Difference between revisions of "Gossen Metrawatt Metrahit 16I"
Jump to navigation
Jump to search
m (Link to manufacturer page) |
(→Measurement functions and ranges: Cosmetics.) |
||
(4 intermediate revisions by the same user not shown) | |||
Line 3: | Line 3: | ||
| name = Gossen Metrawatt Metrahit 16I | | name = Gossen Metrawatt Metrahit 16I | ||
| status = supported | | status = supported | ||
| source_code_dir = | | source_code_dir = gmc-mh-1x-2x | ||
| counts = 3100 (31000) | | counts = 3100 (31000) | ||
| categories = CAT III 1000 V, CAT IV 600 V | | categories = CAT III 1000 V, CAT IV 600 V | ||
Line 44: | Line 44: | ||
| 30 mV | | 30 mV | ||
| 10 μV | | 10 μV | ||
| ±(0 | | ±(0.5% + 3) | ||
| mq V, unit V, mqflags DC | | mq V, unit V, mqflags DC | ||
| | | Input impendance >10 GΩ; not covered by autorange, use manual range | ||
|- | |- | ||
| 300 mV | | 300 mV | ||
| 100 μV | | 100 μV | ||
| ±(0 | | ±(0.5% + 3) | ||
| " | |||
| " | | " | ||
|- | |- | ||
| 3 V | | 3 V | ||
| 1 mV | | 1 mV | ||
| rowspan="3" | ±(0 | | rowspan="3" | ±(0.25% + 1) | ||
| " | | " | ||
| | | | ||
Line 77: | Line 77: | ||
| 600 V | | 600 V | ||
| 1 V | | 1 V | ||
| ±(0 | | ±(0.35% + 1) | ||
| " | | " | ||
| | | | ||
Line 137: | Line 137: | ||
| 30 A | | 30 A | ||
| 10 mA | | 10 mA | ||
| rowspan="2" | ±(2 | | rowspan="2" | ±(2.5% + 3) (>10 Digit) | ||
| mq Curr., unit A, mqflags AC,DC,RMS | | mq Curr., unit A, mqflags AC,DC,RMS | ||
| rowspan="2" | Current clamp with 100 mV/A required, e.g. GMC WZ12B | | rowspan="2" | Current clamp with 100 mV/A required, e.g. GMC WZ12B | ||
Line 150: | Line 150: | ||
| 30.00 Ω | | 30.00 Ω | ||
| 10 mΩ | | 10 mΩ | ||
| rowspan="2" | ±(0 | | rowspan="2" | ±(0.5% + 3) | ||
| mq Res., unit Ω, mqflags - | | mq Res., unit Ω, mqflags - | ||
| | | | ||
Line 163: | Line 163: | ||
| 3.000 kΩ | | 3.000 kΩ | ||
| 1 Ω | | 1 Ω | ||
| rowspan="3" | ±(0 | | rowspan="3" | ±(0.4% + 1) | ||
| " | | " | ||
| | | | ||
Line 182: | Line 182: | ||
| 3.000 MΩ | | 3.000 MΩ | ||
| 1 kΩ | | 1 kΩ | ||
| ±(0 | | ±(0.6% + 1) | ||
| " | | " | ||
| | | | ||
Line 189: | Line 189: | ||
| 30.00 MΩ | | 30.00 MΩ | ||
| 10 kΩ | | 10 kΩ | ||
| ±(2 | | ±(2.0% + 1) | ||
| " | | " | ||
| | | | ||
Line 195: | Line 195: | ||
|- | |- | ||
| Ω Iso 500V | | Ω Iso 500V | ||
| 1 | | 1.600 GΩ | ||
| 1 kΩ - 1 MΩ | | 1 kΩ - 1 MΩ | ||
| rowspan="2" | ±(3% + 2) | | rowspan="2" | ±(3% + 2) | ||
Line 203: | Line 203: | ||
|- | |- | ||
| Ω Iso 1000V | | Ω Iso 1000V | ||
| 3 | | 3.100 GΩ | ||
| 1 kΩ - 1 MΩ | | 1 kΩ - 1 MΩ | ||
| mq Res., unit Ω, mqflags - | | mq Res., unit Ω, mqflags - | ||
Line 210: | Line 210: | ||
|- | |- | ||
| Diode/Cont. | | Diode/Cont. | ||
| 3. | | 3.000 V– | ||
| 1 mV | | 1 mV | ||
| ±(0 | | ±(0.25% + 1) | ||
| mq V, unit V, mqflags DC,DIODE | | mq V, unit V, mqflags DC,DIODE | ||
| Diode and cont. are identical on this device, just without beeper for diode. | | Diode and cont. are identical on this device, just without beeper for diode. | ||
Line 220: | Line 220: | ||
| 30.00 nF | | 30.00 nF | ||
| 10 pF | | 10 pF | ||
| rowspan="3" | ±(1 | | rowspan="3" | ±(1.0% + 3) | ||
| mq Cap., unit F, mqflags - | | mq Cap., unit F, mqflags - | ||
| | | | ||
Line 239: | Line 239: | ||
| 30.00 µF | | 30.00 µF | ||
| 10 nF | | 10 nF | ||
| ±(3 | | ±(3.0% + 3) | ||
| " | | " | ||
| | | | ||
Line 248: | Line 248: | ||
| 0.01 Hz | | 0.01 Hz | ||
| rowspan="4" | ±(0.1% + 3) | | rowspan="4" | ±(0.1% + 3) | ||
| mq Freq., unit Hz | | mq Freq., unit Hz | ||
| | | The protocol does not allow to differ AC and DC. | ||
|- | |- | ||
Line 255: | Line 255: | ||
| 0.1 Hz | | 0.1 Hz | ||
| " | | " | ||
| | | | ||
|- | |- | ||
Line 274: | Line 274: | ||
| 0.01 Hz | | 0.01 Hz | ||
| rowspan="4" | ±(0.1% + 3) | | rowspan="4" | ±(0.1% + 3) | ||
| mq Freq., unit Hz | | mq Freq., unit Hz | ||
| | | The protocol does not allow to differ AC and DC. | ||
|- | |- |
Latest revision as of 20:33, 21 September 2016
Status | supported |
---|---|
Source code | gmc-mh-1x-2x |
Counts | 3100 (31000) |
IEC 61010-1 | CAT III 1000 V, CAT IV 600 V |
Connectivity | Infrared (RS232) |
Measurements | voltage, current, resistance, capacitance, temperature, diode, continuity |
Features | autorange, data hold, bargraph |
Website | [1] |
The Gossen Metrawatt Metrahit 16I is a 3100 counts handheld digital multimeter with IR (RS232) connectivity and insulation measurement (500/1000 V).
