Line 3: Line 3:
| name            = Siemens B1105
| name            = Siemens B1105
| status          = supported
| status          = supported
| source_code_dir = gmc_mh_1x_2x
| source_code_dir = gmc-mh-1x-2x
| counts          = 310000
| counts          = 310000
| categories      = CAT III 1000 V, CAT IV 600 V (older version)<br>CAT III 600 V, CAT IV 300 V (newer version)
| categories      = CAT III 1000 V, CAT IV 600 V (older version)<br>CAT III 600 V, CAT IV 300 V (newer version)
Line 52: Line 52:
| 300 mV
| 300 mV
| 1 μV
| 1 μV
| ±(0,02% + 0,010% + 5)  
| ±(0.02% + 0.010% + 5)  
| mq V, unit V, mqflags DC
| mq V, unit V, mqflags DC
|  
|  
Line 59: Line 59:
| 3 V
| 3 V
| 10 μV
| 10 μV
| rowspan="4" | ±(0,02% + 0,005% + 5)  
| rowspan="4" | ±(0.02% + 0.005% + 5)  
| "
| "
|  
|  
Line 85: Line 85:
| 300 mV
| 300 mV
| 10 µV
| 10 µV
| ±(0,5% + 30) (> 500 Digit))
| ±(0.5% + 30) (> 500 Digit))
| mq V, unit V, mqflags AC,RMS
| mq V, unit V, mqflags AC,RMS
|
|
Line 92: Line 92:
| 3 V
| 3 V
| 100 μV
| 100 μV
| rowspan="4" | ±(0,2% + 30)
| rowspan="4" | ±(0.2% + 30)
| "
| "
|  
|  
Line 118: Line 118:
| 300 mV
| 300 mV
| 10 μV
| 10 μV
| ±(0,5% + 30) (> 500 Digit))
| ±(0.5% + 30) (> 500 Digit))
| mq V, unit V, mqflags AC,DC,RMS
| mq V, unit V, mqflags AC,DC,RMS
|  
|  
Line 125: Line 125:
| 3 V
| 3 V
| 100 μV
| 100 μV
| rowspan="4" | ±(0,2% + 30)
| rowspan="4" | ±(0.2% + 30)
| "
| "
|  
|  
Line 150: Line 150:
| V db≈
| V db≈
| -48 - 63 dB
| -48 - 63 dB
| 0,01 dB
| 0.01 dB
| ± 0,1 dB  
| ± 0.1 dB  
| mq V, unit V, mqflags AC,RMS
| mq V, unit V, mqflags AC,RMS
| Actually the 5 AC ranges, see there for accuracy.
| Actually the 5 AC ranges, see there for accuracy.
Line 158: Line 158:
| V db≈ rel
| V db≈ rel
| -48 - 63 dB
| -48 - 63 dB
| 0,01 dB
| 0.01 dB
| ± 0,1 dB  
| ± 0.1 dB  
| mq V, unit V, mqflags AC,RMS
| mq V, unit V, mqflags AC,RMS
| Actually the 5 AC ranges, see there for accuracy. To activate, press yellow button in dB mode, then do second measurement. Cannot be distinguished from non-rel measurement remotely.
| Actually the 5 AC ranges, see there for accuracy. To activate, press yellow button in dB mode, then do second measurement. Cannot be distinguished from non-rel measurement remotely.
Line 167: Line 167:
| 300.00 µA
| 300.00 µA
| 1 nA
| 1 nA
| 0,05% + 0,02% + 5
| 0.05% + 0.02% + 5
| mq Curr., unit A, mqflags DC
| mq Curr., unit A, mqflags DC
|  
|  
Line 174: Line 174:
| 3.0000 mA
| 3.0000 mA
| 10 nA
| 10 nA
| 0,05% + 0,01% + 5
| 0.05% + 0.01% + 5
| "
| "
|  
|  
Line 181: Line 181:
| 30.000 mA
| 30.000 mA
| 100 nA
| 100 nA
| 0,05% + 0,01% + 5
| 0.05% + 0.01% + 5
| "
| "
|  
|  
Line 188: Line 188:
| 300.00 mA
| 300.00 mA
| 1 µA
| 1 µA
| 0,1% + 0,01% + 5
| 0.1% + 0.01% + 5
| "
| "
|  
|  
Line 195: Line 195:
| 3.0000 A
| 3.0000 A
| 100 µA
| 100 µA
| 0,2% + 0,05% + 5
| 0.2% + 0.05% + 5
| "
| "
|  
|  
Line 202: Line 202:
| 10.000 A
| 10.000 A
| 1 mA
| 1 mA
| 0,2% + 0,05% + 5  
| 0.2% + 0.05% + 5  
| "
| "
|  
|  
Line 210: Line 210:
| 300.00 µA
| 300.00 µA
| 10 nA
| 10 nA
| rowspan="4" | 0,5% + 30
| rowspan="4" | 0.5% + 30
| mq Curr., unit A, mqflags AC,DC,RMS
| mq Curr., unit A, mqflags AC,DC,RMS
|  
|  
Line 235: Line 235:
| 3.0000 A
| 3.0000 A
| 100 µA
| 100 µA
| 0,7% + 30
| 0.7% + 30
| "
| "
|  
|  
Line 242: Line 242:
| 10.000 A
| 10.000 A
| 1 mA
| 1 mA
| 0,5% + 30
| 0.5% + 30
| "
| "
|  
|  
Line 249: Line 249:
| 300.00 Ω
| 300.00 Ω
| 1 mΩ
| 1 mΩ
| rowspan="4" | 0,05% + 0.01% + 5
| rowspan="4" | 0.05% + 0.01% + 5
| mq Res., unit Ω, mqflags -
| mq Res., unit Ω, mqflags -
|
|
Line 274: Line 274:
| 3.0000 MΩ
| 3.0000 MΩ
| 10 Ω
| 10 Ω
| 0,1% + 0.02% + 5
| 0.1% + 0.02% + 5
| "
| "
|  
|  
Line 384: Line 384:
| Pt100/Pt1000<br> –200.0 - +100.0 °C
| Pt100/Pt1000<br> –200.0 - +100.0 °C
| rowspan="4" | 0.1 °C
| rowspan="4" | 0.1 °C
| ±(0,5 K + 3)
| ±(0.5 K + 3)
| mq Temp., unit °C, mqflags -
| mq Temp., unit °C, mqflags -
|
|
Line 390: Line 390:
|-
|-
| Pt100/Pt1000<br> +100.0 - 850.0 °C
| Pt100/Pt1000<br> +100.0 - 850.0 °C
| ±(0,2% + 3)
| ±(0.2% + 3)
| mq Temp., unit °C, mqflags -
| mq Temp., unit °C, mqflags -
|
|


