1-------------------------------------------------------------------------------
2README.devices
3-------------------------------------------------------------------------------
4
5This README contains various notes for users of libsigrok (or frontends
6that are based on libsigrok) about device- and/or driver-specific issues.
7
8
9Firmware
10--------
11
12Some devices supported by libsigrok need a firmware to be uploaded every time
13the device is connected to the PC (usually via USB), before it can be used.
14
15The default locations where libsigrok expects the firmware files are:
16
17 $HOME/.local/share/sigrok-firmware
18 $prefix/share/sigrok-firmware
19 /usr/local/share/sigrok-firmware
20 /usr/share/sigrok-firmware
21
22($prefix is usually /usr/local or /usr, depending on your ./configure options)
23
24For further information see the section below and also:
25
26 http://sigrok.org/wiki/Firmware
27
28
29Per-driver firmware requirements
30--------------------------------
31
32The following drivers/devices require a firmware upload upon connection:
33
34 - asix-sigma: The ASIX SIGMA and SIGMA2 require various firmware files,
35 depending on the settings used. These files are available from our
36 'sigrok-firmware' repository/project under a license which allows us
37 to redistribute them.
38
39 - fx2lafw: Logic analyzers based on the Cypress FX2(LP) chip need the
40 firmware files from the 'sigrok-firmware-fx2lafw' repository/project.
41 The firmware is written from scratch and licensed under the GNU GPLv2+.
42
43 - hantek-6xxx: Certain oscilloscopes based on the Cypress FX2(LP) chip, such
44 as the Hantek 6022BE/6022BL, SainSmart DDS120, and Rocktech BM102, need the
45 firmware files from the 'sigrok-firmware-fx2lafw' repository/project.
46 The firmware is written from scratch and licensed under the GNU GPLv2+.
47
48 - hantek-dso: The Hantek DSO-2090 (and other supported models of the same
49 series of Hantek PC oscilloscopes) need firmware files.
50 These can be extracted from the vendor's Windows drivers using a tool
51 from our 'sigrok-util' repository/project.
52
53 - lecroy-logicstudio: The LeCroy LogicStudio requires FPGA bitstream files.
54 These can be extracted from the vendor's Windows software using a tool
55 from our 'sigrok-util' repository/project.
56 Additionally, it requires a Cypress FX2 firmware. This can be extracted
57 from the vendor's Windows software using another tool. Details:
58
59 http://sigrok.org/wiki/LeCroy_LogicStudio#Firmware
60
61 - saleae-logic16: The Saleae Logic16 needs a firmware file for the
62 Cypress FX2 chip in the device, as well as two FPGA bitstream files.
63 These can be extracted from the vendor's Linux application using a tool
64 from our 'sigrok-util' repository/project.
65
66 - sysclk-lwla:
67
68 - The Sysclk LWLA1034 requires various bitstream files.
69 These files are available from our 'sigrok-firmware' repository/project
70 under a license which allows us to redistribute them.
71
72 - The Sysclk LWLA1016 requires various bitstream files.
73 These can be extracted from the vendor's Windows drivers using a tool
74 from our 'sigrok-util' repository/project.
75
76The following drivers/devices do not need any firmware upload:
77
78 - agilent-dmm
79 - appa-55ii
80 - arachnid-labs-re-load-pro
81 - atten-pps3xxx
82 - baylibre-acme
83 - beaglelogic
84 - brymen-bm86x
85 - brymen-dmm
86 - cem-dt-885x
87 - center-3xx (including all subdrivers)
88 - chronovu-la
89 - colead-slm
90 - conrad-digi-35-cpu
91 - demo
92 - fluke-dmm
93 - ftdi-la
94 - gmc-mh-1x-2x (including all subdrivers)
95 - gwinstek-gds-800
96 - hameg-hmo
97 - hp-3457a
98 - hung-chang-dso-2100
99 - ikalogic-scanalogic2
100 - ikalogic-scanaplus
101 - kecheng-kc-330b
102 - kern-scale
103 - lascar-el-usb
104 - link-mso19
105 - manson-hcs-3xxx
106 - maynuo-m97
107 - mic-985xx (including all subdrivers)
108 - motech-lps-30x
109 - norma-dmm
110 - openbench-logic-sniffer
111 - pce-322a
112 - pipistrello-ols
113 - rigol-ds
114 - scpi-pps
115 - serial-dmm (including all subdrivers)
116 - serial-lcr (including all subdrivers)
117 - teleinfo
118 - testo
119 - tondaj-sl-814
120 - uni-t-dmm (including all subdrivers)
121 - uni-t-ut32x
122 - victor-dmm
123 - yokogawa-dlm
124 - zeroplus-logic-cube
125
126
127Specifying serial ports
128-----------------------
129
130Many devices supported by libsigrok use serial port based cables (real RS232
131or USB-to-serial ones) to connect to a PC.
