Difference between revisions of "Siemens B1105"
Jump to navigation
Jump to search
Line 32: | Line 32: | ||
</gallery> | </gallery> | ||
--> | --> | ||
== Measurement functions and ranges == | |||
Note: Work in progress, currently simply wrong! | |||
{|border="0" style="font-size: smaller" class="alternategrey sortable sigroktable" | |||
|- | |||
! Measurement function | |||
! Range | |||
! Resolution | |||
! Accuracy | |||
! sigrok result | |||
! Remarks | |||
|- | |||
| rowspan="5" | V= | |||
| 300 mV | |||
| 1 μV | |||
| ±(0,02% + 0,010% + 5) | |||
| mq V, unit V, mqflags DC | |||
| | |||
|- | |||
| 3 V | |||
| 10 μV | |||
| rowspan="4" | ±(0,02% + 0,005% + 5) | |||
| " | |||
| | |||
|- | |||
| 30 V | |||
| 100 μV | |||
| " | |||
| | |||
|- | |||
| 300 V | |||
| 1 mV | |||
| " | |||
| | |||
|- | |||
| 1000 V | |||
| 10 mV | |||
| " | |||
| | |||
|- | |||
| rowspan="5" | V≈ | |||
| 300 mV | |||
| 10 µV | |||
| ±(0,5% + 30) (> 500 Digit)) | |||
| mq V, unit V, mqflags AC,RMS | |||
| | |||
|- | |||
| 3 V | |||
| 100 μV | |||
| rowspan="4" | ±(0,2% + 30) | |||
| " | |||
| | |||
|- | |||
| 30 V | |||
| 100 μV | |||
| " | |||
| | |||
|- | |||
| 300 V | |||
| 1 mV | |||
| " | |||
| | |||
|- | |||
| 1000 V | |||
| 10 mV | |||
| " | |||
| | |||
|- | |||
| rowspan="5" | V≃ | |||
| 300 mV | |||
| 10 μV | |||
| ±(0,5% + 30) (> 500 Digit)) | |||
| mq V, unit V, mqflags AC,DC,RMS | |||
| | |||
|- | |||
| 3 V | |||
| 100 μV | |||
| rowspan="4" | ±(0,2% + 30) | |||
| " | |||
| | |||
|- | |||
| 30 V | |||
| 100 μV | |||
| " | |||
| | |||
|- | |||
| 300 V | |||
| 1 mV | |||
| " | |||
| | |||
|- | |||
| 1000 V | |||
| 10 mV | |||
| " | |||
| | |||
|- | |||
| V db≈ | |||
| -48 - 63 dB | |||
| 0,01 dB | |||
| | |||
| mq V, unit V, mqflags AC,RMS | |||
| Actually the 5 AC ranges, see there for accuracy. | |||
|- | |||
| V db≈ rel | |||
| -48 - 63 dB | |||
| 0,01 dB | |||
| | |||
| mq V, unit V, mqflags AC,RMS | |||
| Actually the 5 AC ranges, see there for accuracy. To activate, press yellow button in dB mode, then do second measurement. Cannot be distinguished from non-rel measurement remotely. | |||
|- | |||
| rowspan="6" | A= | |||
| 300.00 µA | |||
| 1 nA | |||
| | |||
| mq Curr., unit A, mqflags DC | |||
| | |||
|- | |||
| 3.0000 mA | |||
| 10 nA | |||
| | |||
| " | |||
| | |||
|- | |||
| 30.000 mA | |||
| 100 nA | |||
| | |||
| " | |||
| | |||
|- | |||
| 300.00 mA | |||
| 1 µA | |||
| | |||
| " | |||
| | |||
|- | |||
| 3.0000 A | |||
| 100 µA | |||
| | |||
| " | |||
| | |||
|- | |||
| 10.000 A | |||
| 1 mA | |||
| | |||
| " | |||
