Difference between revisions of "Mooshim Engineering Mooshimeter"
m (Category Bluetooth added) |
|||
(9 intermediate revisions by 2 users not shown) | |||
Line 1: | Line 1: | ||
{{Infobox multimeter | {{Infobox multimeter | ||
| image = [[File: | | image = [[File:Mooshimeter_mugshot.png|180px]] | ||
| name = Mooshimeter | | name = Mooshimeter | ||
| status = | | status = supported | ||
| source_code_dir = | | source_code_dir = mooshimeter-dmm | ||
| counts = 24 bit | | counts = 24 bit | ||
| categories = CAT III (600V) | | categories = CAT III (600V) | ||
| connectivity = | | connectivity = BLE | ||
| measurements = voltage, resistance, diode, continuity, frequency, current, temperature | | measurements = voltage, resistance, diode, continuity, frequency, current, temperature | ||
| features = autorange, logging to SD card, current / voltage simultaneously | | features = autorange, logging to SD card, current / voltage simultaneously | ||
| website = https://moosh.im/mooshimeter/ | | website = [https://moosh.im/mooshimeter/ moosh.im] | ||
}} | }} | ||
The '''Mooshimeter''' is a 24-bit, two channel, CAT III (600V) remote access digital multimeter with Bluetooth Low Energy connectivity. | The '''Mooshim Engineering Mooshimeter''' (DMM-BLE-2X01A) is a 24-bit, two channel, CAT III (600V) remote access digital multimeter with Bluetooth Low Energy (BLE) connectivity. | ||
See [[Mooshim Engineering Mooshimeter/Info]] for more details (such as '''hcitool/gatttool''' output) about the device. | |||
== Hardware == | == Hardware == | ||
Line 18: | Line 20: | ||
'''Multimeter''': | '''Multimeter''': | ||
* BLE SoC: [http://www.ti.com/product/CC2540 TI CC2540] (custom firmware) | |||
* ADC: [http://www.ti.com/product/ADS1292 TI ADS1292] | |||
== Photos == | == Photos == | ||
Line 24: | Line 28: | ||
<gallery> | <gallery> | ||
File:Mooshimeter-2X01A.jpg | File:Mooshimeter-2X01A.jpg|<small>Device, top</small> | ||
File:Mooshimeter_Component_Side.jpg | File:Mooshimeter_Component_Side.jpg|<small>PCB, top</small> | ||
File:Mooshimeter_Connection_Side.jpg | File:Mooshimeter_Connection_Side.jpg|<small>PCB, bottom</small> | ||
File:Mooshimeter_BLE_SoC.jpg | File:Mooshimeter_BLE_SoC.jpg|<small>TI CC2540</small> | ||
File:Mooshimeter_ADC.jpg | File:Mooshimeter_ADC.jpg|<small>TI ADS1292</small> | ||
</gallery> | </gallery> | ||
'''Another teardown''': | |||
<gallery> | |||
File:Mooshimeter top.jpg|<small>Device, top</small> | |||
File:Mooshimeter bottom.jpg|<small>Device, bottom</small> | |||
File:Mooshimeter pcb top.jpg|<small>PCB, top</small> | |||
File:Mooshimeter pcb bottom.jpg|<small>PCB, bottom</small> | |||
File:Mooshimeter case.jpg|<small>Device, case</small> | |||
</gallery> | |||
== Usage == | |||
Example usage using [[sigrok-cli]]: | |||
<small> | |||
$ '''sigrok-cli --driver mooshimeter-dmm:conn=68-9e-19-cb-e1-71 --scan''' | |||
The following devices were found: | |||
mooshimeter-dmm - Mooshim Engineering Mooshimeter with 3 channels: CH1 CH2 P | |||
$ '''sigrok-cli --driver mooshimeter-dmm:conn=68-9e-19-cb-e1-71 --show''' | |||
Driver functions: | |||
Multimeter | |||
Scan options: | |||
conn | |||
mooshimeter-dmm - Mooshim Engineering Mooshimeter with 3 channels: CH1 CH2 P | |||
Supported configuration options: | |||
continuous: on, off | |||
limit_time: | |||
avg_samples: - supported values: | |||
32 | |||
64 (current) | |||
128 | |||
256 | |||
samplerate - supported samplerates: | |||
125 Hz | |||
250 Hz | |||
500 Hz | |||
1 kHz | |||
2 kHz | |||
4 kHz | |||
8 kHz | |||
