Difference between revisions of "Siemens B1105"

From sigrok
Jump to navigation Jump to search
Line 155: Line 155:
| -48 - 63 dB
| -48 - 63 dB
| 0,01 dB
| 0,01 dB
|  
| ± 0,1 dB
| mq V, unit V, mqflags AC,RMS
| mq V, unit V, mqflags AC,RMS
| Actually the 5 AC ranges, see there for accuracy.
| Actually the 5 AC ranges, see there for accuracy.
Line 163: Line 163:
| -48 - 63 dB
| -48 - 63 dB
| 0,01 dB
| 0,01 dB
|  
| ± 0,1 dB
| mq V, unit V, mqflags AC,RMS
| mq V, unit V, mqflags AC,RMS
| Actually the 5 AC ranges, see there for accuracy. To activate, press yellow button in dB mode, then do second measurement. Cannot be distinguished from non-rel measurement remotely.
| Actually the 5 AC ranges, see there for accuracy. To activate, press yellow button in dB mode, then do second measurement. Cannot be distinguished from non-rel measurement remotely.
Line 171: Line 171:
| 300.00 µA
| 300.00 µA
| 1 nA
| 1 nA
|  
| 0,05% + 0,02% + 5
| mq Curr., unit A, mqflags DC
| mq Curr., unit A, mqflags DC
|  
|  
Line 178: Line 178:
| 3.0000 mA
| 3.0000 mA
| 10 nA
| 10 nA
|  
| 0,05% + 0,01% + 5
| "
| "
|  
|  
Line 185: Line 185:
| 30.000 mA
| 30.000 mA
| 100 nA
| 100 nA
|  
| 0,05% + 0,01% + 5
| "
| "
|  
|  
Line 192: Line 192:
| 300.00 mA
| 300.00 mA
| 1 µA
| 1 µA
|  
| 0,1% + 0,01% + 5
| "
| "
|  
|  
Line 199: Line 199:
| 3.0000 A
| 3.0000 A
| 100 µA
| 100 µA
|  
| 0,2% + 0,05% + 5
| "
| "
|  
|  
Line 206: Line 206:
| 10.000 A
| 10.000 A
| 1 mA
| 1 mA
|  
| 0,2% + 0,05% + 5
| "
| "
|  
|  
Line 214: Line 214:
| 300.00 µA
| 300.00 µA
| 10 nA
| 10 nA
|  
| rowspan="4" | 0,5% + 30
| mq Curr., unit A, mqflags AC,DC,RMS
| mq Curr., unit A, mqflags AC,DC,RMS
|  
|  
Line 221: Line 221:
| 3.0000 mA
| 3.0000 mA
| 100 nA
| 100 nA
|
| "
| "
|  
|  
Line 228: Line 227:
| 30.000 mA
| 30.000 mA
| 1 µA
| 1 µA
|
| "
| "
|  
|  
Line 235: Line 233:
| 300.00 mA
| 300.00 mA
| 10 µA
| 10 µA
|
| "
| "
|  
|  
Line 242: Line 239:
| 3.0000 A
| 3.0000 A
| 100 µA
| 100 µA
|  
| 0,7% + 30
| "
| "
|  
|  
Line 249: Line 246:
| 10.000 A
| 10.000 A
| 1 mA
| 1 mA
|  
| 0,5% + 30
| "
| "
|  
|  
Line 256: Line 253:
| 300.00 Ω
| 300.00 Ω
| 1 mΩ
| 1 mΩ
|  
| rowspan="4" | 0,05% + 0.01% + 5
| mq Res., unit Ω, mqflags -
| mq Res., unit Ω, mqflags -
|
|
Line 263: Line 260:
| 3.0000 kΩ
| 3.0000 kΩ
| 10 mΩ
| 10 mΩ
|
| "
| "
|  
|  
Line 270: Line 266:
| 30.000 kΩ
| 30.000 kΩ
| 100 mΩ
| 100 mΩ
|
| "
| "
|  
|  
Line 277: Line 272:
| 300.00 kΩ
| 300.00 kΩ
| 1 Ω
| 1 Ω
|
| "
| "
|  
|  
Line 284: Line 278:
| 3.0000 MΩ
| 3.0000 MΩ
| 10 Ω
| 10 Ω
|  
| 0,1% + 0.02% + 5
| "
| "
|  
|  
Line 291: Line 285:
| 30.000 MΩ
| 30.000 MΩ
| 100 Ω
| 100 Ω
|  
| 1% + 0.2% + 5
| "
| "
|  
|  
Line 299: Line 293:
| 300 Ω
| 300 Ω
| 100 mΩ
| 100 mΩ
|  
| 1% + 3%
| mq Res., unit Ω, mqflags -
| mq Res., unit Ω, mqflags -
|  
|  
Line 307: Line 301:
| 300 mV
| 300 mV
| 100 µV
| 100 µV
|  
| 0.2% + 3
| mq V, unit V, mqflags DC,DIODE
| mq V, unit V, mqflags DC,DIODE
|  
|  
Line 315: Line 309:
| 3 V
| 3 V
| 100 µV
| 100 µV
|  
| 0.2% + 3
| mq V, unit V, mqflags DC,DIODE
| mq V, unit V, mqflags DC,DIODE
|  
|  
Line 323: Line 317:
| 3.000 nF
| 3.000 nF
| 1 pF
| 1 pF
|  
| rowspan="5" | 1.0% + 0.2
| mq Cap., unit F, mqflags -
| mq Cap., unit F, mqflags -
|
|
Line 330: Line 324:
| 30.00 nF
| 30.00 nF
| 10 pF
| 10 pF
|
| "
| "
|  
|  
Line 337: Line 330:
| 300.0 nF
| 300.0 nF
| 100 pF
| 100 pF
|
| "
| "
|  
|  
Line 344: Line 336:
| 3.000 µF
| 3.000 µF
| 1 nF
| 1 nF
|
| "
| "
|  
|  
Line 351: Line 342:
| 30.00 µF
| 30.00 µF
| 10 nF
| 10 nF
|
| "
| "
|  
|  

