Difference between revisions of "Gossen Metrawatt Metrahit 18S"
Jump to navigation
Jump to search
m (→Measurement functions and ranges: Cosmetics.) |
|||
(19 intermediate revisions by the same user not shown) | |||
Line 1: | Line 1: | ||
{{Infobox multimeter | {{Infobox multimeter | ||
| image = | | image = [[File:Gossen Metrawatt Metrahit 18S small.png|180px]] | ||
| name = [[Gossen Metrawatt]] Metrahit 18S | | name = [[Gossen Metrawatt]] Metrahit 18S | ||
| status = | | status = supported | ||
| source_code_dir = | | source_code_dir = gmc-mh-1x-2x | ||
| counts = 31000 | | counts = 31000 | ||
| categories = CAT III 1000 V, CAT IV 600 V | | categories = CAT III 1000 V, CAT IV 600 V | ||
| connectivity = Infrared (RS232) | | connectivity = Infrared (RS232) | ||
| measurements = voltage, current, resistance, capacitance, temperature, diode, continuity | | measurements = voltage, current, resistance, capacitance, temperature, diode, continuity | ||
| features = autorange, data hold, bargraph | | features = autorange, data hold, bargraph | ||
| website = [http:// | | <!-- website = [http://...] --> | ||
}} | }} | ||
Line 16: | Line 16: | ||
== Hardware == | == Hardware == | ||
* '''Battery''': ''9 V IEC 6 F 22'' or ''IEC 6 LR 61'' or corresponding rechargeable battery | |||
** FF (UR) 1,6 A/ | * '''Fuse''': | ||
** | ** Current 300 mA: FF (UR) 1,6 A/500G; 6,3x32 mm | ||
** Current 3A, 10A: | |||
*** Siba Ultrarapid 16A/600 V≈ 10mm x 38 mm | |||
*** Littlefuse KLK 15A/600 V≈ 13/32” / 1 1/2” | |||
*** Bussmann KLK 15A/600 V≈ 13/32” / 1 1/2” | |||
* TODO. | * TODO. | ||
== Photos == | == Photos == | ||
<gallery> | <gallery> | ||
File: | File:metrahit_18S_1.jpg|<small>Package</small> | ||
File: | File:metrahit_18S_2.jpg|<small>Inside</small> | ||
File:metrahit_18S_4.jpg|<small>Display</small> | |||
File:metrahit_18S_3.jpg|<small>IR Interface</small> | |||
File: | |||
File: | |||
</gallery> | </gallery> | ||
== Measurement functions and ranges == | == Measurement functions and ranges == | ||
Line 54: | Line 50: | ||
| 300 mV | | 300 mV | ||
| 10 μV | | 10 μV | ||
| ±(0 | | ±(0.05% + 3) | ||
| mq V, unit V, mqflags DC | | mq V, unit V, mqflags DC | ||
| | | Input Impendance >10 GΩ | ||
|- | |- | ||
| 3 V | | 3 V | ||
| 100 μV | | 100 μV | ||
| rowspan="4" | ±(0 | | rowspan="4" | ±(0.05% + 3) | ||
| " | | " | ||
| | | | ||
Line 67: | Line 63: | ||
|- | |- | ||
| 30 V | | 30 V | ||
| | | 1 mV | ||
| " | | " | ||
| | | | ||
Line 73: | Line 69: | ||
|- | |- | ||
| 300 V | | 300 V | ||
| | | 10 mV | ||
| " | | " | ||
| | | | ||
Line 79: | Line 75: | ||
|- | |- | ||
| 1000 V | | 1000 V | ||
| | | 100 mV | ||
| " | | " | ||
| | | | ||
Line 87: | Line 83: | ||
| 300 mV | | 300 mV | ||
| 10 µV | | 10 µV | ||
| ±(0 | | ±(0.5% + 30) (> 500 Digit)) | ||
| mq V, unit V, mqflags AC,RMS | | mq V, unit V, mqflags AC,RMS | ||
| | | | ||
Line 94: | Line 90: | ||
| 3 V | | 3 V | ||
| 100 μV | | 100 μV | ||
| rowspan="4" | ±(0 | | rowspan="4" | ±(0.5% + 30) (> 300 Digit) | ||
| " | | " | ||
| | | | ||
Line 100: | Line 96: | ||
|- | |- | ||
| 30 V | | 30 V | ||
| | | 1 mV | ||
| " | | " | ||
| | | | ||
Line 106: | Line 102: | ||
|- | |- | ||
| 300 V | | 300 V | ||
| | | 10 mV | ||
| " | | " | ||
| | | | ||
Line 112: | Line 108: | ||
|- | |- | ||
| 1000 V | | 1000 V | ||
| | | 100 mV | ||
| " | | " | ||
| | | | ||
Line 120: | Line 116: | ||
| 300 mV | | 300 mV | ||
| 10 μV | | 10 μV | ||
| ±(0 | | ±(0.5% + 30) (> 500 Digit)) | ||
| mq V, unit V, mqflags AC,DC,RMS | | mq V, unit V, mqflags AC,DC,RMS | ||
| | | | ||
Line 127: | Line 123: | ||
| 3 V | | 3 V | ||
| 100 μV | | 100 μV | ||
| rowspan="4" | ±(0 | | rowspan="4" | ±(0.5% + 30) (> 300 Digit) | ||
| " | | " | ||
| | | | ||
