Difference between revisions of "Brymen BM525s"

From sigrok
Jump to navigation Jump to search
(photo nits)
(recordings are supported)
 
Line 15: Line 15:


The DMM can record up to 87000 (single display) or 43500 (dual display) measurements in up to 999 "pages" (recording sessions, limited to one of the meter's functions) at a rate between 20/s and one every 600s.
The DMM can record up to 87000 (single display) or 43500 (dual display) measurements in up to 999 "pages" (recording sessions, limited to one of the meter's functions) at a rate between 20/s and one every 600s.
The current mainline driver supports live readings, but support for retrieval of recorded data is currently missing.


== Photos ==
== Photos ==
Line 37: Line 35:
Live readings require one HID report to request a measurement, and results in three HID reports of response data which carry 24 bytes of information. Payload consists of bitmaps for LCD segments, which naturally fit the serial-dmm implementation. Current mainline fully supports live readings in all the meter's functions.
Live readings require one HID report to request a measurement, and results in three HID reports of response data which carry 24 bytes of information. Payload consists of bitmaps for LCD segments, which naturally fit the serial-dmm implementation. Current mainline fully supports live readings in all the meter's functions.


The protocol for recordings differs. One HID report (different from the RTD request) initiates the transmission of data, two other HID reports either advance to the next data chunk, or repeat the most recently communicated data chunk. These packets are referred to as HEAD, NEXT, and CURR requests, and result in four HID reports of response data, which carry 24 bytes of information, and a checksum. The response data forms an adjacent stream of header information, and all measurements of all recording sessions. There is no means of random access, the retrieval of a single recording session involves the communication of all previously recorded information. Reading out recordings is currently not yet implemented in the mainline driver for this DMM.
The protocol for recordings differs. One HID report (different from the RTD request) initiates the transmission of data, two other HID reports either advance to the next data chunk, or repeat the most recently communicated data chunk. These packets are referred to as HEAD, NEXT, and CURR requests, and result in four HID reports of response data, which carry 24 bytes of information, and a checksum. The response data forms an adjacent stream of header information, and all measurements of all recording sessions. There is no means of random access, the retrieval of a single recording session involves the communication of all previously recorded information. Reading out recordings is implemented but has not seen much testing.


== Resources ==
== Resources ==

Latest revision as of 20:38, 22 October 2020

Brymen BM525s
Bm525s-mugshot.png
Status supported
Source code serial-dmm
Counts 10000
IEC 61010-1 CAT IV (1000V)
Connectivity USB
Measurements voltage, current, resistance, capacitance, diode, continuity, conductivity, frequency, duty cycle, temperature
Features autorange, data hold, crest, backlight, true-RMS, logging meter
Website brymen.com

The Brymen BM525s is a 10000 counts, dual display, CAT IV (1000V), handheld logging digital multimeter with USB connectivity. See the BU-86x adapter.

The DMM can record up to 87000 (single display) or 43500 (dual display) measurements in up to 999 "pages" (recording sessions, limited to one of the meter's functions) at a rate between 20/s and one every 600s.

Photos

Cable:

See BU-86x (USB HID).

Protocol

The vendor documents both protocols for live readings (referred to as "real-time download" RTD) as well as recordings (referred to as "memory data sets" xMD).

Live readings require one HID report to request a measurement, and results in three HID reports of response data which carry 24 bytes of information. Payload consists of bitmaps for LCD segments, which naturally fit the serial-dmm implementation. Current mainline fully supports live readings in all the meter's functions.

The protocol for recordings differs. One HID report (different from the RTD request) initiates the transmission of data, two other HID reports either advance to the next data chunk, or repeat the most recently communicated data chunk. These packets are referred to as HEAD, NEXT, and CURR requests, and result in four HID reports of response data, which carry 24 bytes of information, and a checksum. The response data forms an adjacent stream of header information, and all measurements of all recording sessions. There is no means of random access, the retrieval of a single recording session involves the communication of all previously recorded information. Reading out recordings is implemented but has not seen much testing.

Resources