|-
|-
| K NiCr-Ni -270-+1372,0 °C
| K NiCr-Ni -270-+1372.0 °C
| ±(0,7% + 3)
| ±(0.7% + 3)
| "
| "
|
|
Line 402: Line 402:
|-
|-
| J Fe-CuNi -210 - +1200.0 °C
| J Fe-CuNi -210 - +1200.0 °C
| ±(0,8% + 3)
| ±(0.8% + 3)
| "
| "
| Not tested yet due to lack of corresponding sensor.
| Not tested yet due to lack of corresponding sensor.
Line 437: Line 437:
|   
|   
|  
|  
| style="background: #ff0000; "| mqflags += HOLD
| style="background: #ff0000; "| <s>mqflags += HOLD</s>
| Disabled while IR interface is active on [[User:Matthias Heidbrink | Matthias']] test device.
| Disabled while IR interface is active on [[User:Matthias Heidbrink | Matthias']] test device.
   
   
Line 461: Line 461:


The column "sigrok result" contains in short form what the driver generates for the resp. data type.
The column "sigrok result" contains in short form what the driver generates for the resp. data type.


[[Category:Device]]
[[Category:Device]]
[[Category:Multimeter]]
[[Category:Multimeter]]
[[Category:Supported]]
[[Category:Supported]]

Latest revision as of 20:27, 21 September 2016

Siemens B1105
Siemens B1105 small.png
Status supported
Source code gmc-mh-1x-2x
Counts 310000
IEC 61010-1 CAT III 1000 V, CAT IV 600 V (older version)
CAT III 600 V, CAT IV 300 V (newer version)
Connectivity Infrared (RS232/USB)
Measurements voltage, current, resistance, capacitance, temperature, diode, continuity
Features autorange, data hold
Website [1]

The Siemens B1105 is a 310000 counts 1000V CAT III/600 V CAT IV (older version) or 600V CAT III/300 V CAT IV (newer version) handheld digital multimeter with IR (RS232/USB) connectivity.

It is an OEM version of the Gossen Metrawatt  Metrahit 28S. The only differences seem to be the type, logo and color (light grey-blue print on knob, light grey-blue holster) and the voltage rating.

Hardware

  • Power supply:
    • Battery: 2x 1,5 V IEC R 6, IEC LR 6 or corresponding NiCd rechargeable batteries
    • External power supply NA4/500 by Gossen Metrawatt, 4.2 V, 500 mA (for older version)
    • External power supply NA5/600 by Gossen Metrawatt, 5 V, 600 mA (for newer version, 3-pin plug)
  • Fuse:
    • Current 300 mA: FF (UR) 1,6 A/1000 V AC/DC (10 kA) 6,3 mm x 32 mm
    • Current 3A/10A: FF (UR) 10 A/1000 V AC/DC (30 kA) 10 mm x 38 mm

