Gossen Metrawatt Metrahit 14S
Status planned
Source code gmx_mh_1x_2x
Counts 3100
IEC 61010-1 CAT III 1000 V, CAT IV 600 V
Connectivity Infrared (RS232)
Measurements voltage, current, resistance, capacitance, temperature, diode, continuity
Features autorange, data hold, bargraph

The Gossen Metrawatt Metrahit 14S is a 3100 counts handheld digital multimeter with IR (RS232) connectivity.

This page is a work in progress, information incomplete and possibly inaccurate!


Measurement functions and ranges

Measurement function Range Resolution Accuracy sigrok result Remarks
V= 30.00 mV 10 μV ±(1,0% + 3) mq V, unit V, mqflags DC
300.0 mV 100 μV "
3.000 V 1 mV ±(0,15% + 1) "
30.00 V 10 mV "
300.0 V 100 mV "
1000 V 1 V ±(0,2% + 1) "
V≈ 3.000 V 100 μV ±(0,4% + 2) "
30.00 V 100 μV "
300.0 V 1 mV "
1000 V 10 mV "
A=  300.0 µA 100 nA mq Curr., unit A, mqflags DC
3.000 mA 1 µA "
30.00 mA 10 µA "
300.0 mA 100 µA "
3.000 A 1 mA "
10.00 A 10 mA "
A≈ 3.000 mA 1 µA mq Curr., unit A, mqflags AC,DC,RMS
30.00 mA 10 µA " ?
300.0 mA 100 µA "
3.000 A 100 µA " ?
10.00 A 10 mA "
Ω 30.00 Ω 10 mΩ mq Res., unit Ω, mqflags -
300 Ω 100 mΩ "
3.000 kΩ 100 mΩ "
30.00 kΩ 1 Ω "
300.0 kΩ 10 Ω "
3.000 MΩ 100 Ω "
30.00 MΩ 1 kΩ "
Diode/Cont. 3.000 V– 1 mV mq V, unit V, mqflags DC,DIODE Diode and cont. are identical on this device, just without beeper for diode.
°C Pt100 –200.0 - +200.0 °C 0.1 °C 2 Kelvin + 5 mq Temp., unit °C, mqflags -
Pt100 +200.0 - +850.0 °C ±(1.0% + 5) "
Pt1000 –100.0 - +200.0 °C 2 Kelvin + 2 "
Pt1000 +200.0 - +850.0 °C ±(1.0% + 2) "
Data Hold mqflags += HOLD Disabled while IR interface is active.
MIN mqflags += MIN Influences only display, via IR flag is set, but actual current value is delivered.
MAX mqflags += MAX Influences only display, via IR flag is set, but actual current value is delivered.

The column "sigrok result" contains in short form what the driver generates for the resp. data type.