| | |||
|- | |||
| rowspan="6" | A≃ | |||
| 300.00 µA | |||
| 10 nA | |||
| | |||
| mq Curr., unit A, mqflags AC,DC,RMS | |||
| | |||
|- | |||
| 3.0000 mA | |||
| 100 nA | |||
| | |||
| " | |||
| | |||
|- | |||
| 30.000 mA | |||
| 1 µA | |||
| | |||
| " | |||
| | |||
|- | |||
| 300.00 mA | |||
| 10 µA | |||
| | |||
| " | |||
| | |||
|- | |||
| 3.0000 A | |||
| 100 µA | |||
| | |||
| " | |||
| | |||
|- | |||
| 10.000 A | |||
| 1 mA | |||
| | |||
| " | |||
| | |||
|- | |||
| rowspan="6" | Ω | |||
| 300.00 Ω | |||
| 1 mΩ | |||
| | |||
| mq Res., unit Ω, mqflags - | |||
| | |||
|- | |||
| 3.0000 kΩ | |||
| 10 mΩ | |||
| | |||
| " | |||
| | |||
|- | |||
| 30.000 kΩ | |||
| 100 mΩ | |||
| | |||
| " | |||
| | |||
|- | |||
| 300.00 kΩ | |||
| 1 Ω | |||
| | |||
| " | |||
| | |||
|- | |||
| 3.0000 MΩ | |||
| 10 Ω | |||
| | |||
| " | |||
| | |||
|- | |||
| 30.000 MΩ | |||
| 100 Ω | |||
| | |||
| " | |||
| | |||
|- | |||
| Ohm/Cont. | |||
| 300 Ω | |||
| 100 mΩ | |||
| | |||
| mq Res., unit Ω, mqflags - | |||
| | |||
|- | |||
| Diode/Cont. | |||
| 300 mV | |||
| 100 µV | |||
| | |||
| mq V, unit V, mqflags DC,DIODE | |||
| | |||
|- | |||
| Diode | |||
| 3 V | |||
| 100 µV | |||
| | |||
| mq V, unit V, mqflags DC,DIODE | |||
| | |||
|- | |||
| rowspan="8" | Cap. F | |||
| 3.000 nF | |||
| 1 pF | |||
| | |||
| mq Cap., unit F, mqflags - | |||
| | |||
|- | |||
| 30.00 nF | |||
| 10 pF | |||
| | |||
| " | |||
| | |||
|- | |||
| 300.0 nF | |||
| 100 pF | |||
| | |||
| " | |||
| | |||
|- | |||
| 3.000 µF | |||
| 1 nF | |||
| | |||
| " | |||
| | |||
|- | |||
| 30.00 µF | |||
| 10 nF | |||
| | |||
| " | |||
| | |||
|- | |||
| 300.0 µF | |||
| 100 nF | |||
| rowspan="3" | ±(5.0% + 6) | |||
| " | |||
| | |||
|- | |||
| 3.000 mF | |||
| 1 µF | |||
| " | |||
| | |||
|- | |||
| 30.00 mF | |||
| 1 µF | |||
| " | |||
| | |||
|- | |||
| rowspan="3" | Hz≈, ≃ | |||
| 300.00 Hz | |||
| 0.001 Hz | |||
| rowspan="3" | ±(0.05% + 1) | |||
| mq Freq., unit Hz, mqflags AC,DC | |||
| To activate in AC, press yellow button twice. | |||
|- | |||
| 3.0000 kHz | |||
| 0.01 Hz | |||
| " | |||
| | |||
|- | |||
| 300.000 kHz | |||
| 1 Hz | |||
| " | |||
| | |||
|- | |||
| rowspan="4" | °C | |||
| Pt100/Pt1000<br> –200.0 - +100.0 °C | |||
| rowspan="4" | 0.1 °C | |||
| ±(0,5 K + 3) | |||
| mq Temp., unit °C, mqflags - | |||
| | |||
|- | |||
| Pt100/Pt1000<br> +100.0 - 850.0 °C | |||
| ±(0,2% + 3) | |||
| mq Temp., unit °C, mqflags - | |||
| | |||
|- | |||
| K NiCr-Ni -270-+1372,0 °C | |||
| ±(0,7% + 3) | |||
| " | |||
| | |||
|- | |||
| J Fe-CuNi -210 - +1200.0 °C | |||