Photos

Measurement functions and ranges

Measurement function Range Resolution Accuracy sigrok result Remarks
V= 300 mV 1 μV ±(0.02% + 0.010% + 5) mq V, unit V, mqflags DC
3 V 10 μV ±(0.02% + 0.005% + 5) "
30 V 100 μV "
300 V 1 mV "
1000 V 10 mV "
V≈ 300 mV 10 µV ±(0.5% + 30) (> 500 Digit)) mq V, unit V, mqflags AC,RMS
3 V 100 μV ±(0.2% + 30) "
30 V 100 μV "
300 V 1 mV "
1000 V 10 mV "
V≃ 300 mV 10 μV ±(0.5% + 30) (> 500 Digit)) mq V, unit V, mqflags AC,DC,RMS
3 V 100 μV ±(0.2% + 30) "
30 V 100 μV "
300 V 1 mV "
1000 V 10 mV "
V db≈ -48 - 63 dB 0.01 dB ± 0.1 dB mq V, unit V, mqflags AC,RMS Actually the 5 AC ranges, see there for accuracy.
V db≈ rel -48 - 63 dB 0.01 dB ± 0.1 dB mq V, unit V, mqflags AC,RMS Actually the 5 AC ranges, see there for accuracy. To activate, press yellow button in dB mode, then do second measurement. Cannot be distinguished from non-rel measurement remotely.
A=  300.00 µA 1 nA 0.05% + 0.02% + 5 mq Curr., unit A, mqflags DC
3.0000 mA 10 nA 0.05% + 0.01% + 5 "
30.000 mA 100 nA 0.05% + 0.01% + 5 "
300.00 mA 1 µA 0.1% + 0.01% + 5 "
3.0000 A 100 µA 0.2% + 0.05% + 5 "
10.000 A 1 mA 0.2% + 0.05% + 5 "
A≃  300.00 µA 10 nA 0.5% + 30 mq Curr., unit A, mqflags AC,DC,RMS
3.0000 mA 100 nA "
30.000 mA 1 µA "
300.00 mA 10 µA "
3.0000 A 100 µA 0.7% + 30 "
10.000 A 1 mA 0.5% + 30 "
Ω 300.00 Ω 1 mΩ 0.05% + 0.01% + 5 mq Res., unit Ω, mqflags -
3.0000 kΩ 10 mΩ "
30.000 kΩ 100 mΩ "
300.00 kΩ 1 Ω "
3.0000 MΩ 10 Ω 0.1% + 0.02% + 5 "
30.000 MΩ 100 Ω 1% + 0.2% + 5 "
Ohm/Cont. 300 Ω 100 mΩ 1% + 3% mq Res., unit Ω, mqflags -
Diode/Cont. 300 mV 100 µV 0.2% + 3 mq V, unit V, mqflags DC,DIODE
Diode 3 V 100 µV 0.2% + 3 mq V, unit V, mqflags DC,DIODE
Cap. F 3.000 nF 1 pF 1.0% + 0.2 mq Cap., unit F, mqflags -
30.00 nF 10 pF "
300.0 nF 100 pF "
3.000 µF 1 nF "
30.00 µF 10 nF "
300.0 µF 100 nF ±(5.0% + 6) "
3.000 mF 1 µF "
30.00 mF 1 µF "
Hz≈, ≃ 300.00 Hz 0.001 Hz ±(0.05% + 1) mq Freq., unit Hz, mqflags AC,DC To activate in AC, press yellow button twice.
3.0000 kHz 0.01 Hz "
300.000 kHz 1 Hz "
°C Pt100/Pt1000
–200.0 - +100.0 °C
0.1 °C ±(0.5 K + 3) mq Temp., unit °C, mqflags -
Pt100/Pt1000
+100.0 - 850.0 °C
±(0.2% + 3) mq Temp., unit °C, mqflags -
K NiCr-Ni -270-+1372.0 °C ±(0.7% + 3) "
J Fe-CuNi -210 - +1200.0 °C ±(0.8% + 3) " Not tested yet due to lack of corresponding sensor.
°F " " " mq Temp., unit °F, mqflags - Details and ranges like °C.


Event count mq V, unit V, mqflags AC,RMS,REL Counts events >= 2500 digits that last >=2s with a break >=2s. To activate, switch to voltage ACDC, press yellow button twice. Press up to 4 times to cycle parameters, press long to stop. See manual. Not useful remotely because the current measurement value is sent, not the counter.
Stop Watch 100 ns (Time data) Max. 100 Min. To activate, select a 3V-1000V≃ range manually, press yellow button, press AUTO for start/stop, DATA for reset. Using the stop watch delivers only voltage data and sometimes seems to bring the device into a mode where it even sends invalid data packets.
Data Hold mqflags += HOLD Disabled while IR interface is active on  Matthias' test device.
MIN mqflags += MIN Influences only display, via IR flag is set, but actual current value is delivered.
MAX mqflags += MAX Influences only display, via IR flag is set, but actual current value is delivered.

The Accuracy row was simplified, see manual for exact specs. If the value contains three values, the middle value is % or range.

The column "sigrok result" contains in short form what the driver generates for the resp. data type.