Revision as of 22:03, 7 November 2013

Siemens B1105
Siemens B1105 small.png
Status in progress
Counts 310000
IEC 61010-1 CAT III 1000 V, CAT IV 600 V
Connectivity Infrared (RS232)
Measurements voltage, current, resistance, capacitance, temperature, diode, continuity
Features autorange, data hold
Website [1]

The Siemens B1105 is a 310000 counts 1000V CAT III/600 V CAT IV handheld digital multimeter with IR (RS232) connectivity.

It is an OEM version of the Gossen Metrawatt  Metrahit 28S. The only differences seem to be the type, logo and color (light grey-blue print on knob, light grey-blue holster) and the voltage rating.

Hardware

  • Power supply:
    • Battery: 2x 1,5 V IEC R 6, IEC LR 6 or corresponding NiCd rechargeable batteries
    • External power supply NA5/600 by Gossen Metrawatt, 5V, 600 mA
  • Fuse:
    • Current 300 mA: FF (UR) 1,6 A/1000 V AC/DC (10 kA) 6,3 mm x 32 mm
    • Current 3A/10A: FF (UR) 10 A/1000 V AC/DC (30 kA) 10 mm x 38 mm

Photos

Using the device with sigrok

The device is supported in Send Mode using a RS232 interface by the driver gmc-mh-sm-rs232 (in development) and will be detected as a Metrahit 28S. Set the multimeter to use the communication adaptor RS232.

The device might be not detected by the driver if it is in a mode where it sends data rarely, either with a huge send interval or with a measurement function that delivers a very low data rate, e.g. temperature measurements with K type sensor. Then it should be set to a mode with higher data rate during the initialisation of the driver.