| ±(0,8% + 3) | |||
| " | |||
| | |||
|- | |||
| °F | |||
| | |||
| | |||
| | |||
| mq Temp., unit °C, mqflags - | |||
| Not tested yet. | |||
|- | |||
| Event count | |||
| | |||
| | |||
| | |||
| style="background: #FFFF00; "| mq V, unit V, mqflags AC,RMS,REL | |||
| Counts events >= 2500 digits that last >=2s with a break >=2s. To activate, switch to voltage ACDC, press yellow button twice. Press up to 4 times to cycle parameters, press long to stop. See manual. Not useful remotely because the current measurement value is sent, not the counter. | |||
|- | |||
| Stop Watch | |||
| | |||
| 100 ns | |||
| | |||
| style="background: #ff0000; "| (Time data) | |||
| Max. 100 Min. To activate, select a 3V-1000V≃ range manually, press yellow button, press AUTO for start/stop, DATA for reset. Delivers only invalid data ("Overflow") via IR on [[User:Matthias Heidbrink | Matthias']] test device. | |||
|- | |||
| Data Hold | |||
| | |||
| | |||
| | |||
| style="background: #ff0000; "| mqflags += HOLD | |||
| Disabled while IR interface is active on [[User:Matthias Heidbrink | Matthias']] test device. | |||
|- | |||
| MIN | |||
| | |||
| | |||
| | |||
| style="background: #ff0000; "| mqflags += MIN | |||
| Influences only display, via IR flag is set, but actual current value is delivered. | |||
|- | |||
| MAX | |||
| | |||
| | |||
| | |||
| style="background: #ff0000; "| mqflags += MAX | |||
| Influences only display, via IR flag is set, but actual current value is delivered. | |||
|- | |||
|} | |||
The column "sigrok result" contains in short form what the driver generates for the resp. data type. | |||
[[Category:Device]] | [[Category:Device]] | ||
[[Category:Multimeter]] | [[Category:Multimeter]] | ||
[[Category:In progress]] | [[Category:In progress]] |
Revision as of 12:08, 3 November 2013
Status | in progress |
---|---|
Counts | 310000 |
IEC 61010-1 | CAT III 1000 V, CAT IV 600 V |
Connectivity | Infrared (RS232) |
Measurements | voltage, current, resistance, capacitance, temperature, diode, continuity |
Features | autorange, data hold, bargraph, backlight |
Website | [1] |
The Siemens B1105 is a 310000 counts 1000V CAT III/600 V CAT IV handheld digital multimeter with IR (RS232) connectivity.
It is an OEM version of the Gossen Metrawatt Metrahit 28S. The only differences are the type, logo and color (grey-blue print on knob, grey-blue holster) and the voltage rating. See Gossen Metrawatt Metrahit 28S for further information.
Photos
Measurement functions and ranges
Note: Work in progress, currently simply wrong!
