Difference between revisions of "Gossen Metrawatt Metrahit 25S"

From sigrok
Jump to navigation Jump to search
 
(7 intermediate revisions by the same user not shown)
Line 1: Line 1:
{{Infobox multimeter
{{Infobox multimeter
| <!-- image          = -- [[File:Siemens B1105 small.png|180px]] -->
| image          = [[File:Gossen_Metrawatt_Metrahit_25S_Logo.png|180px]]
| name            = Gossen Metrawatt Metrahit 25S
| name            = Gossen Metrawatt Metrahit 25S
| status          = in progress
| status          = supported
| source_code_dir = gmc_mh_1x_2x
| source_code_dir = gmc-mh-1x-2x
| counts          = 31000
| counts          = 31000
| categories      = CAT III 1000 V, CAT IV 600 V
| categories      = CAT III 1000 V, CAT IV 600 V
Line 36: Line 36:


== Using the device with sigrok ==
== Using the device with sigrok ==
The device is supported in ''Send Mode'' using a [[Gossen_Metrawatt#Metrahit_RS232 | RS232 interface]] by the driver gmc-mh-1x-2x-rs232. Set the multimeter to use the communication adaptor ''RS232''.
The device is supported in ''Send Mode'' using a [[Gossen_Metrawatt#Metrahit_RS232 | RS232 interface]] by the driver gmc-mh-1x-2x-rs232 (actually this should work, but with [[User:Matthias Heidbrink | Matthias']] test device it did not, so it's untested). Set the multimeter to use the communication adaptor ''RS232''.


The multimeter is also supported in bidirectional mode using a [[Gossen_Metrawatt#Metrahit_BD232 | BD232]] or [[Gossen_Metrawatt#Metrahit_SI232-II | SI232-II]] interface by the driver gmc-mh-2x-bd232.  Set the multimeter to use the communication adaptor ''BD232''.
The multimeter is also supported in bidirectional mode using a [[Gossen_Metrawatt#Metrahit_BD232 | BD232]] or [[Gossen_Metrawatt#Metrahit_SI232-II | SI232-II]] interface by the driver gmc-mh-2x-bd232.  Set the multimeter to use the communication adaptor ''BD232''.
Line 57: Line 57:
| 300 mV
| 300 mV
| 10 μV
| 10 μV
| ±(0,02% + 0,010% + 5)  
| rowspan="5" |±(0.05% 3)  
| mq V, unit V, mqflags DC
| mq V, unit V, mqflags DC
|  
|  
Line 64: Line 64:
| 3 V
| 3 V
| 100 μV
| 100 μV
| rowspan="4" | ±(0,02% + 0,005% + 5)
| "
| "
|  
|  
Line 90: Line 89:
| 300 mV
| 300 mV
| 100 µV
| 100 µV
| ±(0,5% + 30) (> 500 Digit))
| ±(0.5% + 30) (> 500 Digit))
| mq V, unit V, mqflags AC,RMS
| mq V, unit V, mqflags AC,RMS
|
|
Line 97: Line 96:
| 3 V
| 3 V
| 100 μV
| 100 μV
| rowspan="4" | ±(0,2% + 30)
| rowspan="4" | ±(0.2% + 30)
| "
| "
|  
|  
Line 122: Line 121:
| V db≈
| V db≈
| -48 - 63 dB
| -48 - 63 dB
| 0,01 dB
| 0.01 dB
| ± 0,1 dB  
| ± 0.1 dB  
| mq V, unit V, mqflags AC,RMS
| mq V, unit V, mqflags AC,RMS
| Actually the 5 AC ranges, see there for accuracy.
| Actually the 5 AC ranges, see there for accuracy.
|-
| V db≈ rel
| -48 - 63 dB
| 0,01 dB
| ± 0,1 dB
| mq V, unit V, mqflags AC,RMS
| Actually the 5 AC ranges, see there for accuracy. To activate, press yellow button in dB mode, then do second measurement. Cannot be distinguished from non-rel measurement remotely.