Measurement functions and ranges

Measurement function Range Resolution Accuracy sigrok result Remarks
V= 300 mV 1 μV ±(0,02% + 0,010% + 5) mq V, unit V, mqflags DC
3 V 10 μV ±(0,02% + 0,005% + 5) "
30 V 100 μV "
300 V 1 mV "
1000 V 10 mV "
V≈ 300 mV 10 µV ±(0,5% + 30) (> 500 Digit)) mq V, unit V, mqflags AC,RMS
3 V 100 μV ±(0,2% + 30) "
30 V 100 μV "
300 V 1 mV "
1000 V 10 mV "
V≃ 300 mV 10 μV ±(0,5% + 30) (> 500 Digit)) mq V, unit V, mqflags AC,DC,RMS
3 V 100 μV ±(0,2% + 30) "
30 V 100 μV "
300 V 1 mV "
1000 V 10 mV "
V db≈ -48 - 63 dB 0,01 dB ± 0,1 dB mq V, unit V, mqflags AC,RMS Actually the 5 AC ranges, see there for accuracy.
V db≈ rel -48 - 63 dB 0,01 dB ± 0,1 dB mq V, unit V, mqflags AC,RMS Actually the 5 AC ranges, see there for accuracy. To activate, press yellow button in dB mode, then do second measurement. Cannot be distinguished from non-rel measurement remotely.
A=  300.00 µA 1 nA 0,05% + 0,02% + 5 mq Curr., unit A, mqflags DC
3.0000 mA 10 nA 0,05% + 0,01% + 5 "
30.000 mA 100 nA 0,05% + 0,01% + 5 "
300.00 mA 1 µA 0,1% + 0,01% + 5 "
3.0000 A 100 µA 0,2% + 0,05% + 5 "
10.000 A 1 mA 0,2% + 0,05% + 5 "
A≃  300.00 µA 10 nA 0,5% + 30 mq Curr., unit A, mqflags AC,DC,RMS
3.0000 mA 100 nA "
30.000 mA 1 µA "
300.00 mA 10 µA "
3.0000 A 100 µA 0,7% + 30 "
10.000 A 1 mA 0,5% + 30 "
Ω 300.00 Ω 1 mΩ 0,05% + 0.01% + 5 mq Res., unit Ω, mqflags -
3.0000 kΩ 10 mΩ "
30.000 kΩ 100 mΩ "
300.00 kΩ 1 Ω "
3.0000 MΩ 10 Ω 0,1% + 0.02% + 5 "
30.000 MΩ 100 Ω 1% + 0.2% + 5 "
Ohm/Cont. 300 Ω 100 mΩ 1% + 3% mq Res., unit Ω, mqflags -
Diode/Cont. 300 mV 100 µV 0.2% + 3 mq V, unit V, mqflags DC,DIODE
Diode 3 V 100 µV 0.2% + 3 mq V, unit V, mqflags DC,DIODE
Cap. F 3.000 nF 1 pF 1.0% + 0.2 mq Cap., unit F, mqflags -
30.00 nF 10 pF "
300.0 nF 100 pF "
3.000 µF 1 nF "
30.00 µF 10 nF "
300.0 µF 100 nF ±(5.0% + 6) "
3.000 mF 1 µF "
30.00 mF 1 µF "
Hz≈, ≃ 300.00 Hz 0.001 Hz ±(0.05% + 1) mq Freq., unit Hz, mqflags AC,DC To activate in AC, press yellow button twice.
3.0000 kHz 0.01 Hz "
300.000 kHz 1 Hz "
°C Pt100/Pt1000
–200.0 - +100.0 °C
0.1 °C ±(0,5 K + 3) mq Temp., unit °C, mqflags -
Pt100/Pt1000
+100.0 - 850.0 °C
±(0,2% + 3) mq Temp., unit °C, mqflags -
K NiCr-Ni -270-+1372,0 °C ±(0,7% + 3) "
J Fe-CuNi -210 - +1200.0 °C ±(0,8% + 3) " Not tested yet due to lack of corresponding sensor.
°F " " " mq Temp., unit °F, mqflags - Details and ranges like °C.


Event count mq V, unit V, mqflags AC,RMS,REL Counts events >= 2500 digits that last >=2s with a break >=2s. To activate, switch to voltage ACDC, press yellow button twice. Press up to 4 times to cycle parameters, press long to stop. See manual. Not useful remotely because the current measurement value is sent, not the counter.
Stop Watch 100 ns (Time data) Max. 100 Min. To activate, select a 3V-1000V≃ range manually, press yellow button, press AUTO for start/stop, DATA for reset. Using the stop watch delivers only voltage data and sometimes seems to bring the device into a mode where it even sends invalid data packets.
Data Hold mqflags += HOLD Disabled while IR interface is active on  Matthias' test device.
MIN mqflags += MIN Influences only display, via IR flag is set, but actual current value is delivered.
MAX mqflags += MAX Influences only display, via IR flag is set, but actual current value is delivered.

The column "sigrok result" contains in short form what the driver generates for the resp. data type.