channel_config: | |||
$ '''sigrok-cli --driver mooshimeter-dmm:conn=68-9e-19-cb-e1-71 --samples 4''' | |||
CH1: 0.00 A DC AUTO | |||
CH2: 0.00 V DC AUTO | |||
CH1: 0.00 A DC AUTO | |||
CH2: 0.00 V DC AUTO | |||
$ '''sigrok-cli --driver mooshimeter-dmm:conn=68-9e-19-cb-e1-71 -c channel_config="Aux;0.1/T" --samples 4''' | |||
CH1: -0.19 V DC | |||
CH2: 298.30 °C DC AUTO | |||
CH1: -0.19 V DC | |||
CH2: 298.30 °C DC AUTO | |||
</small> | |||
See also [[libsigrok]]'s [https://sigrok.org/gitweb/?p=libsigrok.git;a=blob;f=README.devices;h=a7e50a395302d9b5a6e9d351a138ea54dc5b8e0b;hb=HEAD#l480 README.devices] for more info. | |||
== Protocol == | == Protocol == | ||
Line 35: | Line 96: | ||
Though the BLE SoC powering the Mooshimeter appears to be a TI CC2540, | Though the BLE SoC powering the Mooshimeter appears to be a TI CC2540, | ||
the firmware running on it is not the standard "UART to BLE" that is | the firmware running on it is not the standard "UART to BLE" that is | ||
commonly found on them. | commonly found on them. Instead, it implements the entire meter control | ||
and its own custom protocol. | and its own custom protocol. | ||
Line 42: | Line 103: | ||
The lowest layer is the BLE GATT stack, which provides two characteristics: | The lowest layer is the BLE GATT stack, which provides two characteristics: | ||
one to write packets to the meter and one to receive them from it. | one to write packets to the meter and one to receive them from it. The | ||
MTU for a packet in either direction is 20 bytes. | MTU for a packet in either direction is 20 bytes. This is implemented | ||
in the GATT abstraction, so we can talk to it via simple write commands | in the GATT abstraction, so we can talk to it via simple write commands | ||
and a read callback. | and a read callback. | ||
The next layer is the serial stream: each BLE packet in either direction | The next layer is the serial stream: each BLE packet in either direction | ||
has a 1-byte header of a sequence number. | has a 1-byte header of a sequence number. Despite what the documentation | ||
says, this is present in both directions (not just meter output) and is | says, this is present in both directions (not just meter output) and is | ||
NOT reset on the meter output on BLE connection. | NOT reset on the meter output on BLE connection. So the implementation | ||
here needs to provide an output sequence number and incoming reassembly | here needs to provide an output sequence number and incoming reassembly | ||
for out of order packets (I haven't actually observed this, but | for out of order packets (I haven't actually observed this, but | ||
Line 61: | Line 121: | ||
! 1-19 bytes | ! 1-19 bytes | ||
|- | |- | ||
|SeqNum | | SeqNum | ||
| Serial Data | | Serial Data | ||
|} | |} | ||
On top of the serial layer is the "config tree" layer. This is how | |||
the meter actually exposes data and configuration. The tree itself | |||
is composed of nodes, each with a string name, data type, and a list | |||
of children (zero or more). For value containing (non-informational) | |||
nodes, they also contain a 7-bit unique identifier. | |||
Access to the config tree is provided by packets on the serial stream, each packet | |||
config tree is provided by packets on the serial stream, each packet | |||
has a 1-byte header, where the uppermost bit (0x80) is set when writing | has a 1-byte header, where the uppermost bit (0x80) is set when writing | ||
