Difference between revisions of "BBC Goertz Metrawatt M2110"
Jump to navigation
Jump to search
(Created page with "{{Infobox multimeter | image = <!-- 180px --> | name = BBC Goertz Metrawatt M2110 | status = in...") |
|||
Line 15: | Line 15: | ||
= Hardware = | = Hardware = | ||
'''RS232 Interface:''' | |||
* 150/300/600/1200/2400/4800/9600 baud, configurable via rotary switch | * 150/300/600/1200/2400/4800/9600 baud, configurable via rotary switch | ||
* 7 data bits | * 7 data bits | ||
Line 24: | Line 23: | ||
* Output intervals 0,66s/10s/1m/10m/1h/manual/remote controlled, configurable via rotary switch | * Output intervals 0,66s/10s/1m/10m/1h/manual/remote controlled, configurable via rotary switch | ||
* Manual output sends value on button ''Start/Reset'' pressed. | * Manual output sends value on button ''Start/Reset'' pressed. | ||
* Remote output sends data on CTS set or ENQ sent, configurable via DIP switches | * Remote controlled output sends data on CTS set or ENQ sent, configurable via DIP switches | ||
* Max. 1,5 measurements/s. | |||
''RS232 cable:'' | |||
* 25 pin "female" connector | |||
* Null modem cable required (DTE configuration)! | |||
== Photos == | == Photos == |
Revision as of 20:03, 18 December 2013
Status | in progress |
---|---|
Source code | serial-dmm |
Counts | 30000 |
Connectivity | RS232 |
Measurements | voltage, current, resistance, capacitance, temperature, diode, continuity |
Features | manual range |
The BBC Goertz Metrawatt M2110 is a historic (about 1984) 30000 count bench multimeter with RS232 connectivity.
Hardware
RS232 Interface:
- 150/300/600/1200/2400/4800/9600 baud, configurable via rotary switch
- 7 data bits
- No/even/odd parity, configurable via DIP switches
- 1 or 2 stop bits, configurable via DIP switch
- Optional CTS handshake, configurable via DIP switch
- Output intervals 0,66s/10s/1m/10m/1h/manual/remote controlled, configurable via rotary switch
- Manual output sends value on button Start/Reset pressed.
- Remote controlled output sends data on CTS set or ENQ sent, configurable via DIP switches
- Max. 1,5 measurements/s.
RS232 cable:
- 25 pin "female" connector
- Null modem cable required (DTE configuration)!
Photos
Protocol
msg := SDDDDDD | 'OVERRNG' CR LF S := '-' | ' ' D := '0'..'9'|'.'