]> sigrok.org Git - libsigrok.git/blobdiff - src/dmm/ut71x.c
several DMMs: set DC flag for diode mode
[libsigrok.git] / src / dmm / ut71x.c
index 0f0843e087ee2b7a577e738ac98124f37248bdc5..6dbf23a47e1cbb5ca5f4881d24d4cc1c8fd9ba92 100644 (file)
@@ -44,7 +44,7 @@ static const int exponents[16][8] = {
        {  0,  -4,  -3,  -2, -1,  0,  0,  0 }, /* DC V */
        {  0,  -4,  -3,  -2, -1,  0,  0,  0 }, /* AC V */
        { -5,   0,   0,   0,  0,  0,  0,  0 }, /* DC mV */
-       {  0,  -1,   0,   1,  2,  3,  4,  0 }, /* Resistance */
+       {  0,  -2,  -1,   0,  1,  2,  3,  0 }, /* Resistance */
        {  0, -12, -11, -10, -9, -8, -7, -6 }, /* Capacitance */
        { -1,   0,   0,   0,  0,  0,  0,  0 }, /* Temp (C) */
        { -8,  -7,   0,   0,  0,  0,  0,  0 }, /* uA */
@@ -81,9 +81,21 @@ static int parse_value(const uint8_t *buf, struct ut71x_info *info, float *resul
                       buf[0], buf[1], buf[2], buf[3], buf[4]);
                return SR_ERR;
        }
+
        for (i = 0, intval = 0; i < num_digits; i++)
                intval = 10 * intval + (buf[i] - '0');
 
+       /*
+        * For measurements that only have 4000 instead of 40000 counts
+        * (resistance, continuity) we have to use an additional factor of 10.
+        *
+        * This seems to vary between DMMs. E.g. the Voltcraft VC920 and VC940
+        * have 4000 counts for resistance, whereas the Tenma 72-9380A,
+        * 72-7730 and 72-7732 have 40000 counts for resistance.
+        */
+       if (num_digits == 4)
+               intval *= 10;
+
        /* Apply sign. */
        intval *= info->is_sign ? -1 : 1;
 
@@ -272,7 +284,7 @@ static void handle_flags(struct sr_datafeed_analog *analog,
        if (info->is_auto)
                analog->meaning->mqflags |= SR_MQFLAG_AUTORANGE;
        if (info->is_diode)
-               analog->meaning->mqflags |= SR_MQFLAG_DIODE;
+               analog->meaning->mqflags |= SR_MQFLAG_DIODE | SR_MQFLAG_DC;
 }
 
 static gboolean flags_valid(const struct ut71x_info *info)