(i.e. never by the meter) and the remaining 7 bits are the node identifier. | (i.e. never by the meter) and the remaining 7 bits are the node identifier. | ||
The length of the packets varies based on the datatype of the tree node. | The length of the packets varies based on the datatype of the tree node. | ||
This means that any lost/dropped packets can make the stream unrecoverable | This means that any lost/dropped packets can make the stream unrecoverable | ||
(i.e. there's no defined sync method other that reconnection). | (i.e. there's no defined sync method other that reconnection). | ||
are emitted by the meter in response to a read or write command (write | |||
Packets are emitted by the meter in response to a read or write command (write | |||
commands simply back the value) and at unsolicited times by the meter | commands simply back the value) and at unsolicited times by the meter | ||
(e.g. continuous sampling and periodic battery voltage). | (e.g. continuous sampling and periodic battery voltage). A read packet | ||
send to the meter looks like: | send to the meter looks like: | ||
Line 125: | Line 187: | ||
| NodeValue | | NodeValue | ||
|} | |} | ||
For the data in the tree, all values are little endian (least significant | For the data in the tree, all values are little endian (least significant | ||
bytes first). | bytes first). The supported type codes are: | ||
{| class="wikitable" | {| class="wikitable" | ||
Line 162: | Line 223: | ||
Plain and Link nodes are present to provide information and/or choices | Plain and Link nodes are present to provide information and/or choices | ||
but do not provide commands codes for direct access (see serialization | but do not provide commands codes for direct access (see serialization | ||
below). | below). Chooser nodes are written with indices described by their Plain | ||
type children (e.g. to select a choice identified by the second child | type children (e.g. to select a choice identified by the second child | ||
of a chooser, write 1 to the chooser node itself). | of a chooser, write 1 to the chooser node itself). | ||
Line 179: | Line 240: | ||
| ADMIN:DIAGNOSTIC || 2 || String | | ADMIN:DIAGNOSTIC || 2 || String | ||
|} | |} | ||
The handshake sequence is to read the contents of ADMIN:TREE, which contains | The handshake sequence is to read the contents of ADMIN:TREE, which contains | ||
the zlib compressed tree serialization, then write the CRC of the compressed | the zlib compressed tree serialization, then write the CRC of the compressed | ||
data back to ADMIN:CRC32 (which the meter will echo back). | data back to ADMIN:CRC32 (which the meter will echo back). Only after | ||
that is done will the meter accept access to the rest of the tree. | that is done will the meter accept access to the rest of the tree. | ||
Line 204: | Line 264: | ||
Once the tree has been deserialized, each node needs it identifier | Once the tree has been deserialized, each node needs it identifier | ||
assigned. | assigned. This is a depth first tree walk, assigning sequential identifiers | ||
first the the current node (if it needs one), then repeating recursively | first the the current node (if it needs one), then repeating recursively | ||
for each of its children. | for each of its children. Plain and Link nodes are skipped in assignment | ||
but not the walk (so the recursion still happens, but the identifier | but not the walk (so the recursion still happens, but the identifier | ||
is not incremented). | is not incremented). | ||
