-ALSA driver
------------
-
-The 'alsa' driver can be used to sample analog data using a PC's soundcard.
-I.e. the sound card can act as a simple oscilloscope (with some limitations)
-using commercial or DIY "sound card scope probe" cables.
-
-Since ALSA is a Linux-specific sound system, this driver will inherently
-only compile and work on Linux.
-
-We might write additional drivers to make a similar functionality available
-on other OSes at some point.
-
-
-Link Instruments MSO-19
------------------------
-
-The driver for the Link Instruments MSO-19 mixed-signal oscilloscope (MSO)
-is currently using libudev (which is a library for the Linux-specific udev
-system). Thus, this driver can currently only compile and work on Linux.
-
-We plan to fix this though, so that it'll work on all other OSes, too.
-
-
-ChronoVu LA8 USB VID/PIDs
--------------------------
-
-The ChronoVu LA8 logic analyzer is available in two revisions. Previously,
-the LA8 shipped with a USB VID/PID of 0403:6001, which is the standard ID
+ - Voltcraft VC-870: Press the "REL/PC" button for roughly 1 second.
+
+
+ASIX OMEGA in RTM CLI mode
+--------------------------
+
+The asix-sigma driver can detect the Omega devices' presence, but does
+not support their protocol and emits a diagnostics message. The firmware
+image is not available for distribution, and information on the protocol
+is not available. That's why native support is in some distant future.
+Yet basic operation of Omega devices is available by using the vendor's
+command line application for real time mode (RTM CLI).
+
+The vendor application targets Windows (on x86), but also executes on
+Linux when 32bit libraries for FTDI communication are provided. The
+user manual discusses the installation. The sigrok asix-omega-rtm-cli
+driver uses the vendor provided omegartmcli.exe binary to configure the
+device for streaming, and to acquire sample data.
+
+Either make an "omegartmcli" executable available in PATH. This can be
+the vendor's executable or some wrapper around it or a symlink to it.
+Or specify the executable's location in the OMEGARTMCLI environment
+variable. The sigrok driver accepts an optional serial number (six or
+eight hex digits) to select one out of several connected devices.
+
+ (optional)
+ $ export "OMEGARTMCLI=$HOME/.wine/drive_c/progx86/ASIX/SIGMA/omegartmcli.exe"
+
+ (optional)
+ $ OMEGASN=":conn=sn=a6030123"
+
+ (example use)
+ $ sigrok-cli -d asix-omega-rtm-cli${OMEGASN} --show
+ $ sigrok-cli -d asix-omega-rtm-cli${OMEGASN} -o capture.sr --time 10s
+ $ sigrok-cli -d asix-omega-rtm-cli${OMEGASN} -o capture.sr --samples 100m
+ $ pulseview -d asix-omega-rtm-cli${OMEGASN}
+
+The RTM mode of operation samples 16 channels at a fixed rate of 200MHz.
+Hardware triggers are not available in this mode. Glib should handle
+platform specific details of external process execution, but the driver
+was only tested on Linux so far. Acquisition start in sigrok applications
+may take some time before sample data becomes available (roughly one
+second here on a slow machine). This is an implementation detail of the
+RTM CLI approach including execution under wine.
+
+The reliability of that setup in the presence of fast changing input
+signals is yet to get determined. It's assumed that slow input signals
+are operational. It's essential that the _average_ rate of changes in
+the input signal in combination with the hardware compression are such
+that the FTDI FIFO can communicate all involved data via USB2.0 to the
+application. Intermediate bursts of rapid changes shall not be an issue
+given the Omega devices' deep memory which RTM uses for buffering.
+
+Native support for the Asix Omega devices depends on the availability of
+a protocol description and use of the protocol depends on the firmware's
+availability at the user's site. Which then would allow to capture to
+DRAM at high rates without the communication bottleneck, before the data
+gets communicated to the PC after the acquisition has completed. Compare
+the native sigrok support for Asix Sigma.
+
+
+ChronoVu LA8/LA16 USB VID/PIDs
+------------------------------
+
+The ChronoVu LA8/LA16 logic analyzer is available in two revisions. Previously,
+the device shipped with a USB VID/PID of 0403:6001, which is the standard ID