132
133For all these devices, you need to specify the serial port they are connected
134to (e.g. using the 'conn' option in sigrok-cli). It is not possible to scan
135for such devices without specifying a serial port.
136
137Example:
138
139 $ sigrok-cli --driver <somedriver>:conn=/dev/ttyUSB0 ...
140
141The following drivers/devices require a serial port specification:
142
143 - agilent-dmm
144 - appa-55ii
145 - atten-pps3xxx
146 - brymen-dmm
147 - cem-dt-885x
148 - center-3xx (including all subdrivers)
149 - colead-slm
150 - conrad-digi-35-cpu
151 - fluke-dmm
152 - gmc-mh-1x-2x (including all subdrivers)
153 - hameg-hmo
154 - link-mso19
155 - mic-985xx (including all subdrivers)
156 - norma-dmm
157 - openbench-logic-sniffer
158 - rigol-ds (for RS232; not required for USBTMC or TCP)
159 - serial-dmm (including all subdrivers)
160 - serial-lcr (including all subdrivers)
161 - teleinfo
162 - tondaj-sl-814
163
164The following drivers/devices do not require a serial port specification:
165
166 - asix-sigma
167 - brymen-bm86x
168 - chronovu-la
169 - demo
170 - fx2lafw
171 - hantek-dso
172 - ikalogic-scanalogic2
173 - ikalogic-scanaplus
174 - kecheng-kc-330b
175 - lascar-el-usb
176 - pipistrello-ols
177 - rigol-ds (USBTMC or TCP)
178 - saleae-logic16
179 - sysclk-lwla
180 - uni-t-dmm (including all subdrivers)
181 - uni-t-ut32x
182 - victor-dmm
183 - yokogawa-dlm (USBTMC or TCP)
184 - zeroplus-logic-cube
185
186
187Specifying serial port parameters
188---------------------------------
189
190Every serial device's driver has default serial port parameters like baud
191rate, number of data bits, stop bits and handshake status. If a device requires
192different parameters, pass them as option "serialcomm" with the driver name.
193See libsigrok docs for the function serial_set_paramstr() for complete specs.
194
195Example:
196
197 $ sigrok-cli --driver <somedriver>:conn=<someconn>:serialcomm=9600/7n1/dtr=1
198
199
200Permissions of serial port based devices
201----------------------------------------
202
203When using devices supported by libsigrok that use serial port based cables
204(real RS232 or USB-to-serial ones) to connect to a PC, you need to ensure
205that the user running the libsigrok frontend has (read/write) permissions to
206access the serial port device (e.g. /dev/ttyS0, /dev/ttyUSB0, and so on).
207
208You can use 'chmod' to apply permissions as you see fit, and/or 'chown' to
209change the owner of the serial port device to a certain user or group.
210
211For USB-to-serial based devices, we recommended using our udev rules file
212(see below for details).
213
214
215Permissions for USB devices (udev rules file)
216---------------------------------------------
217
218When using USB-based devices supported by libsigrok, the user running the
219libsigrok frontend (e.g. sigrok-cli) has to have (read/write) permissions
220for the respective USB device.
221
222On Linux, this is accomplished using either 'chmod' (not recommended) or
223using the udev rules file shipped with libsigrok (recommended).
224
225The file is available in contrib/z60_libsigrok.rules. It contains entries
226for all libsigrok-supported (USB-based) devices and changes their group
227to 'plugdev' and the permissions to '664'.
228
229When using a libsigrok package from your favorite Linux distribution, the
230packager will have already taken care of properly installing the udev file
231in the correct (distro-specific) place, and you don't have to do anything.
