* [https://play.google.com/store/apps/details?id=com.mooshim.mooshimeter&hl=en_GB Vendor software (Android)] [https://github.com/mooshim/Mooshimeter-AndroidApp sources] | * [https://play.google.com/store/apps/details?id=com.mooshim.mooshimeter&hl=en_GB Vendor software (Android)] [https://github.com/mooshim/Mooshimeter-AndroidApp sources] | ||
* [https://itunes.apple.com/us/app/mooshimeter/id945898277?mt=8 Vendor software (iOS)] | * [https://itunes.apple.com/us/app/mooshimeter/id945898277?mt=8 Vendor software (iOS)] | ||
* [https://gitlab.com/Sizurka/libsigrok/tree/mooshimeter/src/hardware/mooshimeter-dmm Out of tree driver (Linux only BLE stack)] | |||
Revision as of 00:07, 28 March 2019
Status | planned |
---|---|
Source code | ble-dmm |
Counts | 24 bit |
IEC 61010-1 | CAT III (600V) |
Connectivity | ble (aka Bluetooth Smart) ONLY |
Measurements | voltage, resistance, diode, continuity, frequency, current, temperature |
Features | autorange, logging to SD card, current / voltage simultaneously |
Website | https://moosh.im/mooshimeter/ |
The Mooshimeter is a 24-bit, two channel, CAT III (600V) remote access digital multimeter with Bluetooth Low Energy connectivity.
Hardware
Multimeter:
Photos
Multimeter:
Protocol
Though the BLE SoC powering the Mooshimeter appears to be a TI CC2540, the firmware running on it is not the standard "UART to BLE" that is commonly found on them. Instead, it implements the entire meter control and its own custom protocol.
The Mooshimeter protocol is broken down into several layers in a communication stack.
The lowest layer is the BLE GATT stack, which provides two characteristics: one to write packets to the meter and one to receive them from it. The MTU for a packet in either direction is 20 bytes. This is implemented in the GATT abstraction, so we can talk to it via simple write commands and a read callback.
The next layer is the serial stream: each BLE packet in either direction
has a 1-byte header of a sequence number. Despite what the documentation
says, this is present in both directions (not just meter output) and is
NOT reset on the meter output on BLE connection. So the implementation
here needs to provide an output sequence number and incoming reassembly
for out of order packets (I haven't actually observed this, but
supposedly it happens, which is why the sequence number is present).
So the structure of packets received looks like:
1 byte | 1-19 bytes |
---|---|
SeqNum | Serial Data |
On top of the serial layer is the "config tree" layer. This is how
the meter actually exposes data and configuration. The tree itself
is composed of nodes, each with a string name, data type, and a list
of children (zero or more). For value containing (non-informational)
nodes, they also contain a 7-bit unique identifier. Access to the
config tree is provided by packets on the serial stream, each packet
has a 1-byte header, where the uppermost bit (0x80) is set when writing
(i.e. never by the meter) and the remaining 7 bits are the node identifier.
The length of the packets varies based on the datatype of the tree node.
This means that any lost/dropped packets can make the stream unrecoverable
(i.e. there's no defined sync method other that reconnection). Packets
are emitted by the meter in response to a read or write command (write
commands simply back the value) and at unsolicited times by the meter
(e.g. continuous sampling and periodic battery voltage). A read packet
send to the meter looks like:
1 bit | 7 bits |
---|---|
0 | NodeID |
In response to the read, the meter will send:
1 bit | 7 bits | 1-N bytes |
---|---|---|
0 | NodeID | NodeValue |
A write packet sent to the meter:
1 bit | 7 bits | |
---|---|---|
1 | NodeID | NodeValue |
In response to the write, the meter will send a read response:
1 bit | 7 bits | 1-N bytes |
---|---|---|
0 | NodeID | NodeValue |
For the data in the tree, all values are little endian (least significant
bytes first). The supported type codes are:
Code | Description | Wire Format |
---|---|---|
0 | Plain | |
1 | Link | |
2 | Chooser | uint8_t |
3 | U8 | uint8_t |
4 | U16 | uint16_t |
5 | U32 | uint32_t |
6 | S8 | int8_t |
7 | S16 | int16_t |
8 | S32 | int32_t |
9 | String | uint16_t length; char value[length] |
10 | Binary | uint16_t length; uint8_t value[length] |
11 | Float | float |
Plain and Link nodes are present to provide information and/or choices but do not provide commands codes for direct access (see serialization below). Chooser nodes are written with indices described by their Plain type children (e.g. to select a choice identified by the second child of a chooser, write 1 to the chooser node itself).