So, for example a write to the ADMIN:CRC32 as part of the handshake would | So, for example a write to the ADMIN:CRC32 as part of the handshake would | ||
Line 251: | Line 310: | ||
! NodeID | ! NodeID | ||
! U16 (len) | ! U16 (len) | ||
! | ! String | ||
|- | |- | ||
! 1 byte | ! 1 byte | ||
! 1 byte | ! 1 byte | ||
! | ! 2 bytes | ||
! | ! len (=8) bytes | ||
|- | |- | ||
| | | 0xAB | ||
| | | 0x20 | ||
| | | 0x0008 | ||
| | | BAD\x20DATA | ||
|} | |} | ||
The config tree at the time of writing looks like: | The config tree at the time of writing looks like: | ||
<small> | |||
<ROOT> (PLAIN) | <ROOT> (PLAIN) | ||
ADMIN (PLAIN) | ADMIN (PLAIN) | ||
Line 360: | Line 419: | ||
1.2 (PLAIN) | 1.2 (PLAIN) | ||
REAL_PWR (FLT) = 39 | REAL_PWR (FLT) = 39 | ||
</small> | |||
== Resources == | == Resources == | ||
Line 365: | Line 425: | ||
* [https://moosh.im/mooshimeter/specs/ Specs] | * [https://moosh.im/mooshimeter/specs/ Specs] | ||
* [https://moosh.im/s/manual/MooshimeterManualRev1.pdf Manual] | * [https://moosh.im/s/manual/MooshimeterManualRev1.pdf Manual] | ||
* [https://play.google.com/store/apps/details?id=com.mooshim.mooshimeter&hl=en_GB Vendor software (Android)] [https://github.com/mooshim/Mooshimeter-AndroidApp sources] | * [https://play.google.com/store/apps/details?id=com.mooshim.mooshimeter&hl=en_GB Vendor software (Android)] ([https://github.com/mooshim/Mooshimeter-AndroidApp sources]) | ||
* [https://itunes.apple.com/us/app/mooshimeter/id945898277?mt=8 Vendor software (iOS)] | * [https://itunes.apple.com/us/app/mooshimeter/id945898277?mt=8 Vendor software (iOS)] | ||
[[Category:Device]] | [[Category:Device]] | ||
[[Category:Multimeter]] | [[Category:Multimeter]] | ||
[[Category: | [[Category:Bluetooth]] | ||
[[Category:Supported]] |
Latest revision as of 14:40, 2 November 2020
Status | supported |
---|---|
Source code | mooshimeter-dmm |
Counts | 24 bit |
IEC 61010-1 | CAT III (600V) |
Connectivity | BLE |
Measurements | voltage, resistance, diode, continuity, frequency, current, temperature |
Features | autorange, logging to SD card, current / voltage simultaneously |
Website | moosh.im |
The Mooshim Engineering Mooshimeter (DMM-BLE-2X01A) is a 24-bit, two channel, CAT III (600V) remote access digital multimeter with Bluetooth Low Energy (BLE) connectivity.
See Mooshim Engineering Mooshimeter/Info for more details (such as hcitool/gatttool output) about the device.
Hardware
Multimeter:
- BLE SoC: TI CC2540 (custom firmware)
- ADC: TI ADS1292
Photos
Multimeter:
Another teardown:
Usage
Example usage using sigrok-cli:
$ sigrok-cli --driver mooshimeter-dmm:conn=68-9e-19-cb-e1-71 --scan The following devices were found: mooshimeter-dmm - Mooshim Engineering Mooshimeter with 3 channels: CH1 CH2 P $ sigrok-cli --driver mooshimeter-dmm:conn=68-9e-19-cb-e1-71 --show Driver functions: Multimeter Scan options: conn mooshimeter-dmm - Mooshim Engineering Mooshimeter with 3 channels: CH1 CH2 P Supported configuration options: continuous: on, off limit_time: avg_samples: - supported values: 32 64 (current) 128 256 samplerate - supported samplerates: 125 Hz 250 Hz 500 Hz 1 kHz 2 kHz 4 kHz 8 kHz channel_config: $ sigrok-cli --driver mooshimeter-dmm:conn=68-9e-19-cb-e1-71 --samples 4 CH1: 0.00 A DC AUTO CH2: 0.00 V DC AUTO CH1: 0.00 A DC AUTO CH2: 0.00 V DC AUTO $ sigrok-cli --driver mooshimeter-dmm:conn=68-9e-19-cb-e1-71 -c channel_config="Aux;0.1/T" --samples 4 CH1: -0.19 V DC CH2: 298.30 °C DC AUTO CH1: -0.19 V DC CH2: 298.30 °C DC AUTO
See also libsigrok's README.devices for more info.
