Difference between revisions of "Hantek 1008C"

From sigrok
Jump to navigation Jump to search
m
(3 intermediate revisions by the same user not shown)
Line 8: Line 8:
| samplerate_state = ?
| samplerate_state = ?
| triggers        = ?
| triggers        = ?
| voltages        = ?
| voltages        = ±200mV / ±15V
| memory          = ?
| memory          = ?
| compression      = ?
| compression      = ?
Line 14: Line 14:
}}
}}


The '''Hantek 1008C''' is a USB-based 8-channel oscilloscope, with 8 channel digital pattern generator.
The '''Hantek 1008C''' is a USB-based 8-channel oscilloscope (sampling resolution: 12bits on each channel), with 8 channel digital pattern generator.


See [[Hantek 1008C/Info]] for some more details (such as '''lsusb -vvv''' output) on the device.
See [[Hantek 1008C/Info]] for some more details (such as '''lsusb -vvv''' output) on the device.
Line 82: Line 82:


=== Start of waiting cycle for data to be ready into the buffers ===
=== Start of waiting cycle for data to be ready into the buffers ===
F3
* F3
A2 01 01 01 01 01 01 01 01
* A2 01 01 01 01 01 01 01 01
A4 01
* A4 01
C0
* C0
C2
* C2
A5 5A
* A5 5A
A5 5A
* A5 5A


=== Buffers Reading Cycle ===
=== Buffers Reading Cycle ===
Line 98: Line 98:
== Notes ==
== Notes ==


* Sample rate: maximum measured 2.4MS/s if enable just 1 channel. Drop down to 1.2MS/s if enable 2 channels.
* Sample rate: maximum measured 2.4MS/s if just 1 channel is enabled. Drop down to 1.2MS/s if enable 2 channels.
* Acquisition: on each scan acquire 4000 samples from 2 buffers (2000 samples each buffer)  
* Acquisition: on each scan acquire 4000 samples from 2 buffers (2000 samples each buffer)


== Resources ==
== Resources ==

Revision as of 03:52, 21 November 2017

Hantek 1008C
Hantek 1008C.png
Status planned
Channels 8
Samplerate 2.4MS/s (see notes)
Samplerate (state) ?
Triggers ?
Min/max voltage ±200mV / ±15V
Memory ?
Compression ?
Website hantek.com

The Hantek 1008C is a USB-based 8-channel oscilloscope (sampling resolution: 12bits on each channel), with 8 channel digital pattern generator.

See Hantek 1008C/Info for some more details (such as lsusb -vvv output) on the device.

Hardware

  • CPU: ST Microelectronics STM32F103C6T6A
  • Low-dropout voltage regulator: Advanced Monolithic Systems AMS1117-3.3
  • 8-channel analog multiplexer/demultiplexer: NXP Semiconductor 74HCT4051D (1 per analog channel)
  • Octal 3-state buffer: Fairchild Semiconductor MM74HC244SJ (in the path of digital outputs)
  • 8-bit shift register with 3-state output: 74HC595 (3 total)
  • Dual opamp: in MSOP-8 package, Texas Instruments unknown model (chip markings "TI 25 AVG") (1 per analog channel)

Photos

Protocol

Need to continuously send commands to the OUT Endpoint, and for each command receive the result from the IN Endpoint (both Endpoints can transfer max 64 bytes). The device got logically disconnected automatically if not receive any command after 7 seconds.

List of commands (incomplete)

  • AA -> set which channel to enable (ie. AA 01 01 00 00 00 00 00 00 enable channel 1 and 2)
  • A2 -> set voltage level for each channel (ie. A2 03 03 03 03 03 03 03 03 apply +-15V range on each channel)
  • A3 -> set sampling rate (ie. A3 11)

Sample of initialization cycle

  • B0
  • F3
  • B901BF040000
  • B700
  • BB0800
  • B5
  • B6
  • E5
  • F7
  • F8
  • FA
  • F5
  • A008
  • AA0101010101010101
  • A311
  • C10000
  • A70000
  • AC01F40009C50009C5

Start of waiting cycle for data to be ready into the buffers

  • F3
  • A2 01 01 01 01 01 01 01 01
  • A4 01
  • C0
  • C2
  • A5 5A
  • A5 5A

Buffers Reading Cycle

  • C6 02 (return Buffer 1 size)
  • A6 02 (acquire 64 bytes, 32 samples of 12 bits each, repeated 60 times get 2000 samples)
  • C6 03 (return Buffer 2 size)
  • A6 03 (acquire 64 bytes, 32 samples of 12 bits each, repeated 60 times get 2000 samples)

Notes

  • Sample rate: maximum measured 2.4MS/s if just 1 channel is enabled. Drop down to 1.2MS/s if enable 2 channels.
  • Acquisition: on each scan acquire 4000 samples from 2 buffers (2000 samples each buffer)

Resources