Line 133: | Line 129: | ||
|- | |- | ||
| 30 V | | 30 V | ||
| | | 1 mV | ||
| " | | " | ||
| | | | ||
Line 139: | Line 135: | ||
|- | |- | ||
| 300 V | | 300 V | ||
| | | 10 mV | ||
| " | | " | ||
| | | | ||
Line 145: | Line 141: | ||
|- | |- | ||
| 1000 V | | 1000 V | ||
| | | 100 mV | ||
| " | | " | ||
| | | | ||
|- | |||
| V db≈ | |||
| -48 - 63 dB | |||
| 0.01 dB | |||
| | |||
| mq V, unit V, mqflags AC,RMS | |||
| Actually 5 ranges. | |||
|- | |||
| V db≈ rel | |||
| -48 - 63 dB | |||
| 0.01 dB | |||
| | |||
| mq V, unit V, mqflags AC,RMS | |||
| Actually 5 ranges. To activate, press yellow button in dB mode, then do second measurement. Cannot be distinguished from non-rel measurement remotely. | |||
|- | |- | ||
Line 172: | Line 184: | ||
|- | |- | ||
| | | 300.00 mA | ||
| 10 µA | | 10 µA | ||
| | | | ||
Line 179: | Line 191: | ||
|- | |- | ||
| | | 3.0000 A | ||
| 100 µA | | 100 µA | ||
| | | | ||
Line 215: | Line 227: | ||
|- | |- | ||
| | | 300.00 mA | ||
| 10 µA | | 10 µA | ||
| | | | ||
Line 222: | Line 234: | ||
|- | |- | ||
| | | 3.0000 A | ||
| 100 µA | | 100 µA | ||
| | | | ||
Line 398: | Line 410: | ||
| | | | ||
| | | | ||
| | | style="background: #FFFF00; "| mq V, unit V, mqflags AC,RMS,REL | ||
| | | Counts events >= 2500 digits that last >=2s with a break >=2s. To activate, switch to voltage ACDC, press yellow button twice. Press up to 4 times to cycle parameters, press long to stop. See manual. Not useful remotely because the current measurement value is sent, not the counter. | ||
|- | |- | ||
Line 406: | Line 418: | ||
| | | | ||
| | | | ||
| | | style="background: #ff0000; "| (Time data) | ||
| To activate, select 3V-1000V range manually, press yellow button, press AUTO for start/stop, DATA for reset. Delivers invalid data ("Overflow") via IR on [[User:Matthias Heidbrink | Matthias']] device. | | To activate, select a 3V-1000V range manually, press yellow button, press AUTO for start/stop, DATA for reset. Delivers only invalid data ("Overflow") via IR on [[User:Matthias Heidbrink | Matthias']] device. | ||
|- | |- | ||
Line 414: | Line 426: | ||
| | | | ||
| | | | ||
| mqflags += HOLD | | style="background: #ff0000; "| mqflags += HOLD | ||
| Disabled while IR interface is active. | | Disabled while IR interface is active on [[User:Matthias Heidbrink | Matthias']] device. | ||
|- | |- | ||
Line 422: | Line 434: | ||
| | | | ||
| | | | ||
| mqflags += MIN | | style="background: #ff0000; "| mqflags += MIN | ||
| | | Influences only display, via IR flag is set, but actual current value is delivered. | ||
|- | |- | ||
Line 430: | Line 442: | ||
| | | | ||
| | | | ||
| mqflags += MAX | | style="background: #ff0000; "| mqflags += MAX | ||
| | | Influences only display, via IR flag is set, but actual current value is delivered. | ||
|- | |- | ||
Line 440: | Line 452: | ||
[[Category:Device]] | [[Category:Device]] | ||
[[Category:Multimeter]] | [[Category:Multimeter]] | ||
[[Category: | [[Category:Supported]] |
Latest revision as of 20:29, 21 September 2016
Status | supported |
---|---|
Source code | gmc-mh-1x-2x |
Counts | 31000 |
IEC 61010-1 | CAT III 1000 V, CAT IV 600 V |
Connectivity | Infrared (RS232) |
Measurements | voltage, current, resistance, capacitance, temperature, diode, continuity |
Features | autorange, data hold, bargraph |
The Gossen Metrawatt Metrahit 18S is a 31000 counts handheld digital multimeter with IR (RS232) connectivity.
Hardware
- Battery: 9 V IEC 6 F 22 or IEC 6 LR 61 or corresponding rechargeable battery
- Fuse:
- Current 300 mA: FF (UR) 1,6 A/500G; 6,3x32 mm
- Current 3A, 10A:
- Siba Ultrarapid 16A/600 V≈ 10mm x 38 mm
- Littlefuse KLK 15A/600 V≈ 13/32” / 1 1/2”
- Bussmann KLK 15A/600 V≈ 13/32” / 1 1/2”
- TODO.
