Difference between revisions of "Gossen Metrawatt Metrahit 27I"
Jump to navigation
Jump to search
(Created page (stub/work).) |
m |
||
Line 195: | Line 195: | ||
|- | |- | ||
| | | rowspan="3" | MΩ | ||
| | | 30 MΩ | ||
| | | 10 kΩ | ||
| | | rowspan="2" | 2% + 10 | ||
| mq Res., unit Ω, mqflags - | | mq Res., unit Ω, mqflags - | ||
| | | | ||
|- | |- | ||
| | | 300 MΩ | ||
| | | 100 kΩ | ||
| " | | " | ||
| | | | ||
|- | |- | ||
| | | 3 GΩ | ||
| | | 1 MΩ | ||
| 2% + 10 | |||
| " | | " | ||
| | | | ||
|- | |- | ||
| | | Cont. | ||
| 1 | | 300 Ω | ||
| | | 100 mΩ | ||
| 1% + 5 | |||
| mq Res., unit Ω, mqflags - | |||
| | | | ||
|- | |- | ||
| | | Diode | ||
| | | 3 V | ||
| 100 µV | |||
| 1% + 5 | |||
| mq V, unit V, mqflags DC,DIODE | |||
| 100 | |||
| | |||
| | |||
| | | | ||
Line 284: | Line 246: | ||
|- | |- | ||
| rowspan=" | | rowspan="3" | °C | ||
| Pt100/Pt1000<br> –200.0 - +100.0 °C | | Pt100/Pt1000 <br> –200.0 - +100.0 °C | ||
| rowspan=" | | rowspan="3" | 0.1 °C | ||
| ±( | | ±(1 K + 5) | ||
| mq Temp., unit °C, mqflags - | | mq Temp., unit °C, mqflags - | ||
| | | | ||
Line 293: | Line 255: | ||
|- | |- | ||
| Pt100/Pt1000<br> +100.0 - 850.0 °C | | Pt100/Pt1000<br> +100.0 - 850.0 °C | ||
| ±(0, | | ±(0,5% + 5) | ||
| mq Temp., unit °C, mqflags - | | mq Temp., unit °C, mqflags - | ||
| | | | ||
|- | |- | ||
| | | NI 100/NI1000<br>-60-+180 °C | ||
| ±(0,5 + 5) | |||
| ±(0, | |||
| " | | " | ||
| Not tested yet due to lack of corresponding sensor. | | Not tested yet due to lack of corresponding sensor. |
Revision as of 20:46, 6 April 2014
Status | planned |
---|---|
Source code | gmc_mh_1x_2x |
Counts | 31000 |
IEC 61010-1 | CAT II 600 V |
Connectivity | Infrared (RS232) |
Measurements | voltage, resistance, temperature, diode, continuity |
Features | autorange, data hold, backlight, insulation measurement |
Website | [1] |
Please note: This page is a work in progress and still partly incorrect.
The Gossen Metrawatt Metrahit 27I is a 3100/31000 counts 600 V CAT II handheld digital multimeter with IR (RS232) connectivity. It supports milliohm and insulation (megaohm) measurements.
Hardware
- Power supply:
- Battery: 3x 1,2 V NiMH rechargeable batteries
- External power supply NA HIT 27 (Art. Z218J) by Gossen Metrawatt, 5V, 600 mA
- Fuse:
- FF (UR) 1,6 A/1000 V AC/DC (10 kA) 6,3 mm x 32 mm
Photos
Measurement functions and ranges
Measurement function | Range | Resolution | Accuracy | sigrok result | Remarks |
---|---|---|---|---|---|
V= | 3 V | 100 μV | ±(0,1% + 10) | mq V, unit V, mqflags DC | |
30 V | 1 mV | ±(0,1% + 5) | " | ||
300 V | 10 mV | " | |||
600 V | 100 mV | " | |||
V≈ | 3 V | 100 μV | ±(0,2% + 10) | mq V, unit V, mqflags AC | |
30 V | 1 mV | " | |||
300 V | 10 mV | " | |||
600 V | 100 mV | " | |||
mΩ@1A | 3 mΩ | 0.001 mΩ | 1% + 10 | mq Res., unit Ω, mqflags - | |
30 mΩ | 0.001 mΩ | 0.5% + 10 | |||
300 mΩ | 0.01 mΩ | " | |||
mΩ@200 mA | 30 mΩ | 0.01 mΩ | 0.25% + 10 | mq Res., unit Ω, mqflags - | |
300 mΩ | 0.01 mΩ | 0.5% + 10 | |||
3 Ω | 0.1 mΩ | " | |||
30 Ω | 1 mΩ | " | |||
Ω | 300.00 Ω | 1 mΩ | 0,05% + 0.01% + 5 | mq Res., unit Ω, mqflags - | |
3.0000 kΩ | 10 mΩ | " | |||
30.000 kΩ | 100 mΩ | " | |||
300.00 kΩ | 1 Ω | " | |||
3.0000 MΩ | 10 Ω | 0,1% + 0.02% + 5 | " | ||
30.000 MΩ | 100 Ω | 1% + 0.2% + 5 | " | ||
MΩ | 30 MΩ | 10 kΩ | 2% + 10 | mq Res., unit Ω, mqflags - | |
300 MΩ | 100 kΩ | " | |||
3 GΩ | 1 MΩ | 2% + 10 | " | ||
Cont. | 300 Ω | 100 mΩ | 1% + 5 | mq Res., unit Ω, mqflags - | |
Diode | 3 V | 100 µV | 1% + 5 | mq V, unit V, mqflags DC,DIODE | |
Hz | 300.00 Hz | 0.01 Hz | ±(0.05% + 5) | mq Freq., unit Hz, mqflags AC | |
3.000 kHz | 0.1 Hz | " | |||
°C | Pt100/Pt1000 –200.0 - +100.0 °C |
0.1 °C | ±(1 K + 5) | mq Temp., unit °C, mqflags - | |
Pt100/Pt1000 +100.0 - 850.0 °C |
±(0,5% + 5) | mq Temp., unit °C, mqflags - | |||
NI 100/NI1000 -60-+180 °C |
±(0,5 + 5) | " | Not tested yet due to lack of corresponding sensor. | ||
°F | " | " | " | mq Temp., unit °F, mqflags - | Details and ranges like °C. |
Data Hold | mqflags += HOLD | Disabled while IR interface is active on Matthias' test device. | |||
MIN | mqflags += MIN | Influences only display, via IR flag is set, but actual current value is delivered. | |||
MAX | mqflags += MAX | Influences only display, via IR flag is set, but actual current value is delivered. |
The Accuracy row was simplified, see manual for exact specs. If the value contains three values, the middle value is % or range.
The column "sigrok result" contains in short form what the driver generates for the resp. data type.