Bug 660 - On some runs incorrect data is shown on CH2
Summary: On some runs incorrect data is shown on CH2
Status: CONFIRMED
Alias: None
Product: libsigrok
Classification: Unclassified
Component: Driver: rigol-ds (show other bugs)
Version: unreleased development snapshot
Hardware: All All
: Normal normal
Target Milestone: ---
Assignee: Nobody
URL:
Keywords:
Depends on:
Blocks:
 
Reported: 2015-09-23 15:32 CEST by Uwe Hermann
Modified: 2015-09-23 15:32 CEST (History)
0 users



Attachments

Note You need to log in before you can comment on or make changes to this bug.
Description Uwe Hermann 2015-09-23 15:32:00 CEST
How to reproduce (on a DS1052E with firmware < 2.4 here):

- Use a function generator to put a sine wave on CH1, a rectangle waveform on CH2.
- Start PulseView, select the detected DS1052E.
- Press Run.
- The two waveforms (sine + rectangle) should be displayed correctly.
- Press run again and again...
- Eventually, after several correct runs, it can happen that there's one run where the data is *not* correctly displayed anymore.
- In this specific case it was observed that the CH2 data would no longer be a rectangle but rather "maxed out" (hanging at a high voltage level). The CH1 data was still correctly displayed (sine).

Pressing Run yet again, usually leads to another working run. Continuing to press Run a few more times will eventually yield the above error case again, though.