|-
|-
Line 139: Line 130:
| 300.00 µA
| 300.00 µA
| 10 nA
| 10 nA
| 0,05% + 0,02% + 5
| rowspan="2" | 0.1% + 5
| mq Curr., unit A, mqflags DC
| mq Curr., unit A, mqflags DC
|  
|  
Line 146: Line 137:
| 3.0000 mA
| 3.0000 mA
| 100 nA
| 100 nA
| 0,05% + 0,01% + 5
| "
| "
|  
|  
Line 153: Line 143:
| 30.000 mA
| 30.000 mA
| 1 mA
| 1 mA
| 0,05% + 0,01% + 5
| 0.05% + 5
| "
| "
|  
|  
Line 160: Line 150:
| 300.00 mA
| 300.00 mA
| 10 µA
| 10 µA
| 0,1% + 0,01% + 5
| 0.5% + 5
| "
| "
|  
|  
Line 167: Line 157:
| 3.0000 A
| 3.0000 A
| 100 µA
| 100 µA
| 0,2% + 0,05% + 5
| rowspan="2" | 0.5% + 10
| "
| "
|  
|  
Line 174: Line 164:
| 10.000 A
| 10.000 A
| 1 mA
| 1 mA
| 0,2% + 0,05% + 5
| "
| "
|  
|  
Line 182: Line 171:
| 300.00 µA
| 300.00 µA
| 10 nA
| 10 nA
| rowspan="4" | 0,5% + 30
| rowspan="4" | 0.5% + 30
| mq Curr., unit A, mqflags AC,DC,RMS
| mq Curr., unit A, mqflags AC,DC,RMS
|  
|  
Line 207: Line 196:
| 3.0000 A
| 3.0000 A
| 100 µA
| 100 µA
| 0,7% + 30
| rowspan="2" | 0.75% + 30
| "
| "
|  
|  
Line 214: Line 203:
| 10.000 A
| 10.000 A
| 1 mA
| 1 mA
| 0,5% + 30
| "
| "
|  
|  
Line 220: Line 208:
| rowspan="6" | Ω
| rowspan="6" | Ω
| 300.00 Ω
| 300.00 Ω
| 1
| 10
| rowspan="4" | 0,05% + 0.01% + 5
| rowspan="5" | 0.1% + 5
| mq Res., unit Ω, mqflags -
| mq Res., unit Ω, mqflags -
|
|
Line 227: Line 215:
|-
|-
| 3.0000 kΩ
| 3.0000 kΩ
| 10
| 100
| "
| "
|  
|  
Line 233: Line 221:
|-
|-
| 30.000 kΩ
| 30.000 kΩ
| 100 mΩ
| 1 Ω
| "
| "
|  
|  
Line 239: Line 227:
|-
|-
| 300.00 kΩ
| 300.00 kΩ
| 1 Ω
| 10 Ω
| "
| "
|  
|  
Line 245: Line 233:
|-
|-
| 3.0000 MΩ
| 3.0000 MΩ
| 10 Ω
| 100 Ω
| 0,1% + 0.02% + 5
| "
| "
|  
|  
Line 252: Line 239:
|-
|-
| 30.000 MΩ
| 30.000 MΩ
| 100 Ω
| 1k Ω
| 1% + 0.2% + 5
| 2% + 5
| "
| "
|  
|  
Line 263: Line 250:
| 1% + 3%
| 1% + 3%
| mq Res., unit Ω, mqflags -
| mq Res., unit Ω, mqflags -
|
|-
| Diode/Cont.
| 300 mV
| 100 µV
| 0.2% + 3
| mq V, unit V, mqflags DC,DIODE
|  
|  


Line 276: Line 255:
| Diode
| Diode
| 3 V
| 3 V
| 100 µV
| 1 mV
| 0.2% + 3
| 0.2% + 3
| mq V, unit V, mqflags DC,DIODE
| mq V, unit V, mqflags DC,DIODE
Line 285: Line 264:
| 3.000 nF
| 3.000 nF
| 1 pF
| 1 pF
| rowspan="5" | 1.0% + 0.2
| rowspan="5" | 1.0% + 5
| mq Cap., unit F, mqflags -
| mq Cap., unit F, mqflags -
|
|
Line 316: Line 295:
| 300.0 µF
| 300.0 µF
| 100 nF
| 100 nF
| rowspan="3" | ±(5.0% + 6)
| rowspan="2" | ±(5.0% + 6)
| "
| "
|  
|  
Line 329: Line 308:
| 30.00 mF
| 30.00 mF
| 1 µF
| 1 µF
| ±(5.0% + 30)
| "
| "
|  
|  
Line 335: Line 315:
| rowspan="3" | Hz≈, ≃
| rowspan="3" | Hz≈, ≃
| 300.00 Hz
| 300.00 Hz
| 0.001 Hz
| 0.01 Hz
| rowspan="3" | ±(0.05% + 1)
| rowspan="3" | ±(0.1% + 1)
| mq Freq., unit Hz, mqflags AC,DC
| mq Freq., unit Hz, mqflags AC,DC
| To activate in AC, press yellow button twice.
| To activate in AC, press yellow button twice.
Line 342: Line 322:
|-
|-
| 3.0000 kHz
| 3.0000 kHz
| 0.01 Hz
| 0.1 Hz
| "
| "
|  
|  