Protocol
Though the BLE SoC powering the Mooshimeter appears to be a TI CC2540, the firmware running on it is not the standard "UART to BLE" that is commonly found on them. Instead, it implements the entire meter control and its own custom protocol.
The Mooshimeter protocol is broken down into several layers in a communication stack.
The lowest layer is the BLE GATT stack, which provides two characteristics: one to write packets to the meter and one to receive them from it. The MTU for a packet in either direction is 20 bytes. This is implemented in the GATT abstraction, so we can talk to it via simple write commands and a read callback.
The next layer is the serial stream: each BLE packet in either direction has a 1-byte header of a sequence number. Despite what the documentation says, this is present in both directions (not just meter output) and is NOT reset on the meter output on BLE connection. So the implementation here needs to provide an output sequence number and incoming reassembly for out of order packets (I haven't actually observed this, but supposedly it happens, which is why the sequence number is present). So the structure of packets received looks like:
1 byte | 1-19 bytes |
---|---|
SeqNum | Serial Data |
On top of the serial layer is the "config tree" layer. This is how the meter actually exposes data and configuration. The tree itself is composed of nodes, each with a string name, data type, and a list of children (zero or more). For value containing (non-informational) nodes, they also contain a 7-bit unique identifier.
Access to the config tree is provided by packets on the serial stream, each packet has a 1-byte header, where the uppermost bit (0x80) is set when writing (i.e. never by the meter) and the remaining 7 bits are the node identifier.
The length of the packets varies based on the datatype of the tree node. This means that any lost/dropped packets can make the stream unrecoverable (i.e. there's no defined sync method other that reconnection).
Packets are emitted by the meter in response to a read or write command (write commands simply back the value) and at unsolicited times by the meter (e.g. continuous sampling and periodic battery voltage). A read packet send to the meter looks like:
1 bit | 7 bits |
---|---|
0 | NodeID |
In response to the read, the meter will send:
1 bit | 7 bits | 1-N bytes |
---|---|---|
0 | NodeID | NodeValue |
A write packet sent to the meter:
1 bit | 7 bits | |
---|---|---|
1 | NodeID | NodeValue |
In response to the write, the meter will send a read response:
1 bit | 7 bits | 1-N bytes |
---|---|---|
0 | NodeID | NodeValue |
For the data in the tree, all values are little endian (least significant bytes first). The supported type codes are:
Code | Description | Wire Format |
---|---|---|
0 | Plain | |
1 | Link | |
2 | Chooser | uint8_t |
3 | U8 | uint8_t |
4 | U16 | uint16_t |
5 | U32 | uint32_t |
6 | S8 | int8_t |
7 | S16 | int16_t |
8 | S32 | int32_t |
9 | String | uint16_t length; char value[length] |
10 | Binary | uint16_t length; uint8_t value[length] |
11 | Float | float |
Plain and Link nodes are present to provide information and/or choices but do not provide commands codes for direct access (see serialization below). Chooser nodes are written with indices described by their Plain type children (e.g. to select a choice identified by the second child of a chooser, write 1 to the chooser node itself).
On initial connection only three nodes at fixed identifiers are available:
Node | ID | Type |
---|---|---|
ADMIN:CRC32 | 0 | U32 |
ADMIN:TREE | 1 | Binary |
ADMIN:DIAGNOSTIC | 2 | String |
The handshake sequence is to read the contents of ADMIN:TREE, which contains the zlib compressed tree serialization, then write the CRC of the compressed data back to ADMIN:CRC32 (which the meter will echo back). Only after that is done will the meter accept access to the rest of the tree.
After zlib decompression the tree serialization is as follows:
Type | Description |
---|---|
uint8_t | The node data type code from above |
uint8_t | Name length |
char[length] | Node name (e.g. "ADMIN" or "CRC32") |
uint8_t | Number of children |
Node[count] | Child serialization (length varies) |
Once the tree has been deserialized, each node needs it identifier assigned. This is a depth first tree walk, assigning sequential identifiers first the the current node (if it needs one), then repeating recursively for each of its children. Plain and Link nodes are skipped in assignment but not the walk (so the recursion still happens, but the identifier is not incremented).