Measurement function | Range | Resolution | Accuracy | sigrok result | Remarks |
---|---|---|---|---|---|
V= | 300 mV | 1 μV | ±(0,02% + 0,010% + 5) | mq V, unit V, mqflags DC | |
3 V | 10 μV | ±(0,02% + 0,005% + 5) | " | ||
30 V | 100 μV | " | |||
300 V | 1 mV | " | |||
1000 V | 10 mV | " | |||
V≈ | 300 mV | 10 µV | ±(0,5% + 30) (> 500 Digit)) | mq V, unit V, mqflags AC,RMS | |
3 V | 100 μV | ±(0,2% + 30) | " | ||
30 V | 100 μV | " | |||
300 V | 1 mV | " | |||
1000 V | 10 mV | " | |||
V≃ | 300 mV | 10 μV | ±(0,5% + 30) (> 500 Digit)) | mq V, unit V, mqflags AC,DC,RMS | |
3 V | 100 μV | ±(0,2% + 30) | " | ||
30 V | 100 μV | " | |||
300 V | 1 mV | " | |||
1000 V | 10 mV | " | |||
V db≈ | -48 - 63 dB | 0,01 dB | mq V, unit V, mqflags AC,RMS | Actually the 5 AC ranges, see there for accuracy. | |
V db≈ rel | -48 - 63 dB | 0,01 dB | mq V, unit V, mqflags AC,RMS | Actually the 5 AC ranges, see there for accuracy. To activate, press yellow button in dB mode, then do second measurement. Cannot be distinguished from non-rel measurement remotely. | |
A= | 300.00 µA | 1 nA | mq Curr., unit A, mqflags DC | ||
3.0000 mA | 10 nA | " | |||
30.000 mA | 100 nA | " | |||
300.00 mA | 1 µA | " | |||
3.0000 A | 100 µA | " | |||
10.000 A | 1 mA | " | |||
A≃ | 300.00 µA | 10 nA | mq Curr., unit A, mqflags AC,DC,RMS | ||
3.0000 mA | 100 nA | " | |||
30.000 mA | 1 µA | " | |||
300.00 mA | 10 µA | " | |||
3.0000 A | 100 µA | " | |||
10.000 A | 1 mA | " | |||
Ω | 300.00 Ω | 1 mΩ | mq Res., unit Ω, mqflags - | ||
3.0000 kΩ | 10 mΩ | " | |||
30.000 kΩ | 100 mΩ | " | |||
300.00 kΩ | 1 Ω | " | |||
3.0000 MΩ | 10 Ω | " | |||
30.000 MΩ | 100 Ω | " | |||
Ohm/Cont. | 300 Ω | 100 mΩ | mq Res., unit Ω, mqflags - | ||
Diode/Cont. | 300 mV | 100 µV | mq V, unit V, mqflags DC,DIODE | ||
Diode | 3 V | 100 µV | mq V, unit V, mqflags DC,DIODE | ||
Cap. F | 3.000 nF | 1 pF | mq Cap., unit F, mqflags - | ||
30.00 nF | 10 pF | " | |||
300.0 nF | 100 pF | " | |||
3.000 µF | 1 nF | " | |||
30.00 µF | 10 nF | " | |||
300.0 µF | 100 nF | ±(5.0% + 6) | " | ||
3.000 mF | 1 µF | " | |||
30.00 mF | 1 µF | " | |||
Hz≈, ≃ | 300.00 Hz | 0.001 Hz | ±(0.05% + 1) | mq Freq., unit Hz, mqflags AC,DC | To activate in AC, press yellow button twice. |
3.0000 kHz | 0.01 Hz | " | |||
300.000 kHz | 1 Hz | " | |||
°C | Pt100/Pt1000 –200.0 - +100.0 °C |
0.1 °C | ±(0,5 K + 3) | mq Temp., unit °C, mqflags - | |
Pt100/Pt1000 +100.0 - 850.0 °C |
±(0,2% + 3) | mq Temp., unit °C, mqflags - | |||
K NiCr-Ni -270-+1372,0 °C | ±(0,7% + 3) | " | |||
J Fe-CuNi -210 - +1200.0 °C | ±(0,8% + 3) | " | |||
°F | mq Temp., unit °C, mqflags - | Not tested yet.
| |||
Event count | mq V, unit V, mqflags AC,RMS,REL | Counts events >= 2500 digits that last >=2s with a break >=2s. To activate, switch to voltage ACDC, press yellow button twice. Press up to 4 times to cycle parameters, press long to stop. See manual. Not useful remotely because the current measurement value is sent, not the counter. | |||
Stop Watch | 100 ns | (Time data) | Max. 100 Min. To activate, select a 3V-1000V≃ range manually, press yellow button, press AUTO for start/stop, DATA for reset. Delivers only invalid data ("Overflow") via IR on Matthias' test device. | ||
Data Hold | mqflags += HOLD | Disabled while IR interface is active on Matthias' test device. | |||
MIN | mqflags += MIN | Influences only display, via IR flag is set, but actual current value is delivered. | |||
MAX | mqflags += MAX | Influences only display, via IR flag is set, but actual current value is delivered. |
The column "sigrok result" contains in short form what the driver generates for the resp. data type.