The resolution is one digit higher on the data interface.
Hardware
- Battery: 9 V IEC 6 LR 61
- Fuse: None! (not required, no internal amps ranges)
- TODO.
Photos
Measurement functions and ranges
Measurement function | Range | Resolution | Deviation | sigrok result | Remarks |
---|---|---|---|---|---|
V= | 30 mV | 10 μV | ±(0.5% + 3) | mq V, unit V, mqflags DC | Input impendance >10 GΩ; not covered by autorange, use manual range |
300 mV | 100 μV | ±(0.5% + 3) | " | " | |
3 V | 1 mV | ±(0.25% + 1) | " | ||
30 V | 10 mV | " | |||
300 V | 100 mV | " | |||
600 V | 1 V | ±(0.35% + 1) | " | ||
V≈ | 3 V | 100 μV | ±(1.0% + 3) (> 10 Digit) | mq V, unit V, mqflags AC,RMS | |
30 V | 100 μV | " | |||
300 V | 1 mV | " | |||
1000 V | 10 mV | " | |||
V≃ | 3 V | 100 μV | ±(1.0% + 3) (> 10 Digit) | mq V, unit V, mqflags AC,DC,RMS | |
30 V | 100 μV | " | |||
300 V | 1 mV | " | |||
1000 V | 10 mV | " | |||
A≈ (Clamp only) | 30 A | 10 mA | ±(2.5% + 3) (>10 Digit) | mq Curr., unit A, mqflags AC,DC,RMS | Current clamp with 100 mV/A required, e.g. GMC WZ12B |
100 A | 100 mA | " | |||
Ω | 30.00 Ω | 10 mΩ | ±(0.5% + 3) | mq Res., unit Ω, mqflags - | |
300.0 Ω | 100 mΩ | " | |||
3.000 kΩ | 1 Ω | ±(0.4% + 1) | " | ||
30.00 kΩ | 10 Ω | " | |||
300.0 kΩ | 100 Ω | " | |||
3.000 MΩ | 1 kΩ | ±(0.6% + 1) | " | ||
30.00 MΩ | 10 kΩ | ±(2.0% + 1) | " | ||
Ω Iso 500V | 1.600 GΩ | 1 kΩ - 1 MΩ | ±(3% + 2) | mq Res., unit Ω, mqflags - | Altogether 4 ranges with different resolution |
Ω Iso 1000V | 3.100 GΩ | 1 kΩ - 1 MΩ | mq Res., unit Ω, mqflags - | Altogether 4 ranges with different resolution | |
Diode/Cont. | 3.000 V– | 1 mV | ±(0.25% + 1) | mq V, unit V, mqflags DC,DIODE | Diode and cont. are identical on this device, just without beeper for diode. |
Cap. F | 30.00 nF | 10 pF | ±(1.0% + 3) | mq Cap., unit F, mqflags - | |
300.0 nF | 100 pF | " | |||
3.000 µF | 1 nF | " | |||
30.00 µF | 10 nF | ±(3.0% + 3) | " | ||
Hz= | 300.0 Hz | 0.01 Hz | ±(0.1% + 3) | mq Freq., unit Hz | The protocol does not allow to differ AC and DC. |
3.000 kHz | 0.1 Hz | " | |||
30.00 kHz | 1 Hz | " | |||
100.00 kHz | 10 Hz | " | |||
Hz≈ | 300.0 Hz | 0.01 Hz | ±(0.1% + 3) | mq Freq., unit Hz | The protocol does not allow to differ AC and DC. |
3.000 kHz | 0.1 Hz | " | |||
30.00 kHz | 1 Hz | " | |||
100.00 kHz | 10 Hz | " | |||
°C | Pt100 –200.0 - +200.0 °C | 0.1 °C | 2 Kelvin + 5 | mq Temp., unit °C, mqflags - | |
Pt100 +200.0 - +800.0 °C | ±(1.0% + 5) | " | |||
Pt1000 –100.0 - +200.0 °C | 2 Kelvin + 5 | " | |||
Pt1000 +200.0 - +800.0 °C | ±(1.0% + 5) | " | |||
Event count | Seems not to be present on data output. | ||||
Data Hold | mqflags += HOLD | Disabled while IR interface is active on Matthias' device. | |||
MIN | mqflags += MIN | Influences only display, via IR flag is set, but actual current value is delivered. | |||
MAX | mqflags += MAX | Influences only display, via IR flag is set, but actual current value is delivered. |
The column "sigrok result" contains in short form what the driver generates for the resp. data type.