Voltcraft M-3650CR

From sigrok
Jump to navigation Jump to search
The printable version is no longer supported and may have rendering errors. Please update your browser bookmarks and please use the default browser print function instead.
Voltcraft M-3650CR
Voltcraft m-3650cr.png
Status supported
Source code serial-dmm
Counts 2000
IEC 61010-1
Connectivity RS232
Measurements voltage, current, resistance, capacitance, hFE, diode, continuity, frequency, logic
Features data hold, bargraph, difference

The Voltcraft M-3650CR is a 2000 counts handheld digital multimeter with RS232 connectivity.

It came into the market 1991 and is a rebadged Metex M-3650CR.

Hardware

Multimeter:

  • Fuse: 2A/250V (5x20mm) (for the A jack; the 20A jack is unfused!)
  • TODO.

RS232 cable:

Photos

Multimeter:

RS232 cable:

See Device_cables#Metex_5-pin_RS232_cable.

Protocol

The protocol is (partially) documented in the manual of the Voltcraft M-3650CR and was determined completely by testing. See Multimeter ICs#Alternative Protocol.

Resources

Measurement functions and ranges

Measurement function Range Resolution Accuracy sigrok result Remarks
V= 200.0 mV 100 μV ±(0.3% + 1) mq V, unit V, mqflags DC
2.000 V 1 mV "
20.00 V 10 mV "
200.0 V 100 mV "
1000 V 1 V "
V≈ 200.0 mV 100 μV ±(0.8% + 3) mq V, unit V, mqflags AC
2.000 V 1 mV "
20.00 V 10 mV "
200.0 V 100 mV "
750 V 1 V ±(1.2% + 3) "
A= 2.000 mA 1 µA ±(0.5% + 1) mq Curr., unit A, mqflags DC
200.0 mA 100 µA ±(1,2% + 1) "
20.00 A 10 mA ±(2,0% + 5) " Max. 20 min.; unfused!
A≈ 2.000 mA 10 µA ±(1,0% + 5) mq Curr., unit A, mqflags AC
200.0 mA 100 µA ±(1,8% + 5) "
20.00 A 10 mA ±(3.0% + 7) " Max. 20 min.; unfused!
Ω 200.0 Ω 100 mΩ mq Res., unit Ω, mqflags -
2.000 kΩ 1 Ω "
20.00 kΩ 10 Ω "
200.0 kΩ 100 Ω "
2.000 MΩ 1 kΩ "
20.00 MΩ 10 kΩ "
Cap. F 2000 pF 1 pF ±(2.0% + 3) mq Cap., unit F, mqflags -
200.0 nF 100 pF "
20.00 µF 10 nF ±(3.0% + 5) "
Hz= 20.00 kHz 1 Hz ±(2% + 3) mq Freq., unit Hz, mqflags DC
200.00 kHz 10 Hz "
Logic -
hFE ? IB=10 µA, UCE=2,8 V
Diode mq V, unit V, mqflags DC,DIODE
Data Hold - Not supported by protocol.

The column "sigrok result" contains in short form what the driver generates for the resp. data type.