|-
|-
| 300.000 kHz
| 100.000 kHz
| 1 Hz
| 10 Hz
| "
| "
|  
|  


|-
|-
| rowspan="4" | °C
| rowspan="3" | °C
| Pt100/Pt1000<br> –200.0 - +100.0 °C
| Pt100/Pt1000<br> –200.0 - -100.0 °C
| rowspan="4" | 0.1 °C
| rowspan="3" | 0.1 °C
| ±(0,5 K + 3)
| 1 K
| mq Temp., unit °C, mqflags -
| mq Temp., unit °C, mqflags -
|
|


|-
|-
| Pt100/Pt1000<br> +100.0 - 850.0 °C
| Pt100/Pt1000<br> -100.0 - +100.0 °C
| ±(0,2% + 3)
| 0.8 K
| mq Temp., unit °C, mqflags -
| "
|
|


|-
|-
| K NiCr-Ni -270-+1372,0 °C
| Pt100/Pt1000<br> +100.0 - +850.0 °C
| ±(0,7% + 3)
| ±(0.5% + 3)
| "
| "
|
|
|-
| J Fe-CuNi -210 - +1200.0 °C
| ±(0,8% + 3)
| "
| Not tested yet due to lack of corresponding sensor.


|-
|-
Line 383: Line 357:
| "
| "
| "
| "
| mq Temp., unit °F, mqflags -
| style="background: #FFFF00" | mq Temp., unit °C (!), mqflags -
| Details and ranges like °C.
| Details and ranges like °C. The protocol does not allow to distinguish between °C and °F, so the unit is always °C.




Line 394: Line 368:
|  
|  
| style="background: #FFFF00; "| mq V, unit V, mqflags AC,RMS,REL
| style="background: #FFFF00; "| mq V, unit V, mqflags AC,RMS,REL
Counts events >= 2500 digits that last >=2s with a break >=2s. To activate, switch to voltage ACDC, press yellow button twice. Press up to 4 times to cycle parameters, press long to stop. See manual. Not useful remotely because the current measurement value is sent, not the counter.
|  To activate, switch to voltage AC, press yellow button twice. Press up to 4 times to cycle parameters, press long to stop. See manual. Not useful remotely because the voltage measurement value is sent, not the counter.
   
   
|-
|-
Line 402: Line 376:
|  
|  
| style="background: #ff0000; "| (Time data)
| style="background: #ff0000; "| (Time data)
| Max. 100 Min. To activate, select a 3V-1000V≃ range manually, press yellow button, press AUTO for start/stop, DATA for reset. Using the stop watch delivers only voltage data and sometimes seems to bring the device into a mode where it even sends invalid data packets.
| Max. 100 Min. Press AUTO for start/stop, DATA for reset. Not useful remotely because the stop watch delivers only zeros.
   
   
|-
|-
Line 430: Line 404:
|-
|-
|}
|}
The ''Accuracy'' row was simplified, see manual for exact specs. If the value contains three values, the middle value is % or range.
The ''Accuracy'' row was simplified, see manual for exact specs. If the value contains three values, the middle value is % or range.


The column "sigrok result" contains in short form what the driver generates for the resp. data type.
The column "sigrok result" contains in short form what the driver generates for the resp. data type.


[[Category:Device]]
[[Category:Device]]
[[Category:Multimeter]]
[[Category:Multimeter]]
[[Category:In Progress]]
[[Category:Supported]]

Latest revision as of 20:38, 21 September 2016

Gossen Metrawatt Metrahit 25S
Gossen Metrawatt Metrahit 25S Logo.png
Status supported
Source code gmc-mh-1x-2x
Counts 31000
IEC 61010-1 CAT III 1000 V, CAT IV 600 V
Connectivity Infrared (RS232)
Measurements voltage, current, resistance, capacitance, temperature, diode, continuity
Features autorange, data hold
Website [1]

The Gossen Metrawatt Metrahit 25S is a 4.5 digit, 31000 counts 1000V CAT III/600 V CAT IV handheld digital multimeter with IR (RS232) connectivity.

Hardware

  • Power supply:
    • Battery: 2x 1,5 V IEC R 6, IEC LR 6 or corresponding NiCd rechargeable batteries
  • Fuse:
    • Current 300 mA: FF (UR) 1,6 A/1000 V AC/DC (10 kA) 6,3 mm x 32 mm
    • Current 3A/10A: FF (UR) 10 A/1000 V AC/DC (30 kA) 10 mm x 38 mm

Photos

Using the device with sigrok

The device is supported in Send Mode using a RS232 interface by the driver gmc-mh-1x-2x-rs232 (actually this should work, but with Matthias' test device it did not, so it's untested). Set the multimeter to use the communication adaptor RS232.

