Difference between revisions of "Gossen Metrawatt Metrahit 16I"

From sigrok
Jump to navigation Jump to search
Line 54: Line 54:
| 10 μV
| 10 μV
| ±(0,5% + 3)
| ±(0,5% + 3)
| sigrok result
| mq V, unit V, mqflags DC
| Not covered by autorange, use manual range
| Not covered by autorange, use manual range


Line 61: Line 61:
| 100 μV
| 100 μV
| ±(0,5% + 3)
| ±(0,5% + 3)
|
| "
| Not covered by autorange, use manual range
| Not covered by autorange, use manual range


Line 68: Line 68:
| 1 mV
| 1 mV
| rowspan="3" | ±(0,25% + 1)
| rowspan="3" | ±(0,25% + 1)
|  
| "
|
|


Line 74: Line 74:
| 30 V
| 30 V
| 10 mV
| 10 mV
|  
| "
|
|


Line 80: Line 80:
| 300 V
| 300 V
| 100 mV
| 100 mV
|  
| "
|
|


Line 87: Line 87:
| 1 V
| 1 V
| ±(0,35% + 1)
| ±(0,35% + 1)
|  
| "
|
|


Line 95: Line 95:
| 100 μV
| 100 μV
| rowspan="4" | ±(1.0% + 3) (> 10 Digit)
| rowspan="4" | ±(1.0% + 3) (> 10 Digit)
|  
| mq V, unit V, mqflags AC,RMS
|
|


Line 101: Line 101:
| 30 V
| 30 V
| 100 μV
| 100 μV
|  
| "
|
|


Line 107: Line 107:
| 300 V
| 300 V
| 1 mV
| 1 mV
|  
| "
|
|


Line 113: Line 113:
| 1000 V
| 1000 V
| 10 mV
| 10 mV
|  
| "
|
|


Line 121: Line 121:
| 100 μV
| 100 μV
| rowspan="4" | ±(1.0% + 3) (> 10 Digit)
| rowspan="4" | ±(1.0% + 3) (> 10 Digit)
|  
| mq V, unit V, mqflags AC,DC,RMS
|
|


Line 127: Line 127:
| 30 V
| 30 V
| 100 μV
| 100 μV
|  
| "
|
|


Line 133: Line 133:
| 300 V
| 300 V
| 1 mV
| 1 mV
|  
| "
|
|


Line 139: Line 139:
| 1000 V
| 1000 V
| 10 mV
| 10 mV
|  
| "
|
|


Line 147: Line 147:
| 10 mA
| 10 mA
| rowspan="2" | ±(2,5% + 3) (>10 Digit)
| rowspan="2" | ±(2,5% + 3) (>10 Digit)
|
| mq Curr., unit A, mqflags AC,DC,RMS
| rowspan="2" | Current clamp with 100 mV/A required, e.g. GMC WZ12B
| rowspan="2" | Current clamp with 100 mV/A required, e.g. GMC WZ12B


Line 153: Line 153:
| 100 A
| 100 A
| 100 mA
| 100 mA
|
| "


|-
|-
Line 309: Line 309:
|}
|}


The column "sigrok result" contains in short form what the driver generates for the resp. data type.


[[Category:Device]]
[[Category:Device]]
[[Category:Multimeter]]
[[Category:Multimeter]]
[[Category:In progress]]
[[Category:In progress]]

Revision as of 18:06, 2 November 2013

Gossen Metrawatt Metrahit 16I
Status in progress
Counts 3100 (31000)
IEC 61010-1 CAT III 1000 V, CAT IV 600 V
Connectivity Infrared (RS232)
Measurements voltage, current, resistance, capacitance, temperature, diode, continuity
Features autorange, data hold, bargraph
Website [1]

The Gossen Metrawatt Metrahit 16I is a 3100 counts handheld digital multimeter with IR (RS232) connectivity and insulation measurement (500/1000 V).

The resolution is one digit higher on the data interface.

Hardware

  • Fuse: None!
  • TODO.

Photos

Measurement functions and ranges

Measurement function Range Resolution Deviation sigriok result Remarks
V= 30 mV 10 μV ±(0,5% + 3) mq V, unit V, mqflags DC Not covered by autorange, use manual range
300 mV 100 μV ±(0,5% + 3) " Not covered by autorange, use manual range
3 V 1 mV ±(0,25% + 1) "
30 V 10 mV "
300 V 100 mV "
600 V 1 V ±(0,35% + 1) "
V≈ 3 V 100 μV ±(1.0% + 3) (> 10 Digit) mq V, unit V, mqflags AC,RMS
30 V 100 μV "
300 V 1 mV "
1000 V 10 mV "
V≃ 3 V 100 μV ±(1.0% + 3) (> 10 Digit) mq V, unit V, mqflags AC,DC,RMS
30 V 100 μV "
300 V 1 mV "
1000 V 10 mV "
A≈ (Clamp only)  30 A 10 mA ±(2,5% + 3) (>10 Digit) mq Curr., unit A, mqflags AC,DC,RMS Current clamp with 100 mV/A required, e.g. GMC WZ12B
100 A 100 mA "
Ω 30.00 Ω 10 mΩ ±(0,5% + 3)
300.0 Ω 100 mΩ
3.000 kΩ 1 Ω ±(0,4% + 1)
30.00 kΩ 10 Ω
300.0 kΩ 100 Ω
3.000 MΩ 1 kΩ ±(0,6% + 1)
30.00 MΩ 10 kΩ ±(2,0% + 1)
Diode/Cont. 3.0000 V– 1 mV ±(0,25% + 1)
Cap. F 30.00 nF 10 pF ±(1,0% + 3)
300.0 nF 100 pF
3.000 µF 1 nF
30.00 µF 10 nF ±(3,0% + 3)
Hz 300.00 Hz 0.01 Hz ±(0.1% + 3)
3.0000 kHz 0.1 Hz
30.000 kHz 1 Hz
100.000 kHz 10 Hz
°C Pt100 –200.0 - +200.0 °C 2 Kelvin + 5
+200.0 - +800.0 °C ±(1.0% + 5)
Pt1000 –100.0 - +200.0 °C 2 Kelvin + 5
+200.0 - +800.0 °C ±(1.0% + 5)
Event count
Stop Watch

The column "sigrok result" contains in short form what the driver generates for the resp. data type.