232The packager might also have adapted 'plugdev' and '664' as needed.
233
234If you're building from source, you need to copy the file to the place
235where your distro expects such files. This is beyond the scope of this README,
236but generally the location could be e.g. /etc/udev/rules.d, or maybe
237/lib/udev/rules.d, or something else. Afterwards you might have to restart
238udev, e.g. via '/etc/init.d/udev restart' or similar, and you'll have to
239re-attach your device via USB.
240
241Please consult the udev docs of your distro for details.
242
243
244Cypress FX2 based devices
245-------------------------
246
247Devices using the Cypress FX2(LP) chip without any specific USB VID/PID will
248be enumerated with VID/PID 04b4:8613 (the default for "unconfigured FX2").
249These are usually "FX2 eval boards" (that can also be used as LAs, though).
250
251On Linux, the 'usbtest' driver will usually grab such devices, and they will
252thus not be usable by libsigrok (and frontends).
253
254You can fix this by running 'rmmod usbtest' as root before using the device.
255
256
257UNI-T DMM (and rebranded models) cables
258---------------------------------------
259
260UNI-T multimeters (and rebranded devices, e.g. some Voltcraft models) can
261ship with different PC connectivity cables:
262
263 - UT-D04 (USB/HID cable with Hoitek HE2325U chip, USB VID/PID 04fa:2490)
264 - UT-D04 (USB/HID cable with WCH CH9325 chip, USB VID/PID 1a86:e008)
265 - UT-D02 (RS232 cable)
266
267The above cables are all physically compatible (same IR connector shape)
268with all/most currently known UNI-T multimeters. For example, you can
269use either of the UT-D04 USB/HID cables or the UT-D02 RS232 cable with
270the UNI-T UT61D multimeter.
271
272When using the UT-D02 RS232 cable with any of the supported UNI-T DMMs,
273you have to use the respective driver with a '-ser' drivername suffix
274(internally all of these models are handled by the 'serial-dmm' driver).
275
276You also need to specify the serial port via the 'conn' option, e.g.
277/dev/ttyUSB0 (attached via a USB-to-serial cable) or /dev/ttyS0 (actual
278RS232 port) on Linux (see above).
279
280Finally, the user running the frontend (e.g. sigrok-cli) also needs
281permissions to access the respective serial port (see above).
282
283Examples (sigrok-cli):
284
285 $ sigrok-cli --driver uni-t-ut61e-ser:conn=/dev/ttyUSB0 ...
286 $ sigrok-cli --driver voltcraft-vc820-ser:conn=/dev/ttyS0 ...
287
288When using any of the UT-D04 USB/HID cables you have to use the respective
289driver _without_ the '-ser' drivername suffix (internally all of these models
290are handled by the 'uni-t-dmm' driver).
291
292You also need to specify the USB vendor/device IDs of the cable.
293Autodetection is not possible here, since various other products use the
294USB VID/PID of those cables too, and there is no way to distinguish them.
295
296Since the UT-D04 cables are USB based (but don't use a USB-to-serial chip)
297there is no need to specify a serial port via 'conn', of course.
298However, the user running the frontend does also need to have permissions
299to access the respective USB device (see above).
300
301Examples (sigrok-cli):
302
303 $ sigrok-cli --driver uni-t-ut61e:conn=1a86.e008 ...
304 $ sigrok-cli --driver voltcraft-vc820:conn=04fa.2490 ...
305
306
307UNI-T UT-D04 cable issue on Linux
308---------------------------------
309
310The UNI-T UT-D04 cable with Hoitek HE2325U (or WCH CH9325) chip seems to have
311a very specific problem on Linux. Apparently it requires to be put into
312suspend (and woken up again) before it is usable. This seems to be a
313Linux-only issue, Windows is not affected by this since apparently the
314Windows kernel does this for every USB device, always.
315
316Thus, if you want to use any of the UNI-T DMMs with this specific cable,
317you'll have to run the following script (as root) once, every time you attach
318the cable via USB. The script was written by Ralf Burger.