Photos
Measurement functions and ranges
Measurement function | Range | Resolution | Accuracy | sigrok result | Remarks |
---|---|---|---|---|---|
V= | 300 mV | 10 μV | ±(0.05% + 3) | mq V, unit V, mqflags DC | Input Impendance >10 GΩ |
3 V | 100 μV | ±(0.05% + 3) | " | ||
30 V | 1 mV | " | |||
300 V | 10 mV | " | |||
1000 V | 100 mV | " | |||
V≈ | 300 mV | 10 µV | ±(0.5% + 30) (> 500 Digit)) | mq V, unit V, mqflags AC,RMS | |
3 V | 100 μV | ±(0.5% + 30) (> 300 Digit) | " | ||
30 V | 1 mV | " | |||
300 V | 10 mV | " | |||
1000 V | 100 mV | " | |||
V≃ | 300 mV | 10 μV | ±(0.5% + 30) (> 500 Digit)) | mq V, unit V, mqflags AC,DC,RMS | |
3 V | 100 μV | ±(0.5% + 30) (> 300 Digit) | " | ||
30 V | 1 mV | " | |||
300 V | 10 mV | " | |||
1000 V | 100 mV | " | |||
V db≈ | -48 - 63 dB | 0.01 dB | mq V, unit V, mqflags AC,RMS | Actually 5 ranges. | |
V db≈ rel | -48 - 63 dB | 0.01 dB | mq V, unit V, mqflags AC,RMS | Actually 5 ranges. To activate, press yellow button in dB mode, then do second measurement. Cannot be distinguished from non-rel measurement remotely. | |
A= | 300.00 µA | 100 nA | mq Curr., unit A, mqflags DC | ||
3.0000 mA | 1 µA | " | |||
30.000 mA | 1 µA | " | |||
300.00 mA | 10 µA | " | |||
3.0000 A | 100 µA | " | |||
10.000 A | 1 mA | " | |||
A≃ | 300.00 µA | 100 nA | mq Curr., unit A, mqflags AC,DC,RMS | ||
3.0000 mA | 1 µA | " | |||
30.000 mA | 1 µA | " | |||
300.00 mA | 10 µA | " | |||
3.0000 A | 100 µA | " | |||
10.000 A | 1 mA | " | |||
Ω | 300.00 Ω | 10 mΩ | mq Res., unit Ω, mqflags - | ||
3.0000 kΩ | 100 mΩ | " | |||
30.000 kΩ | 1 Ω | " | |||
300.00 kΩ | 10 Ω | " | |||
3.0000 MΩ | 100 Ω | " | |||
30.000 MΩ | 1 kΩ | " | |||
Diode/Cont. | 3.0000 V– | 1 mV | mq V, unit V, mqflags DC,DIODE | Diode and cont. are identical on this device, just without beeper for diode. | |
Cap. F | 3.000 nF | 1 pF | mq Cap., unit F, mqflags - | ||
30.00 nF | 10 pF | " | |||
300.0 nF | 100 pF | " | |||
3.000 µF | 1 nF | " | |||
30.00 µF | 10 nF | " | |||
300.0 µF | 100 nF | ±(5.0% + 6) | " | ||
3.000 mF | 1 µF | " | |||
10.00 mF | 10 µF | " | |||
Hz≃ | 300.00 Hz | 0.01 Hz | ±(0.1% + 3) | mq Freq., unit Hz, mqflags AC,DC | |
3.0000 kHz | 0.1 Hz | " | |||
30.000 kHz | 1 Hz | " | |||
100.000 kHz | 10 Hz | " | |||
°C | Pt100 –200.0 - +100.0 °C | 0.1 °C | 2 Kelvin + 5 | mq Temp., unit °C, mqflags - | |
Pt100 +100.0 - +800.0 °C | ±(1.0% + 5) | " | |||
Pt1000 –100.0 - +100.0 °C | 2 Kelvin + 5 | " | |||
Pt1000 +100.0 - +800.0 °C | ±(1.0% + 5) | " |
| ||
Event count | mq V, unit V, mqflags AC,RMS,REL | Counts events >= 2500 digits that last >=2s with a break >=2s. To activate, switch to voltage ACDC, press yellow button twice. Press up to 4 times to cycle parameters, press long to stop. See manual. Not useful remotely because the current measurement value is sent, not the counter. | |||
Stop Watch | (Time data) | To activate, select a 3V-1000V range manually, press yellow button, press AUTO for start/stop, DATA for reset. Delivers only invalid data ("Overflow") via IR on Matthias' device. | |||
Data Hold | mqflags += HOLD | Disabled while IR interface is active on Matthias' device. | |||
MIN | mqflags += MIN | Influences only display, via IR flag is set, but actual current value is delivered. | |||
MAX | mqflags += MAX | Influences only display, via IR flag is set, but actual current value is delivered. |
The column "sigrok result" contains in short form what the driver generates for the resp. data type.