The multimeter is also supported in bidirectional mode using a BD232 or SI232-II interface by the driver gmc-mh-2x-bd232. Set the multimeter to use the communication adaptor BD232.

The device might be not detected by the driver if it is in a mode where it sends data rarely, either with a huge send interval or with a measurement function that delivers a very low data rate, e.g. temperature measurements with K type sensor. Then it should be set to a mode with higher data rate during the initialisation of the driver.

Measurement functions and ranges

Measurement function Range Resolution Accuracy sigrok result Remarks
V= 300 mV 10 μV ±(0.05% 3) mq V, unit V, mqflags DC
3 V 100 μV "
30 V 1 mV "
300 V 10 mV "
1000 V 100 mV "
V≈ 300 mV 100 µV ±(0.5% + 30) (> 500 Digit)) mq V, unit V, mqflags AC,RMS
3 V 100 μV ±(0.2% + 30) "
30 V 1 mV "
300 V 10 mV "
1000 V 100 mV "
V db≈ -48 - 63 dB 0.01 dB ± 0.1 dB mq V, unit V, mqflags AC,RMS Actually the 5 AC ranges, see there for accuracy.
A=  300.00 µA 10 nA 0.1% + 5 mq Curr., unit A, mqflags DC
3.0000 mA 100 nA "
30.000 mA 1 mA 0.05% + 5 "
300.00 mA 10 µA 0.5% + 5 "
3.0000 A 100 µA 0.5% + 10 "
10.000 A 1 mA "
A≃  300.00 µA 10 nA 0.5% + 30 mq Curr., unit A, mqflags AC,DC,RMS
3.0000 mA 100 nA "
30.000 mA 1 µA "
300.00 mA 10 µA "
3.0000 A 100 µA 0.75% + 30 "
10.000 A 1 mA "
Ω 300.00 Ω 10 mΩ 0.1% + 5 mq Res., unit Ω, mqflags -
3.0000 kΩ 100 mΩ "
30.000 kΩ 1 Ω "
300.00 kΩ 10 Ω "
3.0000 MΩ 100 Ω "
30.000 MΩ 1k Ω 2% + 5 "
Ohm/Cont. 300 Ω 100 mΩ 1% + 3% mq Res., unit Ω, mqflags -
Diode 3 V 1 mV 0.2% + 3 mq V, unit V, mqflags DC,DIODE
Cap. F 3.000 nF 1 pF 1.0% + 5 mq Cap., unit F, mqflags -
30.00 nF 10 pF "
300.0 nF 100 pF "
3.000 µF 1 nF "
30.00 µF 10 nF "
300.0 µF 100 nF ±(5.0% + 6) "
3.000 mF 1 µF "
30.00 mF 1 µF ±(5.0% + 30) "
Hz≈, ≃ 300.00 Hz 0.01 Hz ±(0.1% + 1) mq Freq., unit Hz, mqflags AC,DC To activate in AC, press yellow button twice.
3.0000 kHz 0.1 Hz "
100.000 kHz 10 Hz "
°C Pt100/Pt1000
–200.0 - -100.0 °C
0.1 °C 1 K mq Temp., unit °C, mqflags -
Pt100/Pt1000
-100.0 - +100.0 °C
0.8 K "
Pt100/Pt1000
+100.0 - +850.0 °C
±(0.5% + 3) "
°F " " " mq Temp., unit °C (!), mqflags - Details and ranges like °C. The protocol does not allow to distinguish between °C and °F, so the unit is always °C.


Event count mq V, unit V, mqflags AC,RMS,REL To activate, switch to voltage AC, press yellow button twice. Press up to 4 times to cycle parameters, press long to stop. See manual. Not useful remotely because the voltage measurement value is sent, not the counter.
Stop Watch 100 ns (Time data) Max. 100 Min. Press AUTO for start/stop, DATA for reset. Not useful remotely because the stop watch delivers only zeros.
Data Hold mqflags += HOLD Disabled while IR interface is active on  Matthias' test device.
MIN mqflags += MIN Influences only display, via IR flag is set, but actual current value is delivered.
MAX mqflags += MAX Influences only display, via IR flag is set, but actual current value is delivered.

The Accuracy row was simplified, see manual for exact specs. If the value contains three values, the middle value is % or range.

The column "sigrok result" contains in short form what the driver generates for the resp. data type.