On initial connection only three nodes at fixed identifiers are available:
Node | ID | Type |
---|---|---|
ADMIN:CRC32 | 0 | U32 |
ADMIN:TREE | 1 | Binary |
ADMIN:DIAGNOSTIC | 2 | String |
The handshake sequence is to read the contents of ADMIN:TREE, which contains
the zlib compressed tree serialization, then write the CRC of the compressed
data back to ADMIN:CRC32 (which the meter will echo back). Only after
that is done will the meter accept access to the rest of the tree.
After zlib decompression the tree serialization is as follows:
Type | Description |
---|---|
uint8_t | The node data type code from above |
uint8_t | Name length |
char[length] | Node name (e.g. "ADMIN" or "CRC32") |
uint8_t | Number of children |
Node[count] | Child serialization (length varies) |
Once the tree has been deserialized, each node needs it identifier assigned. This is a depth first tree walk, assigning sequential identifiers first the the current node (if it needs one), then repeating recursively for each of its children. Plain and Link nodes are skipped in assignment but not the walk (so the recursion still happens, but the identifier is not incremented).
So, for example a write to the ADMIN:CRC32 as part of the handshake would
be a write by us (the host):
SerSeq | NodeID | U32 (CRC) |
---|---|---|
1 byte | 1 byte | 4 bytes |
0x01 | 0x80 | 0xDEADBEEF |
The meter will respond with a packet like:
SerSeq | NodeID | U32 (CRC) |
---|---|---|
1 byte | 1 byte | 4 bytes |
0x42 | 0x00 | 0xDEADBEEF |
A spontaneous error from the meter (e.g. in response to a bad packet) can be emitted like:
SerSeq | NodeID | U16 (len) | String |
---|---|---|---|
1 byte | 1 byte | 2 bytes | len (=8) bytes |
0xAB | 0x20 | 0x0008 | BAD\x20DATA |
The config tree at the time of writing looks like:
<ROOT> (PLAIN) ADMIN (PLAIN) CRC32 (U32) = 0 TREE (BIN) = 1 DIAGNOSTIC (STR) = 2 PCB_VERSION (U8) = 3 NAME (STR) = 4 TIME_UTC (U32) = 5 TIME_UTC_MS (U16) = 6 BAT_V (FLT) = 7 REBOOT (CHOOSER) = 8 NORMAL (PLAIN) SHIPMODE (PLAIN) SAMPLING (PLAIN) RATE (CHOOSER) = 9 125 (PLAIN) 250 (PLAIN) 500 (PLAIN) 1000 (PLAIN) 2000 (PLAIN) 4000 (PLAIN) 8000 (PLAIN) DEPTH (CHOOSER) = 10 32 (PLAIN) 64 (PLAIN) 128 (PLAIN) 256 (PLAIN) TRIGGER (CHOOSER) = 11 OFF (PLAIN) SINGLE (PLAIN) CONTINUOUS (PLAIN) LOG (PLAIN) ON (U8) = 12 INTERVAL (U16) = 13 STATUS (U8) = 14 POLLDIR (U8) = 15 INFO (PLAIN) INDEX (U16) = 16 END_TIME (U32) = 17 N_BYTES (U32) = 18 STREAM (PLAIN) INDEX (U16) = 19 OFFSET (U32) = 20 DATA (BIN) = 21 CH1 (PLAIN) MAPPING (CHOOSER) = 22 CURRENT (PLAIN) 10 (PLAIN) TEMP (PLAIN) 350 (PLAIN) SHARED (LINK) RANGE_I (U8) = 23 ANALYSIS (CHOOSER) = 24 MEAN (PLAIN) RMS (PLAIN) BUFFER (PLAIN) VALUE (FLT) = 25 OFFSET (FLT) = 26 BUF (BIN) = 27 BUF_BPS (U8) = 28 BUF_LSB2NATIVE (FLT) = 29 CH2 (PLAIN) MAPPING (CHOOSER) = 30 VOLTAGE (PLAIN) 60 (PLAIN) 600 (PLAIN) TEMP (PLAIN) 350 (PLAIN) SHARED (LINK) RANGE_I (U8) = 31 ANALYSIS (CHOOSER) = 32 MEAN (PLAIN) RMS (PLAIN) BUFFER (PLAIN) VALUE (FLT) = 33 OFFSET (FLT) = 34 BUF (BIN) = 35 BUF_BPS (U8) = 36 BUF_LSB2NATIVE (FLT) = 37 SHARED (CHOOSER) = 38 AUX_V (PLAIN) 0.1 (PLAIN) 0.3 (PLAIN) 1.2 (PLAIN) RESISTANCE (PLAIN) 1000.0 (PLAIN) 10000.0 (PLAIN) 100000.0 (PLAIN) 1000000.0 (PLAIN) 10000000.0 (PLAIN) DIODE (PLAIN) 1.2 (PLAIN) REAL_PWR (FLT) = 39