Difference between revisions of "Norma DM950"

From sigrok
Jump to navigation Jump to search
(→‎Interface: Added power connector size info)
Line 113: Line 113:


The RS232 interface is optional. All models were available with or without RS232 interface. The RS232 interface was also available separately as an accessory.
The RS232 interface is optional. All models were available with or without RS232 interface. The RS232 interface was also available separately as an accessory.
These Norma devices use the term ''RMS'' for correct measurement of possibly non-sinusiod AC and ''TRMS'' for correct measurement of mixed-mode (AC+DC) voltages and currents.


== Hardware ==
== Hardware ==

Revision as of 06:09, 21 October 2013

Norma DM950
Norma dm950.png
Status in progress
Counts 21000
IEC 61010-1 CAT II 600 V, CAT III 300 V
Connectivity RS232, proprietary cable
Measurements voltage, current, resistance, capacitance, temperature, diode, continuity
Features autorange, data hold, bargraph, (backlight)
Website set.nl

The Norma DM950 is a 21000 counts, handheld digital multimeter with RS232 connectivity.

Models

The LEM Norma DM9x0 series of multimeters are 21000 count CAT II/CAT III hand multimeters with optional RS232 connectivity. The models differ by precision and feature set.

They were also sold as OEM devices by Siemens. The manufacturer was a company named Norma, later LEM Norma in Austria.

It seems that the same device series was sold with two different names and designs:

  • Norma Normameter 910..950 (red company sign and bumper; earlier)
  • LEM Norma DM910..950 (turquoise company sign and bumper, see pics; newer)
Norma LEM Norma Siemens OEM Precision (DC) Backlight RMS TRMS R 20 MΩ R 200 MΩ U AC 1 kΩ/V I 10A f C
Normameter 910 Norma DM910 B1024 ±(0.5% +2d)
Normameter 920 Norma DM920 B1025 ±(0.5% +2d)
Normameter 930 Norma DM930 B1026 ±(0.3% +3d/+2d)
Normameter 940 Norma DM940 B1027 ±(0.1% +3d/+2d)
Normameter 950 Norma DM950 B1028 ±(0.06% +3d/+2d)

The RS232 interface is optional. All models were available with or without RS232 interface. The RS232 interface was also available separately as an accessory.

These Norma devices use the term RMS for correct measurement of possibly non-sinusiod AC and TRMS for correct measurement of mixed-mode (AC+DC) voltages and currents.

Hardware

Multimeter

Interface

The RS232 interface consists of the following components:

  1. Interface board (installed inside of multimeter, based on a NEC 75P116 microcontroller)
  2. Printer cable 5-pin to DB9
  3. PC adaptor (null-modem plug with slightly unusal pinout)

The RS232 interface is designed to allow a direct connection to a serial printer or data recorder. Therefore an adaptor is required to connect it to a PC.

Comm parameters (PC): DTR on, RTS off, RTS/CTS flow control (!).

If the interface cannot be powered from RS232 (e.g. RS232-USB converter with too low voltage), switching on the multimeter with the FUNC button pressed powers the interface from multimeter.

The interface board contains a battery-buffered clock and NVRAM memory.

The printer cable has got a power connector (10-16 V DC, 50 mA; coaxial power connector 5,5/2,1 mm) to power the multimeter and the interface externally. Opening the plug on multimeter side should be avoided, if possible, because the nut on the opposite to the screw is loose beneath the rubber bumper and it can be a bit tricky to get it closed again without removing the glued bumper.

Photos

Multimeter:

RS232 cable:

Protocol

Printer Mode

TODO.

PC communication mode

The multimeter supports about 50 commands, most functions that do not require changing the position of the rotary switch can be remote controlled.

Command Reply Example Remarks
IDN? Device id and rev., interface rev. 1834 065 V1.06,IF V1.02\r
DATA?\n 27 chars ASCII + \r R␣␣␣␣␣␣␣␣␣␣␣␣␣149.10E+3␣OHM\r Similar to common 14 char protocol, but more verbose. ␣ means space.
STATUS?\n 20 hex chars + \r 03030024913100400040\r Binary data, value + status flags.
... .. .... More to do.