319
320See also: http://erste.de/UT61/index.html
321
322 #!/bin/bash
323 for dat in /sys/bus/usb/devices/*; do
324 if test -e $dat/manufacturer; then
325 grep "WCH.CN" $dat/manufacturer > /dev/null && echo auto > ${dat}/power/level && echo 0 > ${dat}/power/autosuspend
326 fi
327 done
328
329
330Enabling multimeter / data logger measurement output
331----------------------------------------------------
332
333Some multimeters or data loggers will not start outputting measurement data
334unless a certain action has been performed by the user beforehand. This is
335usually mentioned in the vendor manual of the respective device, but here's
336a short list for convenience:
337
338 - BBC Goertz Metrawatt M2110: Briefly press the "Start/Reset" button on the
339 interface panel on top.
340 - Digitek DT4000ZC: Briefly press the "RS232" button.
341 - Gossen Metrawatt Metrahit 1x/2x devices, driver gmc-mh-1x-2x-rs232:
342 - Power on the device with the "DATA" button pressed.
343 - Metrahit 2x devices must be configured for the respective interface type.
344 - Gossen Metrawatt Metrahit 2x devices, driver gmc-mh-2x-bd232:
345 - 'BD232' interface:
346 The multimeter must be configured for the respective interface type.
347 - 'SI232-II' interface ("PC Mode"):
348 The multimeter must be configured for interface type 'BD232' (all),
349 'SI232 online' (28-29S) or 'SI232 store' (22-26x). The interface must
350 be configured to the same baud rate as the host (default 9600).
351 Multimeter and interface must be configured to the same address.
352 - Norma DM950: If the interface doesn't work (e.g. USB-RS232 converter), power
353 on the device with "FUNC" pressed (to power the interface from the DMM).
354 - PCE PCE-DM32: Briefly press the "RS232" button.
355 - RadioShack 22-812: Press and hold "SELECT" and "RANGE" together.
356 - TekPower TP4000ZC: Briefly press the "RS232" button.
357 - Tenma 72-7750: Briefly press the "RS232C" button.
358 - UNI-T UT60G: Briefly press the "RS232C" button.
359 - UNI-T UT61B/C/D: Press the "REL/RS232/USB" button for roughly 1 second.
360 - UNI-T UT71x: Press the "SEND/-/MAXMIN" button for roughly 1 second.
361 Briefly pressing the "EXIT" button leaves this mode again.
362 - UNI-T UT325: Briefly press the "SEND" button (as per manual). However, it
363 appears that in practice you don't have to press the button (at least on
364 some versions of the device), simply connect the device via USB.
365 - V&A VA18B/VA40B: Keep the "Hz/DUTY" key pressed while powering on the DMM.
366 - Victor 70C/86C: Press the "REL/RS232" button for roughly 1 second.
367 - Voltcraft VC-830: Press the "REL/PC" button for roughly 2 seconds.
368 - Voltcraft VC-870: Press the "REL/PC" button for roughly 1 second.
369
370
371ChronoVu LA8/LA16 USB VID/PIDs
372------------------------------
373
374The ChronoVu LA8/LA16 logic analyzer is available in two revisions. Previously,
375the device shipped with a USB VID/PID of 0403:6001, which is the standard ID
376for FTDI FT232 USB chips.
377
378Since this made it hard to distinguish the LA8/LA16 from any other device
379with this FTDI chip connected to the PC, the vendor later shipped the
380device with a USB VID/PID of 0403:8867.
381
382The 'chronovu-la' driver in libsigrok supports both VID/PID pairs and
383automatically finds devices with either VID/PID pair.
384
385
386OLS
387---
388
389The Dangerous Prototypes Openbench Logic Sniffer (OLS) logic analyzer
390driver in libsigrok assumes a somewhat recent firmware has been flashed onto
391the OLS (it doesn't need a firmware upload every time it's attached via USB,
392since the firmware is stored in the device permanently).
393
394The most recent firmware version that is tested is 3.07.
395
396If you use any older firmware and your OLS is not found or is not working
397properly, please upgrade to at least this firmware version. Check the
398Dangerous Prototypes wiki for firmware upgrade instructions:
399
400 http://dangerousprototypes.com/docs/Logic_Sniffer_upgrade_procedure
401
402Also, you need to specify a serial port for the OLS in the frontends, e.g.
403using the 'conn' option in sigrok-cli, and you also need to have the
404permissions to access the serial port (see above).
405