 600 mA (newer version)
- Fuse:
- FF (UR) 1,6 A/1000 V AC/DC (10 kA) 6,3 mm x 32 mm
Photos
Measurement functions and ranges
Measurement function | Range | Resolution | Accuracy | sigrok result | Remarks |
---|---|---|---|---|---|
V= | 3 V | 100 μV | ±(0.1% + 10) | mq V, unit V, mqflags DC | |
30 V | 1 mV | ±(0.1% + 5) | " | ||
300 V | 10 mV | " | |||
600 V | 100 mV | " | |||
V≈ | 3 V | 100 μV | ±(0.2% + 10) | mq V, unit V, mqflags AC | |
30 V | 1 mV | " | |||
300 V | 10 mV | " | |||
600 V | 100 mV | " | |||
mΩ@1A | 3 mΩ | 0.001 mΩ | ±(1% + 10) | mq Res., unit Ω, mqflags - | 4-wire Kelvin measurement |
30 mΩ | 0.001 mΩ | ±(0.5% + 10) | " | " | |
300 mΩ | 0.01 mΩ | " | " | ||
mΩ@200 mA | 30 mΩ | 0.01 mΩ | ±(0.25% + 10) | " | " |
300 mΩ | 0.01 mΩ | " | " | ||
3 Ω | 0.1 mΩ | " | " | ||
30 Ω | 1 mΩ | " | " | ||
Ω | 300.00 Ω | 1 mΩ | ±(0.1% + 10) | " | ZERO with yellow button |
3.0000 kΩ | 10 mΩ | ±(0.1% + 5) | " | " | |
30.000 kΩ | 100 mΩ | " | |||
300.00 kΩ | 1 Ω | " | |||
3.0000 MΩ | 10 Ω | " | |||
30.000 MΩ | 100 Ω | ±(1.5% + 10) | " | ||
MΩ | 30 MΩ | 10 kΩ | ±(2% + 10) | " | Test voltage selectable 50/100/250/500 V |
300 MΩ | 100 kΩ | " | " | ||
3 GΩ | 1 MΩ | ±(2% + 10) | " | " | |
Cont. | 310 Ω | 100 mΩ | ±(1% + 5) | mq Res., unit Ω, mqflags - | Signal < about 10 Ω |
Diode | 3 V | 100 µV | ±(1% + 5) | mq V, unit V, mqflags DC,DIODE | |
Hz | 300.00 Hz | 0.01 Hz | ±(0.05% + 5) | mq Freq., unit Hz, mqflags AC | |
3.000 kHz | 0.1 Hz | " | |||
°C | Pt100/Pt1000 –200.0 - +100.0 °C |
0.1 °C | ±(1 K + 5) | mq Temp., unit °C, mqflags - | |
Pt100/Pt1000 +100.0 - 850.0 °C |
±(0.5% + 5) | " | |||
Ni 100/Ni1000 -60-+180 °C |
" | Not tested yet due to lack of corresponding sensor. | |||
°F | " | " | " | mq Temp., unit °F, mqflags - | Details and ranges like °C. |
Data Hold | mqflags += HOLD | Disabled while IR interface is active on Matthias' test device. | |||
MIN | mqflags += MIN | Influences only display, via IR flag is set, but actual current value is delivered. | |||
MAX | mqflags += MAX | Influences only display, via IR flag is set, but actual current value is delivered. |
The Accuracy row was simplified, see manual for exact specs.
The column "sigrok result" contains in short form what the driver generates for the resp. data type.