So, for example a write to the ADMIN:CRC32 as part of the handshake would be a write by us (the host):
SerSeq | NodeID | U32 (CRC) |
---|---|---|
1 byte | 1 byte | 4 bytes |
0x01 | 0x80 | 0xDEADBEEF |
The meter will respond with a packet like:
SerSeq | NodeID | U32 (CRC) |
---|---|---|
1 byte | 1 byte | 4 bytes |
0x42 | 0x00 | 0xDEADBEEF |
A spontaneous error from the meter (e.g. in response to a bad packet) can be emitted like:
SerSeq | NodeID | U16 (len) | String |
---|---|---|---|
1 byte | 1 byte | 2 bytes | len (=8) bytes |
0xAB | 0x20 | 0x0008 | BAD\x20DATA |
The config tree at the time of writing looks like:
<ROOT> (PLAIN) ADMIN (PLAIN) CRC32 (U32) = 0 TREE (BIN) = 1 DIAGNOSTIC (STR) = 2 PCB_VERSION (U8) = 3 NAME (STR) = 4 TIME_UTC (U32) = 5 TIME_UTC_MS (U16) = 6 BAT_V (FLT) = 7 REBOOT (CHOOSER) = 8 NORMAL (PLAIN) SHIPMODE (PLAIN) SAMPLING (PLAIN) RATE (CHOOSER) = 9 125 (PLAIN) 250 (PLAIN) 500 (PLAIN) 1000 (PLAIN) 2000 (PLAIN) 4000 (PLAIN) 8000 (PLAIN) DEPTH (CHOOSER) = 10 32 (PLAIN) 64 (PLAIN) 128 (PLAIN) 256 (PLAIN) TRIGGER (CHOOSER) = 11 OFF (PLAIN) SINGLE (PLAIN) CONTINUOUS (PLAIN) LOG (PLAIN) ON (U8) = 12 INTERVAL (U16) = 13 STATUS (U8) = 14 POLLDIR (U8) = 15 INFO (PLAIN) INDEX (U16) = 16 END_TIME (U32) = 17 N_BYTES (U32) = 18 STREAM (PLAIN) INDEX (U16) = 19 OFFSET (U32) = 20 DATA (BIN) = 21 CH1 (PLAIN) MAPPING (CHOOSER) = 22 CURRENT (PLAIN) 10 (PLAIN) TEMP (PLAIN) 350 (PLAIN) SHARED (LINK) RANGE_I (U8) = 23 ANALYSIS (CHOOSER) = 24 MEAN (PLAIN) RMS (PLAIN) BUFFER (PLAIN) VALUE (FLT) = 25 OFFSET (FLT) = 26 BUF (BIN) = 27 BUF_BPS (U8) = 28 BUF_LSB2NATIVE (FLT) = 29 CH2 (PLAIN) MAPPING (CHOOSER) = 30 VOLTAGE (PLAIN) 60 (PLAIN) 600 (PLAIN) TEMP (PLAIN) 350 (PLAIN) SHARED (LINK) RANGE_I (U8) = 31 ANALYSIS (CHOOSER) = 32 MEAN (PLAIN) RMS (PLAIN) BUFFER (PLAIN) VALUE (FLT) = 33 OFFSET (FLT) = 34 BUF (BIN) = 35 BUF_BPS (U8) = 36 BUF_LSB2NATIVE (FLT) = 37 SHARED (CHOOSER) = 38 AUX_V (PLAIN) 0.1 (PLAIN) 0.3 (PLAIN) 1.2 (PLAIN) RESISTANCE (PLAIN) 1000.0 (PLAIN) 10000.0 (PLAIN) 100000.0 (PLAIN) 1000000.0 (PLAIN) 10000000.0 (PLAIN) DIODE (PLAIN) 1.2 (PLAIN) REAL_PWR (FLT) = 39